VitaFlow LIBERTY Europe

28 september 2026 uppdaterad av: Shanghai MicroPort CardioFlow Medtech Co., Ltd.

VitaFlow Liberty™ Transkateter aortaklaffsystem Klinisk uppföljningsstudie efter marknadsintroduktion

This is a prospective, single-arm, multi-center, observational, post-market clinical investigation to document the clinical safety and performance of MicroPort CardioFlow VitaFlow Liberty™ Transcatheter Aortic Valve System in the routine practice for the treatment of severe aortic valve stenosis. The primary endpoint is the composite rate of all-cause mortality and stroke with disability at 12 months.

Patients will receive transcatheter aortic valve replacement and examinations at the screening, procedure, discharge, and follow-up per local standard of care.

Studieöversikt

Status

Rekrytering

Betingelser

Studietyp

Observationell

Inskrivning (Beräknad)

222

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studiekontakt

Studera Kontakt Backup

Studieorter

      • Galway, Irland
        • Rekrytering
        • Galway University Hospital
        • Kontakt:
          • Darren Mylotte, MD
      • Castel Volturno, Italien
      • Pisa, Italien
        • Rekrytering
        • Ospedale Cisanello
        • Kontakt:
      • Potenza, Italien
        • Rekrytering
        • Azienda Ospedaliera Regionale -S.Carlo
        • Kontakt:
      • Carnaxide, Portugal
        • Rekrytering
        • Hospital de Santa Cruz
        • Kontakt:
      • Lucerne, Schweiz
        • Rekrytering
        • Luzerner Kantonsspital | Herzzentrum
        • Kontakt:
          • Dr.Stefan Toggweiler
          • Telefonnummer: 0412051111
          • E-post: ifo@luks.ch
      • A Coruña, Spanien
        • Rekrytering
        • Hospital Universitario de a Coruna
        • Kontakt:
      • Madrid, Spanien
      • Valladolid, Spanien

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

18 år och äldre (Vuxen, Äldre vuxen)

Tar emot friska volontärer

Nej

Kön som är behöriga för studier

Allt

Testmetod

Icke-sannolikhetsprov

Studera befolkning

Patients with symptomatic, severe calcified aortic stenosis who are at high surgical risk.

Beskrivning

Inclusion Criteria:

  1. Subjects of age≥ 18 years
  2. Subjects with severe, symptomatic, calcific aortic stenosis who are considered at high risk for surgical aortic valve replacement (AVR).
  3. Subject can understand the purpose of the clinical investigation, has signed voluntarily the informed consent form and is agreeing to the scheduled follow up requirements.

Exclusion Criteria:

  1. Pre-existing mechanical heart valve in aortic position
  2. A known hypersensitivity or contraindication to all anticoagulation /antiplatelet regimens (or inability to be anticoagulated for the index procedure), to nickel or titanium, to nitinol, to dairy products, to polyethylene terephthalate (PET) or contrast media
  3. Ongoing sepsis, including active endocarditis
  4. Anatomically not suitable for the VitaFlow Liberty TAV system
  5. LVEF<20%
  6. Estimated life expectancy of less than 12 months
  7. Any medical, social or psychological condition that in the opinion of the Heart Team precludes the subject from receiving transcatheter aortic valve replacement
  8. Currently participating in another drug or device trial (excluding registries) for which the primary endpoint has not been assessed
  9. Inability to comply with the clinical investigation follow-up or other clinical investigation requirements
  10. Patients temporally unable to provide written informed consent (e. g. unconscious emergency patients)

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Observationsmodeller: COHORT
  • Tidsperspektiv: PROSPECTIVE

Kohorter och interventioner

Grupp / Kohort
Intervention / Behandling
Enarms klinisk undersökning

Transcatheter Aortic Valve Replacement with VitaFlow Liberty™ Transcatheter Aortic Valve System.

Patients will receive transcatheter aortic valve replacement and examinations at the screening, procedure, discharge, and follow-up per local standard of care.

