V-Sensors for Vitals: Assessment of the Accuracy of a Vital Sign Sensor Mounted on a Smartphone
An Open Label, Interventional, Clinical Performance and Validation Study to Demonstrate the Clinical Accuracy of of the V-Sensor in the Healthy Adult Population; Single Site in Switzerland
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
The five vital signs are blood pressure, pulse, respiration, SpO2, and temperature. The user tests two devices mounted on two smartphones.
- Temperature: Each user will use the V-sensor mounted on a smartphone to obtain his/her temperature interfaced with a temperature specific 'app'. Before and after the the temperature is obtained using the V-Sensor, the temperature will be obtained using an equivalent thermometer.
- Blood Pressure: Each user will test his/her blood pressure using the two sensors mounted on the two phones interfaced with a blood pressure specific 'app'. Before and after each measurement, a reference measurement using a mercury sphygmomanometer will be taken.
- Pulse, Respiration, SpO2: Each user will test his/her pulse, respiration, and SpO2 using 2 V-sensors mounted on two separate smartphones interfaced with a Pulse, Respiration, and SpO2 'app'. Before and after each measurement, a reference measurement using equivalent reference devices will be taken.
Study Type
Study Type
Enrollment (Anticipated)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Contact
Study Contact
- Name: Kelly A McClary, RN
- Phone Number: +41 (0)797764954
- Email: km@leman-micro.com
Study Contact Backup
- Name: Chris J Elliott, PhD
- Phone Number: +41(0)791733306
- Email: cje@leman-micro.com
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Informed Consent
- 18 and 80 years old
- no history or risk for developing lymphedema
- no arteriovenous fistula in the right arm
- no right limb amputation up to and including the right index finger
- ability to obtain 3 readings from the device during the screening/training session.
Exclusion Criteria:
- history of arrhythmia
- arteriovenous fistula in the right arm
- lymphedema or high risk for developing lymphedema in the right arm
- inability to follow the procedures of the study
- inability to hold the study device as instructed
- inability to obtain 3 readings from the study device during the screening session
- previous enrolment into the current study
- enrolment of the Investigator or Sponsor, his/her family members, employees and other dependent persons.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Other: V-Sensor Device User
This diagnostic medical device is designed to detect the user's 5 vital signs.
Each user tests two sensors, each sensor is mounted on a smartphone.
Each vital sign is compared to an equivalent reference device obtained by the healthcare provider.
|
Participants use device to detect the 5 vital signs (Blood Pressure, Pulse, Respiration, SpO2, Temperature).
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Blood Pressure accuracy to the reference device within the ranges specified by the ISO standard 81060-2
Time Frame: 4 months
|
For systolic and diastolic blood pressures, the mean error of determination of the individual paired determinations of the sphygmomanometer-under-test and the reference sphygmomanometer (Standby Baumanometer) for all participants shall not be greater than 8.0mmHg.
|
4 months
|
|
Pulse Rate accuracy to the reference device within the ranges specified by the ISO 80601-2-61
Time Frame: 4 months
|
Pulse rate accuracy shall be stated as the root-mean-square (rms) difference between paired pulse rate data recorded with the pulse oximeter equipment and with the Nellcor N-PA reference device.
|
4 months
|
|
Blood Oxygen accuracy to the reference device within the ranges specified by the ISO 80601-2-61.
Time Frame: 4 months
|
The SpO2 accuracy shall be stated as a root-mean-square (rms) difference of less than or equal to 4.0% SpO2 over the range of 70% to 100% SaO2
|
4 months
|
|
Respiration rate accuracy
Time Frame: 4 months
|
Respiration rate accuracy shall be stated as the root-mean-square (rms) difference between paired respiratory rate data with the Passport Respiration Rate Sensor PS-2133 reference method; target range +/- 10% of participant's respiratory rate
|
4 months
|
|
Temperature accuracy to the reference device with the ranges specified by the ISO 80601-2-56 & ATSM 1965 Standard.
Time Frame: 4 months
|
Temperature accuracy shall be stated as the root-mean-square (rms) difference between paired temperatures data recorded with the Exergen Temporal Scanner thermometer.
LMD's target is a clinical bias <= 3.0°C and a clinical repeatability <= 3.0°C.
|
4 months
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
No serious adverse device effects are reported.
Time Frame: 4 months
|
Zero serious adverse device effects are reported.
|
4 months
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Study Director: Chris J Elliott, PhD, Leman Micro-Device
Study record dates
Study Major Dates
Study Start (Anticipated)
Study Start
Primary Completion (Anticipated)
Primary Completion
Study Completion (Anticipated)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
Other Study ID Numbers
- LMD-CT-002
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.