Evaluation of Leak-free Bronchoscope Adapter

August 30, 2023 updated by: Beth Hochman, Columbia University

Evaluation of Leak-free Bronchoscope Adapter to Limit Ventilated Air Volume Loss and Pathogen Aerosolization From Intubated Patients Who Receive Bronchoscopy

Patients who develop respiratory failure require the assistance of mechanical ventilation while recovering from their respiratory injury. For example, patients that are affected by COVID-19 can develop such severe respiratory failure that they are unable to breathe on their own and therefore require the assistance of a mechanical ventilator. A Bronchoscope is a tool that can be used to visualize the inside of the lungs to aid in further diagnosis and delivery of therapeutic agents in patients with respiratory disease or failure. However, bronchoscopy can cause air to leak from the patient's lungs into the room due to a poorly designed adapter that is required to facilitate the introduction of the bronchoscope into the lungs. This leak can lead to the patient losing oxygen during the procedure which can be harmful to the patient and delay or limit recovery. Additionally, the air that leaks can lead to the spread of bacteria/viruses into the ambient air, which can contaminate the intensive care unit and expose other patients and healthcare workers to harmful pathogens. This was highlighted most recently during the COVID-19 pandemic where national guidelines limited the use of bronchoscopy in patients affected by the virus due to concerns about spreading the virus within the hospital and to the healthcare workers.

In this study, the investigators will evaluate the efficacy of a novel adapter that avoids air leakage during bronchoscopy. Patients with respiratory failure on mechanical ventilation who require bronchoscopy will be randomized into two groups (groups A and B). These patients will be identified by the intensive care unit physician. Informed consent will be obtained from the patient's health care proxy. Patients in group A will undergo bronchoscopy using the standard commercially available adapter. Patients in group B will undergo bronchoscopy using the newly developed bronchoscope adapter. The efficacy of the bronchoscopy adapter will be evaluated by analyzing ventilator-associated parameters as well as the level of air contamination after the procedure. A bronchoscope adapter that limits air leak would have significant value in improving patient outcomes who receive bronchoscopy. Additionally, this would allow bronchoscopy to be more readily performed in patients with potential bacterial/viral pathogen spread.

Primary objective: Determination of equivalency or superiority of the new "leak-free adapter" compared to the standard commercial adapter to enable safe bronchoscopy in ventilated patients requiring bronchoscopy.

Secondary objective: To determine if the adapter provides improved ventilation during bronchoscopy by limiting tidal volume loss and pathogen aerosolization.

Study Overview

Detailed Description

Patients who develop acute respiratory failure require hospitalization and mechanical ventilatory support. Of all patients admitted to intensive care units in the US, approximately 40% require mechanical ventilation. With the ongoing COVID-19 pandemic, approximately 18% of affected patients are expected to require mechanical ventilation.

Mechanical ventilation requires endotracheal intubation (cannulation of the patient's trachea) to deliver ventilated air to the lungs. Air is delivered by the ventilator to the patient using positive pressure in a closed circuit, ensuring adequate tidal volume to the patient without air escape to the outside environment. Nevertheless, patients placed on mechanical ventilation often develop superimposed infections known as "Ventilator-Associated Pneumonia" (VAP) as well as significant mucus secretions which limit effective mechanical ventilation. A camera tipped device is used to enable more effective bronchoscopy to visualize and explore the tracheobronchial tree for both diagnostic (bacterial/viral sampling) and therapeutic (suctioning obstructive mucus secretions, delivery of therapeutic agents) purposes.

