- ICH GCP
- Amerikanska kliniska prövningsregistret
- Klinisk prövning NCT06453850
Safety and Performance Evaluation of a New Catheter Range for Lead Implantation at Interventricular Septum (POLARIS)
The purpose of the clinical investigation is to evaluate the safety of the FLEXIGO delivery catheter for transvenous pacing lead implantation.
Studieöversikt
Status
Betingelser
Intervention / Behandling
Studietyp
Inskrivning (Faktisk)
Fas
- Inte tillämpbar
Kontakter och platser
Studieorter
-
-
-
Annecy, Frankrike
- Centre Hospitalier Annecy Genevois
-
Clermont-Ferrand, Frankrike
- CHU Clermont-Ferrand
-
Montpellier, Frankrike
- Clinique du Millénaire
-
Rouen, Frankrike
- CHU Rouen Hôpital Charles Nicolle
-
-
-
-
-
Ferrara, Italien
- Arcispedale Sant'Anna - AO Univ.
-
Novara, Italien
- Ospedale Maggiore della Carità - AO Univ.
-
Taranto, Italien
- Ospedale SS Annunziata
-
Treviso, Italien
- Ospedale Ca' Foncello
-
-
-
-
-
Coimbra, Portugal
- ULS de Coimbra
-
Lisbon, Portugal
- ULS Santa Maria
-
Lisbon, Portugal
- ULS Lisboa Ocidental - Hospital de Santa Cruz
-
-
-
-
-
Girona, Spanien
- Hospital Universitari de Girona Doctor Josep Trueta
-
Granada, Spanien
- Hospital Universitario Virgen de las Nieves
-
Las Palmas de Gran Canaria, Spanien
- Hospital Universitario de Gran Canaria Doctor Negrín
-
Murcia, Spanien
- Hospital Clínico Universitario Virgen de la Arrixaca
-
Valencia, Spanien
- Hospital Universitari i Politècnic La Fe
-
-
-
-
-
Bad Bevensen, Tyskland
- Herz- und Gefäßzentrum Bad Bevensen
-
-
Deltagandekriterier
Urvalskriterier
Åldrar som är berättigade till studier
Tar emot friska volontärer
Kön som är behöriga för studier
Beskrivning
Inclusion criteria (phase I):
- Patient indicated for cardiac pacing according to the most recent guidelines from the European Society of Cardiology (ESC)
- Patient planned for a de novo implantation of any Single Chamber (SR) or Dual Chamber (DR) CE marked pacemaker
- Patient planned for a catheter-guided implantation in the interventricular septum area
- Patient planned for the implantation of any CE marked ventricular pacing lead compatible with the FLEXIGO delivery catheter
- Patient reviewed, signed and dated the Informed Consent Form (ICF)
Inclusion criteria (phase II):
- Patient indicated for cardiac pacing or CRT according to the most recent guidelines from the ESC
- Patient planned for a de novo implantation of any SR or DR CE marked pacemaker, or any CE marked CRT-D
- Patient planned for a catheter-guided implantation in the interventricular septum area
- Patient planned for the implantation of any CE marked ventricular pacing lead compatible with the FLEXIGO delivery catheter
- Patient reviewed, signed and dated the ICF
Exclusion criteria (phase I):
- Patient planned for a device upgrade, or a device or a lead replacement
- Patient with a congenital heart disease
- Patient with significant hypertrophic cardiomyopathy, cardiac amyloidosis or obstructive hypertrophic cardiomyopathy
- Patient with a Left Ventricular Ejection Fraction (LVEF) ≤ 35%
- Patient already enrolled in another clinical investigation that could confound the results of this clinical investigation (e.g., clinical investigations involving intra-cardiac device)
- Patient implanted despite a contraindication for cardiac pacing system implantation (i.e. patient implanted with a defibrillator, patient contraindicated for a single dose of 310 μg of dexamethasone sodium phosphate, patient implanted with a tricuspid replacement heart valve, with prior tricuspid valve intervention or with a significant tricuspid valve disease that may lead to future replacement heart valve surgery)
- Minor age patient (i.e. under 18 years of age)
- Incapacitated patient, under guardianship, kept in detention, refusing to cooperate or not able to understand the purpose of this clinical investigation
- Non-menopausal women
Exclusion criteria (phase II):
- Patient planned for a device upgrade, or a device or a lead replacement
- Patient with a congenital heart disease
- Patient with significant hypertrophic cardiomyopathy, cardiac amyloidosis or obstructive hypertrophic cardiomyopathy
- Patient already enrolled in another clinical investigation that could confound the results of this clinical investigation (e.g., clinical investigations involving intra-cardiac device)
- Patient implanted despite a contraindication for cardiac pacing system implantation (i.e. patient implanted with a defibrillator, patient contraindicated for a single dose of 310 μg of dexamethasone sodium phosphate, patient implanted with a tricuspid replacement heart valve, with prior tricuspid valve intervention or with a significant tricuspid valve disease that may lead to future replacement heart valve surgery)
- Minor age patient (i.e. under 18 years of age)
- Incapacitated patient, under guardianship, kept in detention, refusing to cooperate or not able to understand the purpose of this clinical investigation
- Non-menopausal women
Studieplan
Hur är studien utformad?
