Refitting Somofilcon A Sphere Contact Lens Wearers Into Fanfilcon A Sphere Lenses for 4-Weeks of Wear
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Mexico City, Mexico, 54090
- Optometry Clinic, National Autonomous University
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Is between 18 and 40 years of age (inclusive)
- Has had a self-reported visual exam in the last two years
- Is an adapted soft contact lens wearer
- Has a contact lens spherical prescription between - 0.25 to - 8.00 (inclusive)
- Have no less than -0.75D of astigmatism in both eyes.
- Can achieve best corrected spectacle distance visual acuity of 20/25 (0.10 logMAR) or better in each eye.
- Can achieve a distance visual acuity of 20/30 (0.18 logMAR) or better in each eye with the study contact lenses.
- Has clear corneas and no active ocular disease
- Has read, understood and signed the information consent letter.
- Patient contact lens refraction should fit within the available parameters of the study lenses.
- Is willing to comply with the wear schedule (at least 5 days per week, > 8 hours/day assuming there are no contraindications for doing so).
- Is willing to comply with the visit schedule
Exclusion Criteria:
- Has a CL prescription outside the range of the available parameters of the study lenses.
- Has a spectacle cylinder of ≥ 1.00D in either eye.
- Has a history of not achieving comfortable CL wear (5 days per week; > 8 hours/day)
- Has contact lens best corrected distance vision worse than 20/25 (0.10 logMAR) in either eye.
- Presence of clinically significant (grade 2-4) anterior segment abnormalities
- Presence of ocular or systemic disease or need of medications which might interfere with contact lens wear.
Slit lamp findings that would contraindicate contact lens wear such as:
- Pathological dry eye or associated findings
- Pterygium, pinguecula, or corneal scars within the visual axis
- Neovascularization > 0.75 mm in from of the limbus
- Giant papillary conjunctivitis (GCP) worse than grade 1
- Anterior uveitis or iritis (past or present)
- Seborrheic eczema, Seborrheic conjunctivitis
- History of corneal ulcers or fungal infections
- Poor personal hygiene
- Has a known history of corneal hypoesthesia (reduced corneal sensitivity)
- Has aphakia, keratoconus or a highly irregular cornea.
- Has Presbyopia or has dependence on spectacles for near work over the contact lenses.
- Has undergone corneal refractive surgery.
- Is participating in any other type of eye related clinical or research study.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: somofilcon A (habitual) lens, then fanfilcon A (test) lens
Participants are habitual wearers of somofilcon A lens and refitted with fanfilcon A lens.
|
Contact Lens
Other Names:
Contact Lens
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Lens Centration for Somofilcon A (Habitual) Lens
Time Frame: Baseline
|
Lens Centration was measured using a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
|
Baseline
|
|
Lens Centration for Somofilcon A (Habitual) Lens
Time Frame: 4- weeks
|
Lens Centration was measured using a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
|
4- weeks
|
|
Lens Centration for Fanfilcon A (Test) Lens
Time Frame: Baseline
|
Lens Centration was measured using a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
|
Baseline
|
|
Lens Centration for Fanfilcon A (Test) Lens
Time Frame: 2-weeks
|
Lens Centration was measured using a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
|
2-weeks
|
|
Lens Centration for Fanfilcon A (Test) Lens
Time Frame: 4- weeks
|
Lens Centration was measured using a 3-point scale(1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).
|
4- weeks
|
|
Lens Corneal Coverage for Somofilcon A (Habitual) Lens
Time Frame: Baseline
|
Lens Corneal Coverage was rated as either Yes or No (Yes = full corneal coverage all time / no= incomplete corneal coverage)
|
Baseline
|
|
Lens Corneal Coverage for Somofilcon A (Habitual) Lens
Time Frame: 4 -weeks
|
Lens Corneal Coverage was rated as either Yes or No (Yes = full corneal coverage all time / no= incomplete corneal coverage)
|
4 -weeks
|
|
Lens Corneal Coverage for Fanfilcon A (Test) Lens
Time Frame: Baseline
|
Lens Corneal Coverage was rated as either Yes or No (Yes = full corneal coverage all time / no= incomplete corneal coverage)
|
Baseline
|
|
Lens Corneal Coverage for Fanfilcon A (Test) Lens
Time Frame: 2-weeks
|
Lens Corneal Coverage was rated as either Yes or No (Yes = full corneal coverage all time / no= incomplete corneal coverage)
|
2-weeks
|
|
Lens Corneal Coverage for Fanfilcon A (Test) Lens
Time Frame: 4-weeks
|
Lens Corneal Coverage was rated as either Yes or No (Yes = full corneal coverage all time / no= incomplete corneal coverage)
|
4-weeks
|
|
Percentage of Lens Tightness for Somofilcon A (Habitual) Lens
Time Frame: Baseline
|
Lens Tightness: Push Up Test measured by 0% to 100% continuous scale (0%-falls from cornea without lid support, 50% = optimum, and 100% = no movement)
|
Baseline
|
|
Percentage of Lens Tightness for Somofilcon A (Habitual) Lens
Time Frame: 4-Weeks
|
