Assessment of Visual Function With a Portable Brain-computer Interface

February 26, 2020 updated by: NGoggle
  1. The purpose of this study is to evaluate the nGoggle's accuracy and repeatability in detecting visual function loss. In addition, the ability to stage glaucomatous damage and investigate the relationship between nGoggle metrics and neural damage in glaucoma will also be evaluated.
  2. Longitudinal study, including 200 patients with: glaucoma, suspected of having glaucoma, nonglaucomatous optic neuropathies, AMD, retinal degenerations, other diseases involving the visual pathways, besides healthy controls. Subjects will perform standard ophthalmological exams, and the following research tests: electroencephalogram, visual evoked potentials, and questionnaires.
  3. Statistical analyses will be performed by the PI using the software Stata, MATLAB, and MPLUS. Risks are low, consisting of some discomfort, fatigue, dizziness or motion sickness.

Study Overview

Status

Enrolling by invitation

Intervention / Treatment

Detailed Description

Background & Significance

Assessment of functional loss in glaucoma has been traditionally made with standard automated perimetry (SAP). Despite the significant progress achieved with the development of SAP, assessment of visual field loss still has significant drawbacks. SAP testing is limited by subjectivity of patient responses and high test-retest variability, frequently requiring many tests for effective detection of change over time. Moreover, as these tests are generally conducted in clinic-based settings, limited patient availability and health care resources often result in an insufficient number of tests acquired over time, with delayed diagnosis and detection of disease progression. The requirement for highly trained technicians, cost, complexity, and lack of portability of SAP also preclude its use for screening of visual field loss in underserved populations. Furthermore, current SAP testing as performed in clinical practice frequently results in delayed detection of progression, with potential late initiation/escalation of treatment and irreversible functional damage.

As a part of a funded Phase I project, the nGoggle (nGoggle Inc., San Diego, CA), the first device of its kind, allowed a portable and objective brain-based assessment of visual function deficits in glaucoma. The nGoggle integrates a wearable, wireless, no-prep EEG system and a head-mounted display (HMD) to allow users to routinely monitor electrical brain activities associated with visual field stimulation. A recent study comparing the diagnostic accuracies of an earlier version of the nGoggle and SAP in discriminating patients with glaucoma from healthy subjects found that the nGoggle performed at least as well, if not better than, SAP with the additional advantage of portability and objectivity.

The overall goal of this proposal is to test the nGoggle's accuracy and repeatability in detecting visual function loss. In addition, its ability to stage glaucomatous damage, by discriminating patients at different stages of the disease will be evaluated. Finally, the relationship between nGoggle metrics and neural damage in glaucoma as detected by other objective techniques such as optical coherence tomography will also be investigated. Patients with other diseases affecting the visual system will also be included in order to assess whether the nGoggle can differentiate glaucoma from them.

Study Design & Procedures

This will be a longitudinal study in which patients will be examined approximately biannually. The following patients will be recruited:

  1. Glaucoma;
  2. Suspicion of having glaucoma;
  3. Non-glaucomatous optic neuropathies;
  4. Age-related macular degeneration (AMD);
  5. Retinal degenerations;
  6. Other diseases involving the visual pathways such as optic neuritis, tumors or ischemic neuropathy.
  7. Healthy subjects

General evaluation:

As part of the study all participants will undergo the following general evaluation:

  1. Review of medical history;
  2. Comprehensive ophthalmologic examination
  3. Questionnaires: a. Cognitive function using the Montreal Cognitive Assessment (MoCA); b. A short structured interview to obtain vision-related quality of life information using the National Eye Institute 25-Item Visual Functioning Questionnaire (NEI VFQ-25); c. A questionnaire to assess patient's satisfaction with the nGoggle testing, including easiness of placement, comfort, and user-interface.
  4. Electroencephalogram (EEG) and Visual Evoked Potentials (VEP): Visual evoked potentials (VEP) will be evaluated with Diopsys NOVA (Diopsys, Inc.), which is a well-established conventional multifocal VEP device to assess brain functioning following visual stimuli from a computer screen. Subjects will also be monitored with an EEG system integrated with a head-mounted display (nGoggle, Inc.). The nGoggle is a portable visual evoked potential system that uses an EEG to acquire visual evoked potentials in response to visual stimuli presented on a virtual reality goggle.

