A Pivotal Study to Evaluate Guardian™ Connect CGM System Performance in China
Study Overview
Detailed Description
This study is a multi-center, prospective single-arm design without controls. Up to 72 subjects will be enrolled in order to have approximately 60 subjects complete the study. Three investigational centers in China will be used during this study.
During the study, each subject will be randomly assigned to one day of the Yellow Springs Instrument (YSI™*) frequent sample testing (FST) (Day 1, 3-5, or 7).Subjects will wear two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days (one which will be paired to the Guardian Connect app and the other will function as a glucose recorder.).
The Guardian Sensor (3)s will be worn in the abdomen area and self-inserted by the subject on same side or opposite sides.
On the evening prior to FST, subjects will be asked to fast for approximately 12 hours and adjust their insulin and medications according to routine care (for example as they would do for fasting lipid panel). Subjects may fast for shorter period of time based on investigator discretion.
The subject should be in fasting status upon arrival hospital to start FST process. The feeding protocol may be modified based on investigator discretion. The duration of FST will be approximately 7 hours.
During the study, subjects will continue with their current diabetes regimen independent of the study devices. Subjects will be instructed by the investigational center that they are not to use the study devices (except for the CONTOUR™* study meter) for the management of their diabetes. The CONTOUR™* study meter may be used for treatment decisions and calibration of Guardian™ Sensor (3).
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Beijing, China, 100700
- the Seventh Medical Center of PLA General Hospital
-
Hangzhou, China, 310016
- Sir Run Run Shaw Hospital
-
Shanghai, China, 200233
- Shanghai Sixth People's Hospital
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Subject is 14 - 75 years of age at time of screening
- Subject has a clinical diagnosis of type 1 or 2 diabetes as determined via medical record or source documentation by an individual qualified to make a medical diagnosis
- Subject has adequate venous access as assessed by investigator or appropriate staff
- Subject is willing to follow the study procedures and willing to come to study visits
Exclusion Criteria:
- Subject will not tolerate tape adhesive in the area of Guardian™ Sensor (3) placement as assessed by qualified individual.
- Subject has any unresolved adverse skin condition in the area of study device or device placement (e.g., psoriasis, rash, Staphylococcus infection)
- Subject is actively participating in an investigational study (drug or device) wherein they have received treatment from an investigational study (drug or device) in the last 2 weeks
- Subject is female and has a positive pregnancy screening test
- Females of child bearing age and who are sexually active should be excluded if they are not using a form of contraception deemed reliable by investigator
- Subject is female and plans to become pregnant during the course of the study
- Subject has a hematocrit (Hct) lower than the normal reference range
- Subject may not be on the research staff of those performing this study
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
---|---|
Experimental: Study arm
Use two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days and undergo one Yellow Springs Instrument (YSI™) frequent sample testing (FST) (Day 1, 3-5, or 7).
|
Use two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
Percentage of Guardian™ Sensor (3) Values That Are Within 20% Agreement of YSI
Time Frame: 168 Hours
|
Percentage of Guardian™ Sensor (3) Values that are within 20% agreement of Yellow Springs Instrument (YSI) plasma glucose values (±20 mg/dL (1.1 mmol/L) when reference blood glucose (YSI) less than or equal to (≤) 80 mg/dL (4.4 mmol/L) ) during YSI frequent sampling testing (FST) days.
|
168 Hours
|
Consensus Error Grid Analysis of Mean Rate of Paired Sensor Values and YSI Values in Zone A+B
Time Frame: 168 Hours
|
A mean rate in Zone A+B of Consensus Error Grid between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated.
Consensus Error Grid (or Parkes error grid) compared the paired primary sensor and YSI reference glucose values.
Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action", Zone B as "altered clinical action with little or no effect on clinical outcome".
Ideal situation is 100% in Zone A + B.
|
168 Hours
|
Clarke Error Grid Analysis of Mean Rate of Paired Sensor Values and YSI Values in Zone A+B
Time Frame: 168 Hours
|
A mean rate in Zone A+B of Clarke Error Grid between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated.
Clarke Error Grid compared the paired primary sensor and YSI reference glucose values.
Zone A are "clinically accurate in that they would lead to clinically correct treatment decisions".
Zone B "would lead to benign or no treatment".
Ideal situation is 100% in Zone A + B.
|
168 Hours
|
Mean Absolute Relative Difference (MARD)
Time Frame: 168 Hours
|
Mean absolute relative difference (MARD) between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated.
MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100).
Therefore, the unit of MARD is percentage (%).
