Vector Engineering Clinical

Engineering Evaluation of the Vector NIV Device in Chronic Obstructive Pulmonary Disease (COPD) Patients

This study aimed to validate a novel automatic non-invasive ventilation (NIV) mode that continuously adjusts expiratory positive airway pressure (EPAP) to the lowest value that abolishes tidal expiratory flow limitation (EFL). The investigator conducted a prospective, non-randomized, study on stable chronic obstructive pulmonary disease (COPD) patients that may or may not be treated currently with NIV. Patients were studied in a sleep lab on a single night with the auto - titrating EPAP that adjusts to abolish tidal EFL. The primary endpoint was to evaluate the behavior of the EPAP during the night. Additionally, a sub-group of patients used the device at home for a 2 week period. EPAP behavior was assessed during this 2 week period.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Detailed Description

Tidal Expiratory Flow Limitation (EFLT) is the inability to increase expiratory flow despite increasing effort. Especially in patients with mild to severe COPD, EFL can present challenges when receiving mechanical ventilation. To overcome EFLT expiratory positive airway pressure (EPAP) is applied; however, a single level or fixed EPAP may not overcome the airflow obstruction. The current study was undertaken to explore the variability of EFLT determined using Forced Oscillation Technique (FOT) to dynamically measure lung reactance (∆Xrs) and to evaluate the ability of automatically adjusted EPAP (PEEPopt) to over come EFLT overnight and over a two week period.

In this prospective non-randomized trial, an unreleased noninvasive ventilator set in an S/T mode applied continuous oscillations (5 Hz, 1 cmH2O amplitude, 2 cmH20 peak to peak). Response to the oscillations was analyzed to calculate ∆Xrs and EPAP was adjusted automatically between 4 and 20 cmH2O. A fixed pressure support of 6 cmH2O was delivered. ∆Xrs was measured sitting and supine. Participants with evidence of EFLT (∆Xrs > 2.8) were asked to undergo overnight polysomnography (PSG) and then a sub-group of patients used the ventilator at home for two weeks. EFLT within and between participants was analyzed. EFLT behavior during PSG, the response of the device to dynamically abolish EFLT, as well as the impact of this therapy on sleep quality was studied. Objective ventilator adherence data were used to determine usage after the 2-week in-home use. Additionally, experienced NIV participants qualitatively rated therapy comfort compared to their current treatment.

Study Type

Interventional

Enrollment (Actual)

42

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Ohio
      • Youngstown, Ohio, United States, 44512
        • Pulmonary Rehabilitation Associates
    • Pennsylvania
      • Clairton, Pennsylvania, United States, 15025
        • Jefferson Associates and Internal Medicine

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

40 years to 80 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  1. Age > 40 years of age; < 80 years of age
  2. Ability to provide consent
  3. Diagnosis of COPD
  4. Must present with EFL via screening of the vector device at 3 cmH2O
  5. Have an EPAP to abolish EFL greater or equal to 6cmH2O
  6. Must be able to maintain SpO2 greater than 88% at rest and during EPAP Titration
  7. Participants who completed the initial study and who would be willing to use the Vector NIV device at home during the night for a 2 week period OR
  8. Participants prescribed and currently using a PAP or NIV device at home who meet study inclusion/exclusion criteria of primary protocol

Exclusion Criteria:

