Utvärdering av passiv passform i implantatburna fullbågsproteser med tre digitala arbetsflöden, inklusive ett automatiserat AI-assisterat protokoll: En randomiserad klinisk prövning (MFS03-Pdental)

20 november 2026 uppdaterad av: Miguel Gómez Polo, Universidad Complutense de Madrid

The goal of this clinical trial is to evaluate and compare three digital workflows for the fabrication of definitive implant-supported full-arch prostheses in adult patients requiring fixed implant rehabilitation.

The main questions it aims to answer are:

  • Does an automated AI-assisted digital workflow improve the passive fit of definitive full-arch implant-supported prostheses compared with manual and splint-guided alignment workflows?
  • Are there differences in marginal, geometric, mechanical, and radiographic passivity among the three digital workflows? Researchers will compare manual CBCT-STL alignment, splint-guided alignment, and automated AI-assisted CBCT-STL alignment to see if the degree of digital workflow automation affects the passive fit of definitive full-arch implant-supported prostheses.

Participants will:

  • Be adults (18 years and older) indicated for fixed implant-supported full-arch rehabilitation.
  • Receive a definitive, screw-retained, full-arch implant-supported prosthesis fabricated using one of the three assigned digital workflows.
  • Undergo standardized clinical and radiographic assessments at the time of definitive prosthesis placement to evaluate prosthesis passive fit.

Studieöversikt

Detaljerad beskrivning

This randomized controlled clinical trial aims to evaluate and compare the passive fit of definitive full-arch implant-supported prostheses fabricated using three different digital workflows with increasing levels of automation for implant position registration and prosthesis fabrication.

Full-arch implant-supported rehabilitations require a high level of precision to ensure passive fit between the prosthetic framework and the implant-abutment connections. Inadequate passive fit may lead to mechanical complications, biological overload, or long-term prosthetic failure. Digital workflows combining intraoral scanning (IOS) and cone-beam computed tomography (CBCT) have been introduced to improve accuracy; however, differences in data acquisition and dataset alignment strategies may influence the final prosthetic fit.

The study will include adult patients (18 years and older) indicated for fixed implant-supported full-arch rehabilitation in the maxilla or mandible. Eligible participants will be recruited from the Faculty of Dentistry of the Complutense University of Madrid and associated clinical centers. After providing written informed consent, participants will be randomly assigned in a 1:1:1 ratio to one of three digital workflows:

  1. Manual CBCT-STL alignment (MedicalFit 1.0): implant positions are obtained by combining intraoral scans and CBCT data, with dataset registration performed manually by the operator based on visual alignment of scan bodies.
  2. Splint-guided alignment (MedicalFit 2.0): a calibrated rigid reference splint with metallic cylinders is used to stabilize implant positions and assist manual dataset registration, aiming to reduce operator-dependent variability.
  3. Automated AI-assisted CBCT-STL alignment (MedicalFit 3.0 - Pdental): dataset registration and passivation are performed automatically by dedicated software using advanced algorithms for implant detection and alignment, without manual intervention.

All participants will receive a definitive, screw-retained, full-arch implant-supported prosthesis fabricated according to the assigned digital workflow. Prosthetic materials and clinical procedures will follow standard clinical practice. No outcome measures will be assessed during provisional prosthetic phases.

Passive fit will be evaluated exclusively at the time of definitive prosthesis placement using a standardized multi-assessment clinical approach performed at each implant-prosthesis connection.

Marginal passive fit will be assessed clinically by direct inspection using a calibrated periodontal probe. The presence or absence of marginal discrepancies between the prosthesis and the transepithelial abutment will be evaluated based on visual and tactile criteria, including the ability of the probe to penetrate the prosthesis-abutment interface.

Geometric passive fit will be evaluated using the modified Sheffield test. With all prosthetic screws loosened except for one distal screw, the presence of any lifting or separation of the prosthetic framework at the non-tightened connections will be assessed visually and tactually using an explorer probe.

