Comparison Evaluation of Blood Glucose Meter Systems

January 29, 2016 updated by: Ascensia Diabetes Care

Evaluation of Blood Glucose Meter Systems - NACT Study

The purpose of this study was to evaluate the performance of one investigational Bayer Blood Glucose meter (with investigational Bayer test strips) and four Blood Glucose Monitoring Systems from other companies. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems was evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some blood samples were glycolyzed to lower the glucose levels.

Study Overview

Study Type

Interventional

Enrollment (Actual)

146

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Indiana
      • Mishawaka, Indiana, United States, 46544
        • Bayer HealthCare LLC, Diabetes Care

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

Yes

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Males and females
  • 18 years of age or older
  • Willing to complete all study procedures

Exclusion Criteria:

  • Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
  • Hemophilia or any other bleeding disorder
  • Pregnancy

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: DIAGNOSTIC
  • Allocation: NA
  • Interventional Model: SINGLE_GROUP
  • Masking: NONE

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
EXPERIMENTAL: Study Staff Test BGMSs
All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS.
Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of CONTOUR® NEXT EZ BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of FreeStyle Freedom Lite® BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of ACCU-CHEK® Aviva BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of TRUEtrack® BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of OneTouch® Ultra®2 BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
Time Frame: 10 hours
Using the overall Blood Glucose(BG) range (24 to 386mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all |(BG meter)-(BG reference)| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.
10 hours

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL)
Time Frame: 10 hours
Using fresh and glycolyzed samples with Blood Glucose(BG) below 70 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all |(BG meter)-(BG reference)| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.
10 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

March 1, 2012

Primary Completion (ACTUAL)

March 1, 2012

Study Completion (ACTUAL)

March 1, 2012

Study Registration Dates

First Submitted

March 6, 2012

First Submitted That Met QC Criteria

March 6, 2012

First Posted (ESTIMATE)

March 8, 2012

Study Record Updates

Last Update Posted (ESTIMATE)

February 29, 2016

Last Update Submitted That Met QC Criteria

January 29, 2016

Last Verified

January 1, 2016

More Information

Terms related to this study

Other Study ID Numbers

  • CTD-PRO-2012-001-01

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.