A Pivotal Study to Evaluate Guardian™ Connect CGM System Performance in China

August 24, 2020 updated by: Medtronic Diabetes
The purpose of this study is to demonstrate the performance and safety of the Guardian™ Sensor (3) over 170 hours (7 days) when inserted in the abdomen used in subjects age 14 - 75 years with the Guardian™ Connect System.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Detailed Description

This study is a multi-center, prospective single-arm design without controls. Up to 72 subjects will be enrolled in order to have approximately 60 subjects complete the study. Three investigational centers in China will be used during this study.

During the study, each subject will be randomly assigned to one day of the Yellow Springs Instrument (YSI™*) frequent sample testing (FST) (Day 1, 3-5, or 7).Subjects will wear two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days (one which will be paired to the Guardian Connect app and the other will function as a glucose recorder.).

The Guardian Sensor (3)s will be worn in the abdomen area and self-inserted by the subject on same side or opposite sides.

On the evening prior to FST, subjects will be asked to fast for approximately 12 hours and adjust their insulin and medications according to routine care (for example as they would do for fasting lipid panel). Subjects may fast for shorter period of time based on investigator discretion.

The subject should be in fasting status upon arrival hospital to start FST process. The feeding protocol may be modified based on investigator discretion. The duration of FST will be approximately 7 hours.

During the study, subjects will continue with their current diabetes regimen independent of the study devices. Subjects will be instructed by the investigational center that they are not to use the study devices (except for the CONTOUR™* study meter) for the management of their diabetes. The CONTOUR™* study meter may be used for treatment decisions and calibration of Guardian™ Sensor (3).

Study Type

Interventional

Enrollment (Actual)

72

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Beijing, China, 100700
        • the Seventh Medical Center of PLA General Hospital
      • Hangzhou, China, 310016
        • Sir Run Run Shaw Hospital
      • Shanghai, China, 200233
        • Shanghai Sixth People's Hospital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

14 years to 75 years (Child, Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  1. Subject is 14 - 75 years of age at time of screening
  2. Subject has a clinical diagnosis of type 1 or 2 diabetes as determined via medical record or source documentation by an individual qualified to make a medical diagnosis
  3. Subject has adequate venous access as assessed by investigator or appropriate staff
  4. Subject is willing to follow the study procedures and willing to come to study visits

Exclusion Criteria:

  1. Subject will not tolerate tape adhesive in the area of Guardian™ Sensor (3) placement as assessed by qualified individual.
  2. Subject has any unresolved adverse skin condition in the area of study device or device placement (e.g., psoriasis, rash, Staphylococcus infection)
  3. Subject is actively participating in an investigational study (drug or device) wherein they have received treatment from an investigational study (drug or device) in the last 2 weeks
  4. Subject is female and has a positive pregnancy screening test
  5. Females of child bearing age and who are sexually active should be excluded if they are not using a form of contraception deemed reliable by investigator
  6. Subject is female and plans to become pregnant during the course of the study
  7. Subject has a hematocrit (Hct) lower than the normal reference range
  8. Subject may not be on the research staff of those performing this study

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Study arm
Use two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days and undergo one Yellow Springs Instrument (YSI™) frequent sample testing (FST) (Day 1, 3-5, or 7).
Use two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percentage of Guardian™ Sensor (3) Values That Are Within 20% Agreement of YSI
Time Frame: 168 Hours
Percentage of Guardian™ Sensor (3) Values that are within 20% agreement of Yellow Springs Instrument (YSI) plasma glucose values (±20 mg/dL (1.1 mmol/L) when reference blood glucose (YSI) less than or equal to (≤) 80 mg/dL (4.4 mmol/L) ) during YSI frequent sampling testing (FST) days.
168 Hours
Consensus Error Grid Analysis of Mean Rate of Paired Sensor Values and YSI Values in Zone A+B
Time Frame: 168 Hours
A mean rate in Zone A+B of Consensus Error Grid between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. Consensus Error Grid (or Parkes error grid) compared the paired primary sensor and YSI reference glucose values. Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action", Zone B as "altered clinical action with little or no effect on clinical outcome". Ideal situation is 100% in Zone A + B.
168 Hours
Clarke Error Grid Analysis of Mean Rate of Paired Sensor Values and YSI Values in Zone A+B
Time Frame: 168 Hours
A mean rate in Zone A+B of Clarke Error Grid between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. Clarke Error Grid compared the paired primary sensor and YSI reference glucose values. Zone A are "clinically accurate in that they would lead to clinically correct treatment decisions". Zone B "would lead to benign or no treatment". Ideal situation is 100% in Zone A + B.
168 Hours
Mean Absolute Relative Difference (MARD)
Time Frame: 168 Hours
Mean absolute relative difference (MARD) between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple FST days were pooled together for reporting purpose.
168 Hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Jian Zhou, MD, Shanghai 6th People's Hospital
  • Principal Investigator: Xiaofeng Lv, MD, Seventh Medical Center of PLA Army General Hospital
  • Principal Investigator: Hong Li, MD, Sir Run Run Shaw Hospital

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

February 28, 2019

Primary Completion (Actual)

July 25, 2019

Study Completion (Actual)

July 25, 2019

Study Registration Dates

First Submitted

October 15, 2018

First Submitted That Met QC Criteria

October 15, 2018

First Posted (Actual)

October 17, 2018

Study Record Updates

Last Update Posted (Actual)

September 9, 2020

Last Update Submitted That Met QC Criteria

August 24, 2020

Last Verified

August 1, 2020

More Information

Terms related to this study

Other Study ID Numbers

  • CIP320

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.