Mechanical passive fit will be assessed through the tactile sensitivity of the passive screw test. Resistance perceived during screw tightening will be evaluated qualitatively to identify the presence of internal stresses or framework flexure during seating of the definitive prosthesis.

Radiographic passive fit will be assessed by measuring marginal gaps at the prosthesis-abutment interface on standardized periapical radiographs. Radiographic gaps will be quantified and classified using predefined ordinal thresholds to identify clinically relevant discrepancies.

Each assessment will be scored using an ordinal scale (0-2) based on predefined clinical criteria. To reflect clinical decision-making and ensure a conservative interpretation of prosthetic fit, a hierarchical aggregation rule will be applied. For each assessment dimension, the prosthesis-level score will be defined as the worst score observed among all implant-prosthesis connections. The global multi-assessment passive fit score (0-2) will then be defined as the worst score observed across all assessment dimensions.

Accordingly, if any implant-prosthesis connection exhibits a clinically relevant misfit in any assessment, the definitive prosthesis will be classified as non-passive. This approach reflects the clinical principle that lack of passive fit at a single connection compromises the overall prosthetic outcome.

The primary objective of the study is to determine whether increased automation in digital workflows improves the passive fit of definitive full-arch implant-supported prostheses. Secondary objectives include comparing marginal, geometric, mechanical, and radiographic passivity outcomes among the three workflows and exploring the relationship between workflow automation and prosthetic adaptation accuracy.

The study follows CONSORT guidelines and aims to provide clinically relevant evidence to support the optimization of digital workflows in full-arch implant-supported prosthetic rehabilitation

Studietyp

Interventionell

Inskrivning (Beräknad)

30

Fas

  • Inte tillämpbar

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

    • Madrid
      • Madrid, Madrid, Spanien, 28040
        • Rekrytering
        • University Complutense of Madrid
        • Kontakt:
          • Miguel A Gomez-Polo, DDS, PhD, DDS
          • Telefonnummer: +34 91 394 2029
          • E-post: mgomezpo@ucm.es

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

18 år och äldre (Vuxen, Äldre vuxen)

Tar emot friska volontärer

Nej

Kön som är behöriga för studier

Allt

Beskrivning

Inclusion Criteria:

  1. Adults aged 18 years or older.
  2. Patients indicated for fixed implant-supported full-arch rehabilitation in the maxilla or mandible.
  3. Presence of clinically stable, osteointegrated titanium implants suitable for prosthetic rehabilitation.
  4. Absence of clinical or radiographic signs of peri-implant disease.
  5. Ability to understand the study procedures and provide written informed consent.
  6. Willingness and ability to attend all required clinical visits and evaluations.

Exclusion Criteria:

  1. Presence of peri-implant mucositis or peri-implantitis at the time of evaluation.
  2. Implant mobility or implant positioning that prevents proper prosthetic rehabilitation.
  3. Uncontrolled systemic medical conditions that could interfere with study participation or prosthetic treatment.
  4. Cognitive, psychological, or medical conditions that limit the ability to comply with study procedures.
  5. Inability to complete the required clinical evaluations or follow the study protocol, as judged by the investigator.

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Primärt syfte: Behandling
  • Tilldelning: Randomiserad
  • Interventionsmodell: Parallellt uppdrag
  • Maskning: Enda

Vapen och interventioner

Deltagargrupp / Arm
Intervention / Behandling
Aktiv komparator: CBCT-STL Alignment (Medicalfit 1.0)

Participants in this arm will receive a definitive, screw-retained, implant-supported full-arch prosthesis fabricated using a digital workflow based on manual alignment of CBCT and intraoral scan (STL) datasets. Scannable healing abutments (Tissue Shapers-IF; MedicalFit, Úbeda, Spain) will be connected to multi-unit abutments according to the manufacturer's instructions. An intraoral scan and a CBCT scan will be obtained and manually aligned by the operator during the CAD design stage using reference points visible in both datasets. Prosthetic framework passivation is performed manually during the CAD phase. The definitive prosthesis is designed and manufactured using CAD/CAM technology in monolithic zirconia (IPS e.max ZirCAD Prime; Ivoclar, Schaan, Liechtenstein). This workflow is fully operator-dependent and serves as an active comparator.

