Measuring Thoracic Impedance in Hemodialysis Patients With the u-Cor Monitoring System (MaTcH)

October 9, 2020 updated by: Zoll Medical Corporation

Measuring Thoracic Impedance in Hemodialysis Patients With the µ-Cor System A Pre-Market Validation Study (MaTcH)

The purpose of this study is to measure the fluid changes in the body using a device called the µ-Cor System, an investigational device. The µ-Cor System will record the fluid changes in the body, as well as ECG (electrocardiogram or an electrical tracing of your heart rhythm) heart rate, breathing rate, posture and activity at regular intervals.

The information collected by the µ-Cor System will then be compared to the actual fluid removed through dialysis and to the information collected by an FDA (Food and Drug Administration) cleared device called ZOE (NonInvasive Medical Technologies), a monitor which also measures the fluid changes in the body.

The objectives of this study are to document any differences in measurement of thoracic impedance obtained from the µ-Cor System and the ZOE system. Thoracic impedance is a measure of the electrical activity in the chest that varies with changes in body size and composition, fluid volume, breathing status and other variables.

Measurements of the amount of fluid removed during dialysis will also be compared between the µ-Cor System, the ZOE system and the actual fluid removed.

The ability of the µ-Cor System to measure thoracic impedance will be tested by placing the device randomly at one of two locations:

  • Study Arm 1: side location- below left axilla
  • Study Arm 2: front location - upper left pectoral area

Both the µ-Cor System and the ZOE System will be worn simultaneously during one dialysis session. The ZOE system device is placed on the chest- 2 points: 1 at the base of your neck and 1 in your mid chest.

Study Overview

Status

Completed

Detailed Description

The µ-Cor System is intended to record, store, transmit, and display the following physiological data to medical professionals: thoracic impedance, ECG, heart rate, respiration rate, activity and posture.

The µ-Cor System is indicated for patients who are 21 years of age or older who:

  • have fluid-management problems,
  • are taking diuretic medication,
  • are living with heart failure,
  • are living with end-stage renal disease,
  • are recovering from a coronary artery disease-related event, and/or
  • are suffering from recurrent dehydration.

Objectives:

  • This clinical trial is intended to provide evidence of substantial equivalence between the µ-Cor System and the ZOE System in the ability to measure thoracic impedance, by comparing the correlation between µ Cor measurement and ultrafiltration volume (UFV) as compared with the correlation between ZOE measurement and UFV. The actual UFV changes will be used to arbitrate any differences.
  • The ability of the µ-Cor System to measure thoracic impedance will be demonstrated at 2 body locations: side location (below left axilla, Study Arm 1), and front location (upper left pectoral area, Study Arm 2).

Subject population:

Forty patients undergoing hemodialysis (n = 40) in 2 study arms with at least 50% enrollment of patients diagnosed with congestive heart failure (CHF) in each arm:

  • Study Arm 1: 20 patients undergoing hemodialysis, with at least 10 of these patients having CHF, will have the µ-Cor device placed on the side location
  • Study Arm 2: 20 patients undergoing hemodialysis, with at least 10 of these patients having CHF, will have the µ-Cor device placed on the front location

Patients will wear the µ-Cor device and will be connected to the ZOE monitor during one hemodialysis session in the clinic. Patients will be randomly allocated to Study Arm 1 or Study Arm 2, with randomization stratified by CHF status (CHF or non-CHF). During the hemodialysis session, the patient will wear 1 µ-Cor device and will be connected simultaneously to 1 ZOE monitor (via 2 ZOE electrodes) for comparative measurements and UFV correlation, from at least 15 minutes before the start of dialysis to at least 15 minutes after the end of the dialysis session. The µ-Cor will record measurements during the dialysis session, including at least the 15 minutes before and 15 minutes after the session. ZOE (Z0) values will be measured every 6 minutes (± 1 minute) during the dialysis session. ZOE (Z0) values also will be measured every 3 minutes (± 1 minute) before and after the session, for a minimum of 5 measurements. The UFV during the course of hemodialysis will be measured by automated readings provided by the dialysis machine every 6 minutes (± 1 minute). 10. Any IV infusions administered, oral fluid or solid intake, and urine output during the course of hemodialysis will be recorded. Weight, blood pressure, and heart rate will be recorded before and after the dialysis session.The patient's involvement in the study ends after the completion of all procedures planned for the hemodialysis session.