Note that results from multiple FST days were pooled together for reporting purpose.
|
168 Hours
|
Collaborators and Investigators
Sponsor
Investigators
- Principal Investigator: Jian Zhou, MD, Shanghai 6th People's Hospital
- Principal Investigator: Xiaofeng Lv, MD, Seventh Medical Center of PLA Army General Hospital
- Principal Investigator: Hong Li, MD, Sir Run Run Shaw Hospital
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
- CIP320
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Diabetes Mellitus
-
University of Colorado, DenverMassachusetts General Hospital; Beta Bionics, Inc.CompletedDiabetes Mellitus, Type 1 | Type 1 Diabetes | Diabetes type1 | Type 1 Diabetes Mellitus | Autoimmune Diabetes | Diabetes Mellitus, Insulin-Dependent | Juvenile-Onset Diabetes | Diabetes, Autoimmune | Insulin-Dependent Diabetes Mellitus 1 | Diabetes Mellitus, Insulin-Dependent, 1 | Diabetes Mellitus, Brittle | Diabetes Mellitus, Juvenile-Onset and other conditionsUnited States
-
Guang NingRecruitingType 2 Diabetes Mellitus | Type1 Diabetes Mellitus | Monogenetic Diabetes | Pancreatogenic Diabetes | Drug-Induced Diabetes Mellitus | Other Forms of Diabetes MellitusChina
-
Meir Medical CenterCompletedDiabetes Mellitus Type 2 | Diabetes Mellitus, Non-insulin Dependant | Diabetes Mellitus, on Oral Hypoglycemic Treatment | Adult Type Diabetes MellitusIsrael
-
Medical College of WisconsinMedical University of South CarolinaCompletedDiabetes Mellitus | Type 2 Diabetes Mellitus | Adult-Onset Diabetes Mellitus | Non-Insulin-Dependent Diabetes Mellitus | Noninsulin Dependent Diabetes Mellitus, Type IIUnited States
-
Hanmi Pharmaceutical Company LimitedUnknownType2 Diabetes Mellitus | Type1 Diabetes MellitusUnited States
-
Peking Union Medical College HospitalUnknownType 2 Diabetes Mellitus | Type 1 Diabetes Mellitus | Gestational Diabetes Mellitus | Pancreatogenic Diabetes Mellitus | Pregestational Diabetes Mellitus | Diabetes Patients in Perioperative PeriodChina
-
SanofiCompletedType 1 Diabetes Mellitus-Type 2 Diabetes MellitusHungary, Russian Federation, Germany, Poland, Japan, United States, Finland
-
Joslin Diabetes CenterCambridge Medical Technologies, LLCCompletedType 2 Diabetes Mellitus | Type1 Diabetes MellitusUnited States
-
Medical College of WisconsinMedical University of South Carolina; National Institute of Diabetes and Digestive...Active, not recruitingDiabetes Mellitus, Type 2 | Diabetes Mellitus, Type II | Diabetes Mellitus, Adult-Onset | Diabetes Mellitus, Non-Insulin-Dependent | Diabetes Mellitus, Noninsulin DependentUnited States
-
Medical College of WisconsinNational Institute of Diabetes and Digestive and Kidney Diseases (NIDDK)CompletedDiabetes Mellitus, Type 2 | Diabetes Mellitus, Type II | Diabetes Mellitus, Adult-Onset | Diabetes Mellitus, Non-Insulin-Dependent | Diabetes Mellitus, Noninsulin DependentUnited States
Clinical Trials on Guardian™ Sensor (3)
-
Medtronic DiabetesCompletedType 1 Diabetes Mellitus | Type2 Diabetes MellitusUnited States
-
Medtronic DiabetesCompletedType 2 Diabetes Mellitus | Type 1 Diabetes MellitusUnited States
-
Medtronic DiabetesCompletedType 1 Diabetes | Type 2 DiabetesUnited States
-
University of AlbertaCompleted
-
Postgraduate Institute of Medical Education and...RecruitingType1 Diabetes MellitusIndia
-
Vastra Gotaland RegionCompletedDiabetes Mellitus, Type 1 | Blood GlucoseSweden
-
HealthPartners InstituteNational Institute of Diabetes and Digestive and Kidney Diseases (NIDDK); Yale... and other collaboratorsCompletedType 1 Diabetes MellitusUnited States, Germany, Slovenia, Israel
-
University of OklahomaCompletedWeight Gain | Feeding Disorder Neonatal | Very Low Birth Weight InfantUnited States
-
University Hospital Inselspital, BerneCompletedLeft Ventricular Assist DeviceSwitzerland
-
Medtronic DiabetesWithdrawn