Despite clear diagnostic and therapeutic benefits, bronchoscopy is also associated with substantial morbidity and complications. To perform bronchoscopy, a standard (commercially available) adapter allows the bronchoscope to enter the closed ventilation circuit, providing direct access to the patient's airways. However, the standard adapter leads to a significant leak of tidal volume outside of the ventilation circuit to the ambient environment. This air leak from the adapter leads to two major complications: First, the loss of positive pressure ventilation to patients means that oxygen is not being delivered to the patient and therefore leads to hypoxemia, which limits the duration and efficacy of bronchoscopy. Second, escaped air from the ventilator circuit can lead to the generation of aerosols, which can lead to the significant spread of viral and bacterial infectious agents to the ambient environment and surrounding healthcare providers. Intensive care units (ICUs) in the US must employ negative pressure ventilation to limit pathogen spread outside of the room and to the rest of the ICU. ICU rooms equipped with standard room negative pressure ventilation require 46 minutes after bronchoscopy to clear 99% of aerosolized pathogens. Thus, despite the known diagnostic and therapeutic benefits of bronchoscopy, concerns surrounding viral exposure related to an air leak from bronchoscopy adapters led to the development of national guidelines that limited bronchoscopy in patients affected by Covid-19.

To overcome these limitations, the investigators developed a novel bronchoscope adapter capable of effectively limiting all tidal volume loss (i.e., air leak via adapter) associated with the currently available standard adapters. In ex-vivo and animal studies, the investigators noted that a significant portion of tidal volume delivered by the ventilator is lost to the ambient air when using the standard bronchoscope adapter using the standard adapter (~40%). There was no tidal volume loss associated with the Leak-free bronchoscope adapter. The innovation in this adapter's development lies in the capacity to form an improved seal around the bronchoscope that limits tidal volume loss around the bronchoscope and captures air that escapes the circuit with the insertion and removal of the bronchoscope. Of note, the adapter is positioned outside the circuit and is only a conduit for the bronchoscope to enter the circuit.

Here the investigators propose a comparative effectiveness study that evaluates the leak-free bronchoscope adapter's use to enable bronchoscopy similar to the standard adapter. Additionally, the investigators will evaluate the adapter's capacity to deliver either equivalent tidal volume to the currently available adapter in patients undergoing bronchoscopy.

Study Type

Interventional

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • New York
      • New York, New York, United States, 10032
        • Recruiting
        • Columbia University Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  1. All patients diagnosed with respiratory failure who require mechanical ventilation for any reason except those listed below (exclusion criteria).
  2. Male or female
  3. Aged 18 years or above

Exclusion Criteria:

The participant may not enter the study if ANY of the following apply:

  1. Female participants who are pregnant during the study.
  2. SpO2 (oxygen saturation) < 90% at baseline.
  3. The patient requires high ventilation requirements as indicated by a positive end-expiratory pressure (PEEP) greater than 20 mmHg or inspired oxygen requirements greater than 90%.
  4. Hemodynamic instability requiring significant use of vasoactive pressors to maintain blood pressure.
  5. Any other significant disease or disorder that, in the Investigator's opinion, may either put the participants at risk because of participation in the study or may influence the result of the study.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Supportive Care
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: Double

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Leak-free bronchoscope adapter
The mechanical ventilation circuit used in patients with respiratory failure is considered a closed-loop circuit between the patient and the ventilator. This allows air to flow from the ventilator to the patient and back to the circuit without escaping to the ambient environment. To perform bronchoscopy, a standard adapter is spliced into the ventilator circuit, which allows the bronchoscope to enter the ventilation circuit granting access to the patient's airways. This study will compare the efficacy of the standard commercially available adapter to the newly developed leak-free adapter. The intervention is considered the "leak-free bronchoscope adapter."
A novel leak-free bronchoscope adapter will connect to the ventilation circuit in place of a standard adapter.
Active Comparator: Standard bronchoscope adapter
The mechanical ventilation circuit used in patients with respiratory failure is considered a closed-loop circuit between the patient and the ventilator. This allows air to flow from the ventilator to the patient and back to the circuit without escaping to the ambient environment. To perform bronchoscopy, a standard adapter is spliced into the ventilator circuit, which allows the bronchoscope to enter the ventilation circuit granting access to the patient's airways. This study will compare the efficacy of the standard commercially available adapter to the newly developed leak-free adapter. This arm utilizes the standard adapter.
Commercially available bronchoscope adapter (Smiths Medical)

