Efficacy of Celsi Warmer for the Management of Hypothermic Newborns at the Lagos University Teaching Hospital in Lagos, Nigeria
Efficacy of Low-cost Warming Mattress Celsi Warmer for the Management of Hypothermic Newborns at the Lagos University Teaching Hospital in Lagos, Nigeria
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Study Type
Study Type
Enrollment (Estimated)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Contact
Study Contact
- Name: Rebecca Massey, MPH, BSN, RN
- Phone Number: 7133482923
- Email: rebecca.elias@rice.edu
Study Contact Backup
- Name: Sonia Sosa Saenz, BME
- Email: soniasosa@rice.edu
Study Locations
-
-
-
Lagos, Nigeria
- Recruiting
- University of Lagos, Lagos, Nigeria
-
Contact:
- Chinyere Ezeaka, MBBS
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
- Child
Accepts Healthy Volunteers
Description
Inclusion Criteria:
Any neonate who:
- Is currently being treated at study location,
- Is an inborn admission to the neonatal ward,
- Whose parents or guardians provided informed consent,
- Whose parents or guardians providing informed consent are 18 years old or older,
- Has a current weight of greater than or equal to 1.0 kg and less than or equal to 4.0 kg,
- Has been identified as in need of thermal care defined as having a moderate to severe hypothermic temperature (32.0-36.0 °C) as the last temperature recorded in hospital chart, or during recruitment procedures; and
- is unable to be enrolled in Kangaroo Mother Care (KMC) for reasons including, but not limited to: a. Mother/guardian unable or unavailable to provide KMC b. Under observation in the Neonatal Intensive Care Unit (NICU) before transfer to KMC c. No space in KMC d. Clinician's discretion;
- May be receiving other medical treatments, including but not limited to, Continuous Positive Airway Pressure (CPAP), oxygen therapy, intravenous (IV) fluids, management and monitoring of common newborn conditions such as hypoglycemia, and/or hyperbilirubinemia.
Exclusion Criteria:
- Requires mechanical ventilation;
- Is deemed in need of intensive care by the hospital staff who is providing care, including but not limited to: a. neonates with severe anemia and/or any suspected hematological disorders, and/or b. Neonates with obvious congenital anomalies, and/or c. neonates suspected with hypo/hyperthyroidism or any hormonal disorders;
- has been diagnosed with birth asphyxia;
- Presents a condition that precludes the use of the temperature sensor and/or abdominal belt including but not limited to, known umbilical cord infection, known skin infection;
- Whose clinician presents concerns about their participation;
- Is receiving treatment that participation in the study would interfere with (e.g. transferring to KMC).
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
Experimental: Hypothermic Neonates (90)
A trained study clinician will attach the Celsi Warmer temperature sensor to the abdomen and secure it with the abdominal belt according to the device's instruction manual (such that the temperature sensor is positioned using a vertical line from the mid-clavicular line and just below the ribs).
The time of sensor placement will be recorded.
|
A trained study clinician will attach the Celsi Warmer temperature sensor to the abdomen and secure it with the abdominal belt according to the device's instruction manual (such that the temperature sensor is positioned using a vertical line from the mid-clavicular line and just below the ribs).
The time of sensor placement will be recorded.
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Broad Objective - Comparison to the Gold Standard of Rewarming Hypothermic Newborns
Time Frame: 1 year
|
To evaluate the clinical efficacy and safety of the Celsi Warmer in rewarming hypothermic newborns by comparing the Celsi Warmer's accuracy in measuring neonatal temperature compared to the gold standard, axillary temperature readings.
|
1 year
|
Secondary Outcome Measures
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Efficacy of the Device
Time Frame: 1 year
|
To assess the performance of the Celsi Warmer in rewarming hypothermic newborns by measuring the incidence of attainment of normothermia (36.5-37.5°C)
|
1 year
|
|
Safety of the Device - Incidence of Hyperthermia
Time Frame: 1 year
|
To evaluate any incidence of hyperthermia (>37.5°C)
during the thermoregulation intervention.
|
1 year
|
|
Safety of the Device - Rate of Temperature Increase
Time Frame: 1 year
|
To measure the infant's rate of temperature (°C) increase through thermoregulatory intervention.
|
1 year
|
|
Safety of the Device - Local effect of the Abdominal Belt
Time Frame: 1 year
|
To evaluate the local effect of the abdominal belt on newborn's skin by looking at any evidence of skin irritation
|
1 year
|
|
Safety of the Device - Incidence of Rebound Hypothermia
Time Frame: 1 year
|
To report any incidence of rebound hypothermia ( < 36.5°C)
(post-intervention hypothermia) up to 72 hours from the initial encounter
|
1 year
|
Collaborators and Investigators
Sponsor
Sponsor
Collaborators
Collaborators
Investigators
Investigators
- Principal Investigator: Chinyere Ezeaka, MBBS, MPH, College of Medicine, University of Lagos, Lagos, Nigeria
Publications and helpful links
General Publications
- Kumar V, Shearer JC, Kumar A, Darmstadt GL. Neonatal hypothermia in low resource settings: a review. J Perinatol. 2009 Jun;29(6):401-12. doi: 10.1038/jp.2008.233. Epub 2009 Jan 22.
- Lunze K, Bloom DE, Jamison DT, Hamer DH. The global burden of neonatal hypothermia: systematic review of a major challenge for newborn survival. BMC Med. 2013 Jan 31;11:24. doi: 10.1186/1741-7015-11-24.
- Carns J, Kawaza K, Quinn MK, Miao Y, Guerra R, Molyneux E, Oden M, Richards-Kortum R. Impact of hypothermia on implementation of CPAP for neonatal respiratory distress syndrome in a low-resource setting. PLoS One. 2018 Mar 15;13(3):e0194144. doi: 10.1371/journal.pone.0194144. eCollection 2018.
- Cavallin F, Calgaro S, Brugnolaro V, Seni AHA, Muhelo AR, Da Dalt L, Putoto G, Trevisanuto D. Impact of temperature change from admission to day one on neonatal mortality in a low-resource setting. BMC Pregnancy Childbirth. 2020 Oct 23;20(1):646. doi: 10.1186/s12884-020-03343-7.
- Mullany LC. Neonatal hypothermia in low-resource settings. Semin Perinatol. 2010 Dec;34(6):426-33. doi: 10.1053/j.semperi.2010.09.007.
- Ogunlesi TA, Ogunfowora OB, Ogundeyi MM. Prevalence and risk factors for hypothermia on admission in Nigerian babies <72 h of age. J Perinat Med. 2009;37(2):180-4. doi: 10.1515/JPM.2009.014.
Study record dates
Study Major Dates
Study Start (Actual)
Study Start
Primary Completion (Estimated)
Primary Completion
Study Completion (Estimated)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (Actual)
First Posted
Study Record Updates
Last Update Posted (Actual)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- FY2023-69
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.