Study Type

Interventional

Enrollment (Actual)

40

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Colorado
      • Lakewood, Colorado, United States, 80228
        • DaVita Clinical Research
    • Florida
      • Orlando, Florida, United States, 32809
        • Orlando Clinical Research Center
    • Minnesota
      • Minneapolis, Minnesota, United States, 55404
        • DaVita Clinical Research

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

21 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Men and women at least 21 years of age.
  • Is currently scheduled to undergo hemodialysis 3 times per week in a clinic setting and has been on this regimen for at least 3 months.
  • For those patients with CHF: were diagnosed with CHF by a qualified provider and show symptomatic signs of New York Heart Association (NYHA) Class II to IV at enrollment.
  • Is prescribed a net fluid removal of at least 2.5 L during the hemodialysis session.
  • Is willing and able to sign informed consent in English.

Exclusion Criteria:

  • Is a female patient with a known pregnancy or is unsure of pregnancy status.
  • Has known allergies or skin sensitivities to electrode hydrogel and/or acrylic based adhesive.
  • Has skin breakdown in areas where device and electrode placement is required.
  • Was hospitalized within the 2 weeks prior to enrollment.
  • Had intradialytic hypotension requiring administration of intravenous (IV) fluids of ≥250 mL or that resulted in a referral to urgent care, within 2 weeks prior to enrollment.
  • Had myocardial infarction, acute coronary syndrome, or stroke within 4 weeks prior to enrollment.
  • Has active nephrotic syndrome
  • Has severe malnutrition, as diagnosed per a qualified provider.
  • Is participating in another clinical trial.
  • Has an implanted device that might interfere with the µ-Cor.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Other: study arm 1 - below left axilla
20 patients undergoing hemodialysis, with at least 10 of these patients having congestive heart failure, will have the u-Cor system placed below left axilla and the ZOE Fluid Status Monitor
Monitoring System
Other Names:
  • u-Cor V3.0 System
Monitoring System
Other: study arm 2-upper left pectoral area
20 patients undergoing hemodialysis, with at least 10 of these patients having congestive heart failure, will have the u-Cor system placed on the upper left pectoral area and the ZOE Fluid Status Monitor
Monitoring System
Other Names:
  • u-Cor V3.0 System
Monitoring System

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Difference in Correlation of Thoracic Impedance With Ultra-filtration Volume (UFV) Between uCor and ZOE
Time Frame: 15 minutes prior the start of one hemodialysis session to 15 minutes post completion of one hemodialysis session, approximately 2 to 6 hours.
The primary endpoint measurement is the difference of 1) the Pearson correlation of thoracic impedance made by the u-Cor System to UFV and 2) the Pearson correlation of thoracic impedance made by the ZOE to UFV. Differences in correlation coefficients were calculated for each subject and each study arm.
15 minutes prior the start of one hemodialysis session to 15 minutes post completion of one hemodialysis session, approximately 2 to 6 hours.

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Study Director: Madhuri Bhat, MS, Zoll Medical Corporation

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

November 3, 2016

Primary Completion (Actual)

February 1, 2017

Study Completion (Actual)

February 1, 2017

Study Registration Dates

First Submitted

March 1, 2017

First Submitted That Met QC Criteria

March 1, 2017

First Posted (Actual)

March 7, 2017

Study Record Updates

Last Update Posted (Actual)

November 4, 2020

Last Update Submitted That Met QC Criteria

October 9, 2020

Last Verified

October 1, 2020

More Information

Terms related to this study

Other Study ID Numbers

  • 2016-KM-004

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Fluid Retention Tissue

Clinical Trials on u-Cor System

Subscribe