This procedure consists of a fully digital workflow for fabricating a screw-retained, implant-supported full-arch prosthesis using manual alignment between CBCT and intraoral scan (STL) data. Scannable healing abutments (Tissue Shapers-IF; MedicalFit, Úbeda, Spain) are attached to multi-unit abutments torqued to 10 Ncm. A CBCT scan and an intraoral scan are acquired and manually aligned in CAD software using operator-defined reference points. Passive fit is adjusted manually during the CAD design stage. The definitive framework is milled from monolithic zirconia (IPS e.max ZirCAD Prime; Ivoclar, Schaan, Liechtenstein). This workflow is operator-dependent and serves as the control procedure.

Aktiv komparator: Splint-Guided Alignment (Medicalfit 2.0)

Participants in this arm will receive a definitive, screw-retained, implant-supported full-arch prosthesis fabricated using a digital workflow that incorporates a rigid reference splint with metallic abutment cylinders to assist manual dataset alignment. Scannable healing abutments (Tissue Shapers-IF; MedicalFit, Úbeda, Spain) are connected to multi-unit abutments. A calibrated rigid splint is intraorally adapted and connected to the abutment cylinders, and both intraoral and splint scans are obtained. The acquired datasets are imported into CAD software, where CBCT-STL registration is performed manually using the splint as a reference. The definitive prosthetic framework is designed and manufactured using CAD/CAM technology in monolithic zirconia (IPS e.max ZirCAD Prime; Ivoclar, Schaan, Liechtenstein). This workflow remains operator-dependent and serves as an active comparator.

This procedure employs a rigid reference splint with metallic abutment cylinders to guide dataset alignment for full-arch prosthesis fabrication. Scannable healing abutments (Tissue Shapers-IF; MedicalFit, Úbeda, Spain) are connected to multi-unit abutments torqued to 10 Ncm. The splint is attached intraorally and scanned extraorally to capture implant positions. The splint and intraoral scans (STL) are imported into CAD software, where manual alignment is performed using the splint as a reference. The final monolithic zirconia framework (IPS e.max ZirCAD Prime; Ivoclar, Schaan, Liechtenstein) is designed and milled via CAD/CAM. This workflow is operator-dependent and used as an active comparator.

Experimentell: Automated AI-Assisted CBCT-STL Alignment (Medicalfit 3.0)

Participants in this arm will receive a definitive, screw-retained, implant-supported full-arch prosthesis fabricated using an automated AI-assisted digital workflow. Scannable healing abutments (Tissue Shapers-IF; MedicalFit, Úbeda, Spain) are connected to multi-unit abutments according to the manufacturer's protocol. An intraoral scan and a CBCT scan are acquired and imported into dedicated AI-assisted software (Pdental; MedicalFit, Úbeda, Spain). The software automatically detects the scan bodies and performs CBCT-STL dataset alignment prior to CAD/CAM prosthesis fabrication. The definitive prosthesis is designed and manufactured using CAD/CAM technology in monolithic zirconia (IPS e.max ZirCAD Prime; Ivoclar, Schaan, Liechtenstein). This workflow minimizes operator involvement and represents the experimental intervention.

This procedure uses AI-assisted software (Pdental, MedicalFit, Úbeda, Spain) to automate the alignment of CBCT and intraoral scan (STL) datasets for full-arch prosthesis fabrication. After placement of scannable healing abutments (Tissue Shapers-IF; MedicalFit), intraoral and CBCT scans are obtained. The software automatically detects healing abutments, performs CBCT-STL registration, and corrects deviations greater than 120 µm to achieve passive fit within clinically acceptable limits. The corrected digital model is exported in STL format for CAD design and CAM milling of a monolithic zirconia screw-retained prosthesis (IPS e.max ZirCAD Prime; Ivoclar, Schaan, Liechtenstein). This workflow minimizes operator dependency based on an AI-assisted automation. .