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Composite rate of all-cause mortality and stroke with disability
Tidsram: 12 months post procedure

Composite rate of all-cause mortality and stroke with disability

12 months post procedure

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Safety outcomes defined by VARC3
Tidsram: 30 days, 12 months, and 2-5 years post procedure

Rate of all-cause mortality, myocardial infarction, stroke, implanted related new and/or worsen conduction disturbances and arrhythmias, new permanent pacemaker implantation, procedure-related or valve-related hospitalization defined by VARC 3.

30 days, 12 months, and 2-5 years post procedure
Safety outcomes defined by VARC3
Tidsram: 30 days and 12 months post procedure

Rate of type 2-4 bleeding, major vascular and access-related complications, acute kidney injury (stage 2-4) defined by VARC 3

30 days and 12 months post procedure
Safety outcomes defined by VARC3
Tidsram: acute procedure (within 24 hours)

Rate of other acute procedural and technical valve related complications: conversion to surgery; unplanned use of mechanical circulatory support; implantation of multiple (>1) transcatheter valves during the index hospitalization; valve mal-position

acute procedure (within 24 hours)
Device success
Tidsram: 30 days post implantation

Device success defined by VARC 3

30 days post implantation
Device early safety
Tidsram: 30 days post implantation

Device early safety defined by VARC 3

30 days post implantation
Bioprosthesis haemodynamic function
Tidsram: Discharge (24 hours to 7 days), 30 days,12 months and 2-5 years post implantation

Bioprosthesis haemodynamic function (assessed by echocardiography) including effective orifice area (cm2).

Discharge (24 hours to 7 days), 30 days,12 months and 2-5 years post implantation
Bioprosthesis haemodynamic function
Tidsram: Discharge (24 hours to 7 days), 30 days,12 months and 2-5 years post implantation

Bioprosthesis haemodynamic function (assessed by echocardiography) including mean transvalvular gradient (mmHg).

Discharge (24 hours to 7 days), 30 days,12 months and 2-5 years post implantation
Bioprosthesis haemodynamic function
Tidsram: Discharge (24 hours to 7 days), 30 days,12 months and 2-5 years post implantation

Bioprosthesis haemodynamic function (assessed by echocardiography) including aortic regurgitation (paravalvular, central and total).

Discharge (24 hours to 7 days), 30 days,12 months and 2-5 years post implantation
Bioprosthetic valve dysfunction
Tidsram: 12 months and 2-5 years post procedure

Moderate or severe haemodynamic valve deterioration defined by VARC3

12 months and 2-5 years post procedure
Bioprosthetic valve dysfunction
Tidsram: 12 months and 2-5 years post procedure

Bioprosthetic valve failure defined by VARC3

12 months and 2-5 years post procedure
Bioprosthetic valve dysfunction
Tidsram: 12 months and 2-5 years post procedure

Clinically significant valve thrombosis defined by VARC3

12 months and 2-5 years post procedure
NYHA Classification
Tidsram: From baseline to discharge (24 hours to 7 days), 30 days, 12 months and 2-5 years post procedure

Changes in cardiac function at discharge, 30 days and 12 months and 2-5 years post implantation according to the NYHA Classification Scheme compared to baseline

From baseline to discharge (24 hours to 7 days), 30 days, 12 months and 2-5 years post procedure
Rate of Alwide Plus Balloon Catheter pre-dilation success
Tidsram: During procedure

Defined as successful access, delivery and dilation the aortic valve, and retrieval for the AlwideTM Plus balloon catheter

During procedure
Alwide Plus Balloon Catheter related adverse-event
Tidsram: Discharge
Discharge
Predictors for the permanent pacemaker implantation
Tidsram: Baseline or intraprocedure

Including but not limited to pre-existing conduction disturbance, implant depth, LVOT calcium volume , post-dilatation

Baseline or intraprocedure

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Utredare

  • Huvudutredare: Darren Mylotte, MD, Galway University Hospital

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

6 maj 2025

Primärt slutförande (Beräknad)

6 juni 2028

Avslutad studie (Beräknad)

1 juni 2033

Studieregistreringsdatum

Först inskickad

21 juli 2024

Först inskickad som uppfyllde QC-kriterierna

30 juli 2024

Första postat (Faktisk)

2 augusti 2024

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

1 oktober 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

28 september 2026

Senast verifierad

1 december 2025

Mer information

Termer relaterade till denna studie

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

NEJ

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .

Prenumerera