  1. Any major non COPD uncontrolled disease or condition, such as congestive heart failure, malignancy, liver or renal insufficiency (that requires current evaluation for liver or renal transplantation or dialysis), amyotrophic lateral sclerosis, or severe stroke, or other condition as deemed appropriate by investigator as determined by review of medical history and / or participant reported medical history
  2. Suffering from a COPD exacerbation at the time of data collection or in the 30 days prior to data collection
  3. Self-reported Pregnancy
  4. Employee or family member that is affiliated with Philips Respironics
  5. Currently employed by a manufacturer of respiratory products or family member employed by a manufacturer of respiratory products
  6. History of bullous emphysema
  7. History of pneumothorax
  8. Evidence of acute sinusitis or otitis media
  9. Hypotension
  10. Participants at risk for aspiration of gastric contents
  11. Epistaxis
  12. Participants in respiratory failure
  13. Inability to maintain a patent airway or adequately clear secretions
  14. Participants currently using a PAP or NIV device at home with a documented EPAP setting on their current device that is greater than then the mean or final EPAP determined during the therapy session of the screening visit.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Auto-titrating EPAP
ExpiraFlowTM technology- Non-invasive ventilator that auto titrates EPAP to abolish Expiratory flow limitation.
In the Auto-titrating EPAP mode, the device automatically adjusted EPAP to the minimum level able to abolish tidal expiratory flow limitation.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Average EPAP During Overnight PSG
Time Frame: Approximately 8 hours
The expiratory positive airway pressure (EPAP) is the parameter that is automatically adjusted. The ventilator automatically titrates EPAP to the minimum value able to abolish expiratory flow limitation.
Approximately 8 hours
Difference Between Mean Inspiratory and Expiratory Reactance During Overnight PSG
Time Frame: Approximately 8 hours
The difference between mean inspiratory and expiratory reactance is measured using the forced oscillation technique. The difference between mean inspiratory and expiratory reactance is an index of tidal expiratory flow limitation. Values > 2.8 cmH2O*s/L are indicative of tidal expiratory flow limitation.
Approximately 8 hours
Average EPAP for 2 Week Device Take Home Studies
Time Frame: Approximately 8 hours nightly use over 14 days
The expiratory positive airway pressure (EPAP) is the parameter that is automatically adjusted. The ventilator automatically titrates EPAP to the minimum value able to abolish expiratory flow limitation.
Approximately 8 hours nightly use over 14 days
Difference Between Mean Inspiratory and Expiatory Reactance During 2 Week Device Take Home Studies
Time Frame: 2 weeks
The difference between mean inspiratory and expiatory reactance is measured using the forced oscillation technique. The difference between mean inspiratory and expiratory reactance is an index of tidal expiratory flow limitation. Values > 2.8 cmH2O*s/L are indicative of tidal expiratory flow limitation.
2 weeks

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Sleep Quality
Time Frame: Approximately 8 hours
Sleep quality is a key outcome during overnight studies. Nocturnal non-invasive ventilation may affect sleep quality in two ways: 1) it may improve it by reducing sleep-related respiratory problems; 2) it may worsen it if the ventilation mode causes discomfort to the patient. Sleep quality was evaluated by full laboratory polysomnography. Participants after there overnight answered a question of comfort and they answered as 1- being very uncomfortable 2- uncomfortable 3- neutral 4- comfortable 5- very comfortable.
Approximately 8 hours
2 Week Home Use Device Compliance
Time Frame: Nightly use for 14 days
Device usage every day, for 14 days based on data downloaded from the device.
Nightly use for 14 days
Therapy Comfort Survey
Time Frame: End of 2 weeks device take home study

Prior NIV users rated their comfort with the Forced Oscillation Technique (FOT) therapy over their prior non-invasive ventilation (NIV) therapy. Participants were asked the question "how comfortable did the pressure therapy feel compared to your current device" and they answered the questions on a scale of 0 to 5.

5-Very Comfortable 4-Comfortable 3-Neutral 2-Uncomfortable

1-Very Uncomfortable 0-Not applicable

End of 2 weeks device take home study

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

October 4, 2018

Primary Completion (Actual)

March 3, 2020

Study Completion (Actual)

March 3, 2020

Study Registration Dates

First Submitted

January 13, 2021

First Submitted That Met QC Criteria

January 25, 2021

First Posted (Actual)

January 26, 2021

Study Record Updates

Last Update Posted (Actual)

May 26, 2021

Last Update Submitted That Met QC Criteria

May 25, 2021

Last Verified

May 1, 2021

More Information

Terms related to this study

Other Study ID Numbers

  • SRC-HRC-VectorENG -2018-10241

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.