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Change from baseline of tidal volume loss during procedure
Time Frame: Baseline and at one minute into the procedure
To measure tidal volume loss, the patient's ventilator displays both total volume of inspired (VTi) air and expired (VTe) air for each breath given. Immediately prior to the start of bronchoscopy, baseline values for VTi and VTe will be recorded for 10 breaths. Upon initiation of bronchoscopy, the VTi and VTe values will be recorded for 10 breaths one minute after the initiation of the bronchoscopy procedure.
Baseline and at one minute into the procedure
Change from baseline of oxygen saturation levels during procedure
Time Frame: Baseline and at one minute into the procedure
To measure patient oxygen saturation (SpO2) as measured using a pulse oximeter through the procedure. Immediately prior to bronchoscopy, the patient's baseline circulating oxygen saturation will be recorded during 10 ventilated breaths. One minute after initiation of bronchoscopy, the SpO2 level will be recorded for 10 breaths during the procedure. The oxygen saturation sensor is a standard non-invasive device placed on the patient and provides continuous pulse oximetry monitoring.
Baseline and at one minute into the procedure
Change from baseline of Particulate matter (PM) in air immediately post-procedure
Time Frame: Baseline and at one minute post-procedure
Air quality samples will be obtained to monitor the level and degree of circulating particulate matter using an air quality sampler. The sampling device will obtain an air sample immediately before the bronchoscopic procedure as baseline, and a second sample will be obtained immediately after the procedure. The sampling device is a non-invasive tool placed within the room and near the participant during the procedure.
Baseline and at one minute post-procedure
Change from baseline of aerosolized pathogens immediately post-procedure
Time Frame: Baseline and at one minute post-procedure
Air quality samples will be obtained to monitor the level and degree of aerosolized pathogens using an air quality sampler. The sampling device will obtain an air sample immediately before the bronchoscopic procedure as baseline, and a second sample will be obtained immediately following the procedure. The sampling device is a non-invasive tool placed within the room and near the participant during the procedure.
Baseline and at one minute post-procedure
Change from baseline of Pulmonary dynamic compliance immediately post-procedure
Time Frame: Baseline and at one minute post-procedure
Immediately prior to initiating the bronchoscopic procedure, a baseline pulmonary dynamic compliance value will be obtained from the ventilator. The ventilator provides this continuous value in all intubated patients. A repeat value will then be obtained immediately following the completion of the bronchoscopic procedure.
Baseline and at one minute post-procedure

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Change from baseline of Heart rate during procedure
Time Frame: Baseline and at one minute into the procedure
Immediately prior to the initiation of the bronchoscopic procedure, the patient's baseline heart rate will be recorded. These values will be continuously monitored throughout the bronchoscopic procedure and recorded through the bronchoscopic procedure.
Baseline and at one minute into the procedure
Change from baseline of systolic and diastolic blood pressure during procedure
Time Frame: Baseline and at one minute into the procedure
Immediately prior to the initiation of the bronchoscopic procedure, the patient's baseline systolic and diastolic blood pressure will be recorded. These values will be continuously monitored throughout the bronchoscopic procedure and recorded through the bronchoscopic procedure.
Baseline and at one minute into the procedure

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Beth Hochman, MD, Columbia University

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Estimated)

May 1, 2023

Primary Completion (Estimated)

June 1, 2024

Study Completion (Estimated)

June 1, 2024

Study Registration Dates

First Submitted

November 23, 2021

First Submitted That Met QC Criteria

March 9, 2022

First Posted (Actual)

March 21, 2022

Study Record Updates

Last Update Posted (Actual)

September 1, 2023

Last Update Submitted That Met QC Criteria

August 30, 2023

Last Verified

August 1, 2023

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

IPD Plan Description

Patient data is anonymized for study analysis. There are no plans to share data beyond the scope of researchers directly involved with this study.

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Respiratory Insufficiency

Clinical Trials on Leak-free Bronchoscope Adapter

Subscribe