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Multi-assessment global passive fit score of definitive full-arch implant-supported prostheses
Tidsram: At definitive prosthesis placement

Passive fit of the definitive full-arch implant-supported prosthesis will be evaluated using a standardized multi-assessment clinical approach. Marginal, geometric (modified Sheffield test), mechanical (passive screw test sensitivity), and radiographic passive fit will be assessed independently at each implant-prosthesis connection.

Each assessment will be scored using an ordinal scale (0-2) based on predefined clinical criteria:

0 = no detectable discrepancy, indicating adequate passive fit;

1 = minor discrepancy, indicating a cl To obtain the prosthesis-level outcome, a conservative aggregation rule will be applied: for each assessment dimension, the prosthesis score will be defined as the worst score observed among all implant-prosthesis connections. The global multi-assessment passive fit score (0-2) will then be defined as the worst score observed across all assessment dimensions.

At definitive prosthesis placement

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Clinical marginal passive fit (probe assessment)
Tidsram: At definitive prosthesis placement

Marginal passive fit will be evaluated clinically by direct inspection using a calibrated probe to assess the presence of marginal discrepancies between the prosthesis and the transepithelial abutment. Outcomes will be recorded using an ordinal scale (0-2) based on visual and tactile criteria defined as follows:

0 = no visible gap and no probe penetration;

  1. = visible marginal discrepancy without probe penetration;
  2. = clinically detectable gap with probe penetration or visually, indicating lack of passive fit.
At definitive prosthesis placement
Radiographic marginal gap at the prosthesis-abutment interface
Tidsram: At definitive prosthesis placement

Marginal gaps between the definitive prosthesis and the transepithelial abutment will be evaluated on standardized periapical radiographs. Measured gaps will be classified using an ordinal scale (0-2) as follows:

0 = gap < 100 µm;

  1. = gap between 100 and 130 µm;
  2. = gap > 130 µm, indicating a clinically relevant misfit.
At definitive prosthesis placement
Geometric passive fit assessed by the modified Sheffield test
Tidsram: At definitive prosthesis placement

Geometric passive fit will be evaluated using the modified Sheffield test by tightening a single distal prosthetic screw and assessing the presence of lifting or separation at the non-tightened implant-prosthesis connections. Outcomes will be classified as follows:

0 = no visible or tactile separation;

  1. = slight visible separation without probe penetration;
  2. = clear separation with probe penetration, indicating absence of passive fit.
At definitive prosthesis placement
Mechanical passive fit assessed by passive screw test sensitivity
Tidsram: At definitive prosthesis placement

Mechanical passive fit will be evaluated qualitatively based on tactile resistance perceived during tightening of the prosthetic screws. Outcomes will be classified using an ordinal scale (0-2) defined as follows:

0 = smooth screw tightening without perceptible resistance;

  1. = mild or transient resistance without need for force;
  2. = clear resistance or framework flexure during tightening, indicating internal stress and lack of passive fit.
At definitive prosthesis placement

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Utredare

  • Studierektor: Miguel A Gómez Polo, DDS, PhD, School of Dentistry, Complutense University of Madrid. Pza Ramon y Cajal S/N. 28040 Madrid, Spain

Publikationer och användbara länkar

Den som ansvarar för att lägga in information om studien tillhandahåller frivilligt dessa publikationer. Dessa kan handla om allt som har med studien att göra.

Allmänna publikationer

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

1 juli 2026

Primärt slutförande (Beräknad)

1 december 2026

Avslutad studie (Beräknad)

1 mars 2027

Studieregistreringsdatum

Först inskickad

18 december 2025

Först inskickad som uppfyllde QC-kriterierna

18 december 2025

Första postat (Faktisk)

2 januari 2026

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

23 november 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

20 november 2026

Senast verifierad

1 november 2026

Mer information

Termer relaterade till denna studie

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

NEJ

IPD-planbeskrivning

Individual participant data will not be shared due to ethical and privacy considerations. The data collected include detailed clinical and radiographic information that could potentially allow re-identification of participants, even after de-identification. Data will be used solely for the purposes defined in the approved study protocol.

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .

Prenumerera