Risk/Benefit Assessment

Risks:

Risks from participating in the study are very low. The primary risks for all participants are some discomforts, which are similar to those encountered in any complete eye examination. Participants will undergo non-invasive procedures used for diagnostic purposes that are used in standard clinical care. Eventually, it will be necessary to use numbing or dilating drops for fundus examination, as is done in standard clinical care. Some risks of those drops include: dry eye, mild headache, blurred vision, drowsiness, burning eyes, tearing, light sensitivity, redness, and allergy reactions. The only non-FDA approved device used in this study is the nGoggle (nGoggle Inc.), which is an investigational diagnostic device qualified as a non-significant risk device. The subject may experience mild discomfort due to drying of the eye during the study. Should a patient note drying, a drop of artificial tears will be placed over the eye by the study key personnel. In very rare cases, artificial eye tear drops may cause irritation or allergy. There are no known risks associated with the EEG test other than discomfort and fatigue. Subjects might become bored, fatigued or distressed while participating. To minimize these potential effects, the tasks will be kept short and subjects will be instructed that they may request to interrupt the test anytime. All assessments, software, instruments that are not FDA approved will be used as tools to assess disability and device efficacy, and will not be used as a diagnostic procedure without confirmation of the diagnosis by another, medically established diagnostic product or procedure. There is, however, the potential risk of loss of confidentiality. Every effort will be made to keep the information confidential; however, this cannot be guaranteed. Some of the questions that will be asked as part of this study may make the patients feel uncomfortable. Patients may refuse to answer any of the questions and may take a break at any time during the study. Patient may stop their participation in this study at any time.

Unknown/unforeseeable risks

In addition to the risks and discomforts listed here, there may be other risks that are currently not known. Subjects will be informed if any other potential risk becomes known through the duration of the study. Also, the risks and discomforts may occur more often or be more severe than have been seen before and written in this form.

Benefits:

Subjects will be referred from the Duke Eye Center care providers after a comprehensive clinical eye examination from a routine visit. If the last clinical eye examination is not within 6 months of the study visit, the participant will be examined by the PI, with no additional cost to the patient.

Participants will also undergo tests that will allow the validation and development of a transformative device for diagnosing and monitoring ocular diseases, and could be suited for home-based testing of visual function in the future, which could benefit patients with eye diseases.

Study Type

Observational

Enrollment (Anticipated)

200

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • North Carolina
      • Durham, North Carolina, United States, 27710
        • Duke University Eye Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

14 years to 86 years (Adult, Older Adult)

Accepts Healthy Volunteers

N/A

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

The following patients will be recruited from Dr. Tseng's clinic by verbal communication. Staff members at the Duke Eye Center will also be invited to recommend potential participants for the study, in conformance with the Duke Human Subjects Committee Guidelines.

  1. Glaucoma;
  2. Suspicion of having glaucoma;
  3. Non-glaucomatous optic neuropathies;
  4. Age-related macular degeneration (AMD);
  5. Retinal degenerations;
  6. Other diseases involving the visual pathways such as optic neuritis, tumors or ischemic neuropathy.
  7. Healthy subjects

Description

Inclusion Criteria:

  1. Subjects must be between the ages of 18 and 90 years old;
  2. Both males and females will be included.
  3. Be able and willing to provide signed informed consent and follow study instructions

Exclusion Criteria:

1. Subjects will be excluded if they present with any systemic or ocular conditions that in the opinion of the Principal Investigator may prevent them from completing the tests (e.g. history of seizures or other coexistent ocular pathologies).