Designdetaljer
- Primärt syfte: Behandling
- Tilldelning: N/A
- Interventionsmodell: Enskild gruppuppgift
- Maskning: Ingen (Open Label)
Vapen och interventioner
Deltagargrupp / Arm |
Intervention / Behandling |
|---|---|
|
Experimentell: FLEXIGO leveranskateter
|
Use of FLEXIGO delivery catheter for implantation of ventricular pacing lead |
Vad mäter studien?
Primära resultatmått
Resultatmått |
Åtgärdsbeskrivning |
Tidsram |
|---|---|---|
|
Absence of Unanticipated Serious Adverse Device Effect (USADE) related to FLEXIGO delivery catheter
Tidsram: 72 hours after implantation
|
Any event caused by use of the FLEXIGO delivery catheter or any accessory packaged with the catheter, will be considered as an ADE related to the FLEXIGO delivery catheter. The primary endpoint evaluation will be based on Sponsor assessment of safety events, based on investigational site reporting. |
72 hours after implantation
|
Sekundära resultatmått
Resultatmått |
Åtgärdsbeskrivning |
Tidsram |
|---|---|---|
|
Implantation success rates
Tidsram: During implantation
|
- Implantation success rate with FLEXIGO delivery catheter: Proportion of lead successfully implanted at the interventricular septum with use of the FLEXIGO delivery catheter. - Left Bundle Branch Area Pacing (LBBAP) implantation success rate with FLEXIGO delivery catheter: Proportion of lead successfully implanted at the interventricular septum with use of the FLEXIGO delivery catheter and fulfilling LBBAP success criteria according to the operator's decision. - LBBAP implantation success rate (overall): Proportion of lead successfully implanted at the interventricular septum with use of any material and fulfilling LBBAP success criteria. |
During implantation
|
|
Implantation time
Tidsram: During implantation
|
Total procedural time and total fluoroscopic time documented per type of cardiac device implanted and per model of catheter used. |
During implantation
|
|
Implantation usability
Tidsram: During implantation
|
Implantation usability of the FLEXIGO delivery catheter documented through handling scores based on a specific questionnaire addressed to the operator after implantation. |
During implantation
|
|
Serious Adverse Device Effect (SADE) rate related to FLEXIGO delivery catheter or the slitter
Tidsram: 72 hours after implantation
|
Proportion of SADE related to the FLEXIGO delivery catheter or the slitter. |
72 hours after implantation
|
|
Left Bundle Branch Area Pacing (LBBAP) implantation success rate per indication (cardiac pacing and CRT) and per ventricular pacing lead model
Tidsram: During implantation
|
- LBBAP implantation success rate with FLEXIGO delivery catheter (documented in cardiac pacing and CRT-D populations separately): Proportion of lead successfully implanted at the interventricular septum with use of the FLEXIGO delivery catheter and fulfilling LBBAP success criteria according to the operator's decision. - LBBAP implantation success rate (overall, and documented in cardiac pacing and CRT-D populations separately): Proportion of lead successfully implanted at the interventricular septum with use of any material and fulfilling LBBAP success criteria according to the operator's decision. |
During implantation
|
Samarbetspartners och utredare
Sponsor
Studieavstämningsdatum
Studera stora datum
Studiestart (Faktisk)
Primärt slutförande (Faktisk)
Avslutad studie (Faktisk)
Studieregistreringsdatum
Först inskickad
Först inskickad som uppfyllde QC-kriterierna
Första postat (Faktisk)
Uppdateringar av studier
Senaste uppdatering publicerad (Faktisk)
Senaste inskickade uppdateringen som uppfyllde QC-kriterierna
Senast verifierad
Mer information
Termer relaterade till denna studie
Ytterligare relevanta MeSH-villkor
Andra studie-ID-nummer
- DARI01 - POLARIS
Plan för individuella deltagardata (IPD)
Planerar du att dela individuella deltagardata (IPD)?
Läkemedels- och apparatinformation, studiedokument
Studerar en amerikansk FDA-reglerad läkemedelsprodukt
Studerar en amerikansk FDA-reglerad produktprodukt
Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .