- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT01699763
Evaluation of Blood Glucose Monitoring Systems
January 29, 2016 updated by: Ascensia Diabetes Care
Evaluation of Blood Glucose Meter Systems - Contour® NEXT LINK Study
The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and two additional Blood Glucose Monitoring Systems (BGMS) from other companies.
All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study Overview
Status
Completed
Conditions
Intervention / Treatment
Detailed Description
The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and two additional Blood Glucose Monitoring Systems (BGMS) from other companies.
All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Performance of the three systems was evaluated across the glucose range of the BGMSs using capillary blood.
All testing and lancing was performed by study staff and some samples were tested from subject fingertips.
Additionally, some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels.
Study Type
Interventional
Enrollment (Actual)
112
Phase
- Not Applicable
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Locations
-
-
Indiana
-
Mishawaka, Indiana, United States, 46544
- Bayer HealthCare LLC, Diabetes Care
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
18 years and older (Adult, Older Adult)
Accepts Healthy Volunteers
No
Genders Eligible for Study
All
Description
Inclusion Criteria:
- Males and females, 18 years of age or older
- Willing to complete all study procedures
Exclusion Criteria:
- Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
- Hemophilia or any other bleeding disorder
- Pregnancy
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: Subjects with and without Diabetes
All testing and lancing were performed by the study staff; Subjects with and without Diabetes did not perform any lancing or self-testing in this study.
Study Staff lanced the fingers of subjects and tested the blood samples using three Blood Glucose Monitoring Systems (BGMS): Contour® NEXT LINK BGMS; OneTouch® UltraLink® BGMS; Nova Max Link® BGMS.
|
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with and without diabetes (up to 10% of subjects without diabetes were included).
Performance of Contour® NEXT LINK BGMS was evaluated across the glucose range of the BGMSs.
All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with and without diabetes (up to 10% of subjects without diabetes were included).
Performance of OneTouch® UltraLink® BGMS was evaluated across the glucose range of the BGMSs.
All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with and without diabetes (up to 10% of subjects without diabetes were included).
Performance of Nova Max Link® BGMS was evaluated across the glucose range of the BGMSs.
All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
Time Frame: 8 hours
|
Using the overall Blood Glucose (BG) range (34 to 561 mg/dL according to YSI subject plasma results), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI plasma results (BG reference) were compared.
MARD is calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%).
Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention.
Lower MARD value indicates smaller difference between meter value and the reference value.
Higher MARD value indicates larger difference between meter value and the reference value.
|
8 hours
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range (<=80 mg/dL)
Time Frame: 8 hours
|
Using fresh and glycolyzed samples with YSI plasma Blood Glucose (BG) ≤80 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI subject plasma results (BG reference) were compared.
MARD is calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%).
Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention.
Lower MARD value indicates smaller difference between meter value and the reference value.
Higher MARD value indicates larger difference between meter value and the reference value.
|
8 hours
|
|
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL)
Time Frame: 8 hours
|
Using samples with YSI plasma Blood Glucose (BG) >180 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI subject plasma results (BG reference) were compared.
MARD is calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%).
Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention.
Lower MARD value indicates smaller difference between meter value and the reference value.
Higher MARD value indicates larger difference between meter value and the reference value.
|
8 hours
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start
October 1, 2012
Primary Completion (Actual)
October 1, 2012
Study Completion (Actual)
October 1, 2012
Study Registration Dates
First Submitted
October 2, 2012
First Submitted That Met QC Criteria
October 2, 2012
First Posted (Estimate)
October 4, 2012
Study Record Updates
Last Update Posted (Estimate)
February 29, 2016
Last Update Submitted That Met QC Criteria
January 29, 2016
Last Verified
January 1, 2016
More Information
Terms related to this study
Other Study ID Numbers
- CTD-2012-007-01
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Diabetes
-
University of Colorado, DenverMassachusetts General Hospital; Beta Bionics, Inc.CompletedDiabetes Mellitus, Type 1 | Type 1 Diabetes | Diabetes type1 | Type 1 Diabetes Mellitus | Autoimmune Diabetes | Diabetes Mellitus, Insulin-Dependent | Juvenile-Onset Diabetes | Diabetes, Autoimmune | Insulin-Dependent Diabetes Mellitus 1 | Diabetes Mellitus, Insulin-Dependent, 1 | Diabetes Mellitus, Brittle | Diabetes Mellitus, Juvenile-Onset and other conditionsUnited States
-
University of North Carolina, Chapel HillAmerican Diabetes AssociationNot yet recruitingType 2 Diabetes Mellitus (T2DM) | Diabetes (DM) | Insulin Dependent Diabetes | Type 1 Diabetes (T1D) | Diabetes Education | Diabetes Care | Diabetes (Insulin-requiring, Type 1 or Type 2)United States
-
University of North Carolina, Chapel HillAmerican Heart AssociationRecruitingType 2 Diabetes | Nutrition | Diabetes Type 2 | T2DM (Type 2 Diabetes Mellitus) | Diabetes Mellitis | T2DM | Diabetes EducationUnited States
-
University of Colorado, DenverMassachusetts General Hospital; Ann & Robert H Lurie Children's Hospital of... and other collaboratorsRecruitingDiabetes Mellitus | Diabetes | Type 2 Diabetes | Diabetes Mellitus Type 2 | Diabetes Mellitus, Type I | Diabetes Mellitus Type II | Diabetes Mellitus, Insulin-Dependent | Diabetes, Autoimmune | Type 1 Diabetes (T1D) | Diabetes Type 2 on Insulin | Diabetes, Type IIUnited States
-
Guang NingRecruitingType 2 Diabetes Mellitus | Type1 Diabetes Mellitus | Monogenetic Diabetes | Pancreatogenic Diabetes | Drug-Induced Diabetes Mellitus | Other Forms of Diabetes MellitusChina
-
Northern Care Alliance NHS Foundation TrustBrighter ABCompletedDiabetes type1 | Diabetes type2United Kingdom
-
University of Trás-os-Montes and Alto DouroCompletedType 2 Diabetes Mellitus | Diabetes-Related ComplicationsPortugal
-
DiappyMedSanoiaRecruitingDiabetes Type 1 | Pancreatogenous Diabetes | Diabetes Type 2 on InsulinFrance
-
VeraLight, Inc.InLight SolutionsUnknownGestational Diabetes | Insulin Dependent Diabetes | Non Insulin Dependent DiabetesUnited States
-
COUR Pharmaceutical Development Company, Inc.RecruitingType 1 Diabetes | Type 1 Diabetes Mellitus | T1DM | T1D | Type 1 Diabetes in Adolescence | Type 1 Diabetes in Children | Type 1 Diabetes Patients | Type 1 Diabetes Mellitis | T1DM - Type 1 Diabetes Mellitus | Type 1 Diabetes (Juvenile Onset)United States
Clinical Trials on Contour® NEXT LINK BGMS
-
Ascensia Diabetes CareCompletedDiabetesUnited States
-
Ascensia Diabetes CareCompleted
-
Ascensia Diabetes CareCompletedDiabetes MellitusUnited States
-
Ascensia Diabetes CareCompletedDiabetes MellitusAustralia
-
Ascensia Diabetes CareRegulatory and Clinical Research Institute IncCompletedDiabetes MellitusUnited States
-
Ascensia Diabetes CareCompleted
-
Ascensia Diabetes CareCompletedDiabetesUnited States
-
Ascensia Diabetes CareCompletedDiabetesUnited States