Lens Tightness: Push Up Test measured by 0% to 100% continuous scale (0%-falls from cornea without lid support, 50% = optimum, and 100% = no movement)
|
4-Weeks
|
|
Percentage of Lens Tightness for Fanfilcon A (Test) Lens
Time Frame: Baseline
|
Lens Tightness: Push Up Test measured by 0% to 100% continuous scale (0%-falls from cornea without lid support, 50% = optimum, and 100% = no movement)
|
Baseline
|
|
Percentage of Lens Tightness for Fanfilcon A (Test) Lens
Time Frame: 2-Weeks
|
Lens Tightness: Push Up Test measured by 0% to 100% continuous scale (0%-falls from cornea without lid support, 50% = optimum, and 100% = no movement)
|
2-Weeks
|
|
Percentage of Lens Tightness for Fanfilcon A (Test) Lens
Time Frame: 4-Weeks
|
Lens Tightness: Push Up Test measured by 0% to 100% continuous scale (0%-falls from cornea without lid support, 50% = optimum, and 100% = no movement)
|
4-Weeks
|
|
Post-Blink Movement for Somofilcon A (Habitual) Lens
Time Frame: Baseline
|
Post-Blink movement was measured on a scale 0-4 (0=Insufficient, unacceptable movement, 4 = Excessive, unacceptable movement)
|
Baseline
|
|
Post-Blink Movement for Somofilcon A (Habitual) Lens
Time Frame: 4-weeks
|
Post-Blink movement was measured on a scale 0-4 (0=Insufficient, unacceptable movement, 4 = Excessive, unacceptable movement)
|
4-weeks
|
|
Post-Blink Movement for Fanfilcon A (Test) Lens
Time Frame: Baseline
|
Post-Blink movement was measured on a scale 0-4 (0=Insufficient, unacceptable movement, 4 = Excessive, unacceptable movement)
|
Baseline
|
|
Post-Blink Movement for Fanfilcon A (Test) Lens
Time Frame: 2-weeks
|
Post-Blink movement was measured on a scale 0-4 (0=Insufficient, unacceptable movement, 4 = Excessive, unacceptable movement)
|
2-weeks
|
|
Post-Blink Movement for Fanfilcon A (Test) Lens
Time Frame: 4-weeks
|
Post-Blink movement was measured on a scale 0-4 (0=Insufficient, unacceptable movement, 4 = Excessive, unacceptable movement)
|
4-weeks
|
|
Overall Lens Fit Acceptance for Somofilcon A (Habitual) Lens
Time Frame: Baseline
|
Overall fit acceptance was measured using a scale of 0-4 (0= should not be worn, 1 = borderline but unacceptable, 2 = min.
acceptable, early review, 3 = not perfect but OK to dispense, and 4 = perfect)
|
Baseline
|
|
Overall Lens Fit Acceptance for Somofilcon A (Habitual) Lens
Time Frame: 4-weeks
|
Overall fit acceptance was measured using a scale of 0-4 (0= should not be worn, 1 = borderline but unacceptable, 2 = min.
acceptable, early review, 3 = not perfect but OK to dispense, and 4 = perfect)
|
4-weeks
|
|
Overall Lens Fit Acceptance for Fanfilcon A (Test) Lens
Time Frame: Baseline
|
Overall fit acceptance was measured using a scale of 0-4 (0= should not be worn, 1 = borderline but unacceptable, 2 = min.
acceptable, early review, 3 = not perfect but OK to dispense, and 4 = perfect)
|
Baseline
|
|
Overall Lens Fit Acceptance for Fanfilcon A (Test) Lens
Time Frame: 2-weeks
|
Overall fit acceptance was measured using a scale of 0-4 (0= should not be worn, 1 = borderline but unacceptable, 2 = min.
acceptable, early review, 3 = not perfect but OK to dispense, and 4 = perfect)
|
2-weeks
|
|
Overall Lens Fit Acceptance for Fanfilcon A (Test) Lens
Time Frame: 4-weeks
|
Overall fit acceptance was measured using a scale of 0-4 (0= should not be worn, 1 = borderline but unacceptable, 2 = min.
acceptable, early review, 3 = not perfect but OK to dispense, and 4 = perfect)
|
4-weeks
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Average Daily Wearing Time - Somofilcon A (Habitual) Lens
Time Frame: 4-weeks
|
Average Daily Wearing Time for somofilcon A (Habitual) lens - measured by time hours/day
|
4-weeks
|
|
Average Daily Wearing Time - Fanfilcon A (Test) Lens
Time Frame: 2-weeks
|
Average daily wearing time for fanfilcon A (test) lens was measured in hours/day
|
2-weeks
|
|
Average Daily Wearing Time - Fanfilcon A (Test) Lens
Time Frame: 4-weeks
|
Average daily wearing time for fanfilcon A (test) lens was measured in hours/day
|
4-weeks
|
|
Average Comfortable Wearing Time - Somofilcon A (Habitual) Lens
Time Frame: 4 weeks
|
Average Comfortable Wearing Time - somofilcon A (habitual) Lens measured in hours/day.
Assessed using typical time of day when subject first experiences lens awareness or irritation.
|
4 weeks
|
|
Average Comfortable Wearing Time - Fanfilcon A (Test) Lens
Time Frame: 2 weeks
|
Average Comfortable Wearing Time - fanfilcon A (test) Lens measured in hours/day.
Assessed using typical time of day when subject first experiences lens awareness or irritation.
|
2 weeks
|
|
Average Comfortable Wearing Time - Fanfilcon A (Test) Lens
Time Frame: 4 weeks
|
Average Comfortable Wearing Time - fanfilcon A (test) Lens measured in hours/day.
Assessed using typical time of day when subject first experiences lens awareness or irritation.
|
4 weeks
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Ruben Velázquez Guerrero, MSc., FIACLE, School of Optometry, National Autonomous University (UNAM), Mexico City
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Actual)
Primary Completion
Study Completion (Actual)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- EX-MKTG-101
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.