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Cohort
  • Time Perspectives: Prospective

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Glaucoma and glaucoma suspects
Patients with at least two consecutive and reliable standard automated perimetry (SAP) examinations with either a pattern standard deviation (PSD) outside the 95% normal limits or a glaucoma hemifield test (GHT) result outside the 99% normal limits. Patients considered suspects for glaucoma must have an intraocular pressure (IOP) greater than 21 millimeters of mercury (mmHg) or suspicious appearance of the optic nerve head but with reliable normal visual fields, defined as a PSD within 95% confidence limits and a GHT result within normal limits.
The nGoggle is a portable visual evoked potential system that uses an EEG to acquire visual evoked potentials in response to visual stimuli presented on a virtual reality goggle. The investigators will assess the ability of the nGoggle to detect visual field loss in patients with glaucoma, as well as to discriminate among the different stages of the disease, its repeatability and reliability for patient self-testing and patient satisfaction compared to standard automated perimetry (SAP), optical coherence tomography (OCT) metrics and Diopsys NOVA (Diopsys, Inc.). In addition, patients with other diseases affecting the visual system will also be tested in order to assess whether the nGoggle can differentiate glaucoma from them.
Other Names:
  • Portable brain-computer interface
Age-related macular degeneration

Patients will be considered as having AMD if one or more of the following are present on posterior biomicroscopy (fundoscopy), indirect ophthalmoscopy or Optical Coherence Tomography (OCT) exams:

  1. Presence of at least intermediate-size drusen (63µm or larger in diameter)
  2. Retinal pigment epithelium (RPE) abnormalities such as hypopigmentation or hyperpigmentation
  3. Reticular pseudodrusen (also called sub retinal drusenoid deposit)
  4. Presence of any of the following features: geographic atrophy of the RPE, choroidal neovascularization (exudative, wet), polypoidal choroidal vasculopathy, or retinal angiomatous proliferation.
The nGoggle is a portable visual evoked potential system that uses an EEG to acquire visual evoked potentials in response to visual stimuli presented on a virtual reality goggle. The investigators will assess the ability of the nGoggle to detect visual field loss in patients with glaucoma, as well as to discriminate among the different stages of the disease, its repeatability and reliability for patient self-testing and patient satisfaction compared to standard automated perimetry (SAP), optical coherence tomography (OCT) metrics and Diopsys NOVA (Diopsys, Inc.). In addition, patients with other diseases affecting the visual system will also be tested in order to assess whether the nGoggle can differentiate glaucoma from them.
Other Names:
  • Portable brain-computer interface
Other eye diseases
Patients with other retinal degenerations such as retinitis pigmentosa or with other diseases affecting the visual pathways such as tumors, ischemic neuropathy or optic neuritis may also be included. Their diagnosis will be extracted from their clinical visits.
The nGoggle is a portable visual evoked potential system that uses an EEG to acquire visual evoked potentials in response to visual stimuli presented on a virtual reality goggle. The investigators will assess the ability of the nGoggle to detect visual field loss in patients with glaucoma, as well as to discriminate among the different stages of the disease, its repeatability and reliability for patient self-testing and patient satisfaction compared to standard automated perimetry (SAP), optical coherence tomography (OCT) metrics and Diopsys NOVA (Diopsys, Inc.). In addition, patients with other diseases affecting the visual system will also be tested in order to assess whether the nGoggle can differentiate glaucoma from them.
Other Names:
  • Portable brain-computer interface
Healthy subjects
To be considered healthy, subjects have to have IOP < 22 mmHg with no history of elevated IOP and with at least two reliable normal visual fields, defined as a PSD within 95% confidence limits and a GHT result within normal limits.
The nGoggle is a portable visual evoked potential system that uses an EEG to acquire visual evoked potentials in response to visual stimuli presented on a virtual reality goggle. The investigators will assess the ability of the nGoggle to detect visual field loss in patients with glaucoma, as well as to discriminate among the different stages of the disease, its repeatability and reliability for patient self-testing and patient satisfaction compared to standard automated perimetry (SAP), optical coherence tomography (OCT) metrics and Diopsys NOVA (Diopsys, Inc.). In addition, patients with other diseases affecting the visual system will also be tested in order to assess whether the nGoggle can differentiate glaucoma from them.
Other Names:
  • Portable brain-computer interface

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Device accuracy to detect visual field loss
Time Frame: from date of enrollment and every 6 months, up to 5 years
Comparison of diagnostic accuracies will be performed by analyzing areas under the receiver operating characteristic (ROC) curves for each nGoggle and SAP metric by assessment of partial areas under the ROC curves (targeting high specificity>90% region), sensitivities at matched specificities and diagnostic likelihood ratios (LRs). The investigators will also evaluate the correlation between nGoggle multifocal steady-state visual evoked potentials (mfSSVEP) metrics and SAP parameters throughout the disease spectrum.
from date of enrollment and every 6 months, up to 5 years
Device accuracy to detect structural neural loss
Time Frame: from date of enrollment and every 6 months, up to 5 years
Comparison of diagnostic accuracies will be performed by analyzing areas under the receiver operating characteristic (ROC) curves for each nGoggle and OCT parameters, as well as by assessment of partial areas under the ROC curves (targeting high specificity>90% region), sensitivities at matched specificities and diagnostic likelihood ratios (LRs). The investigators will also evaluate the correlation between nGoggle mfSSVEP metrics and SDOCT parameters throughout the disease spectrum. SDOCT testing will be performed with Spectralis OCT (Heidelberg Engineering, Germany). SDOCT parameters to be investigated will include retinal nerve fiber layer (RNFL) thickness, minimum rim width (MRW) and macular thickness.
from date of enrollment and every 6 months, up to 5 years
Device accuracy to detect objective functional loss
Time Frame: from date of enrollment and every 6 months, up to 5 years
Comparison of diagnostic accuracies will be performed by analyzing areas under the receiver operating characteristic (ROC) curves for each nGoggle and Diopsys NOVA parameters, as well as by assessment of partial areas under the ROC curves (targeting high specificity>90% region), sensitivities at matched specificities and diagnostic likelihood ratios (LRs). The investigators will also evaluate the correlation between nGoggle mfSSVEP metrics and those parameters throughout the disease spectrum. SDOCT testing will be performed with Spectralis OCT (Heidelberg Engineering, Germany). SDOCT parameters to be investigated will include retinal nerve fiber layer (RNFL) thickness, minimum rim width (MRW) and macular thickness.
from date of enrollment and every 6 months, up to 5 years
Assessment of repeatability
Time Frame: 6 weeks
Assessment of repeatability of the device parameters will be evaluated through intra-class correlation coefficient (ICC). A subgroup of subjects will undergo 5 visits over a short period of time (6 weeks). A sample size of 35 subjects will provide a width of 0.1 for the 95% for ICC, allowing assessment of repeatability of the devices parameters.
6 weeks

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Patients' satisfaction
Time Frame: from date of enrollment and every 6 months, up to 5 years
The investigators will obtain a survey regarding patients' satisfaction with testing, including easiness of placement, comfort, and user-interface.
from date of enrollment and every 6 months, up to 5 years

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Henry Tseng, PhD, Duke University

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

February 11, 2019

Primary Completion (Anticipated)

January 1, 2024

Study Completion (Anticipated)

January 1, 2024

Study Registration Dates

First Submitted

November 27, 2018

First Submitted That Met QC Criteria

November 28, 2018

First Posted (Actual)

November 30, 2018

Study Record Updates

Last Update Posted (Actual)

February 27, 2020

Last Update Submitted That Met QC Criteria

February 26, 2020

Last Verified

February 1, 2020

More Information

Terms related to this study

Other Study ID Numbers

  • Pro00100894
  • R42EY027651 (U.S. NIH Grant/Contract)

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

Undecided

IPD Plan Description

Undecided.

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.