First-in-Human Trial of the A3-Shield System for Stabilizing Abdominal Aortic Aneurysms (PRINCIPIIS-FIH)

January 5, 2026 updated by: Angiolutions GmbH

PRemarket INvestigation to Evaluate the Clinical Safety and effectIveness of the Protective "A3- Shield" System for the stabIlIzation of Abdominal Aortic aneurysmS - Pilot Phase

This clinical trial evaluates the safety and feasibility of the A3-Shield implant for stabilising small abdominal aortic aneurysms in adult patients. The study enrolls men and women with small, asymptomatic infrarenal abdominal aortic aneurysms who are eligible for an elective catheter-based procedure.

The trial aims to determine whether implantation of the A3-Shield device is technically feasible and whether it is associated with mechanical immobilization of the aneurysm neck. The study also assesses the peri-procedural and safety profile of the device.

Participants will undergo implantation of the A3-Shield device and attend scheduled follow-up visits that include clinical assessments and imaging such as ultrasound and computed tomography. The study does not include randomisation or a separate control group.

Study Overview

Status

Recruiting

Intervention / Treatment

Detailed Description

Current management of abdominal aortic aneurysms is severely limited by the lack of therapeutic options that can halt the underlying dilating biology. At present, there is no available intervention that can sufficiently intercept the size progression of small abdominal aortic aneurysms. This gap is highly relevant because the vast majority of abdominal aortic aneurysms are detected at a small size, and a substantial proportion of these aneurysms enlarge over the following years and eventually require prosthetic repair, which carries procedure-related and long-term risks. The limited efficacy of endovascular aortic repair in small aneurysms and the significant long-term complications observed in large aneurysms treated by endovascular aortic repair are thought to be related, at least in part, to insufficient control of aneurysm sac expansion. Effective stabilisation of aneurysm size at an early stage therefore represents a major unmet need in abdominal aortic aneurysm management.

The A3-Shield implant introduces a new approach to aneurysm treatment by targeting the biomechanical forces that contribute to aneurysm enlargement. Each heartbeat generates aortic pulse waves that propagate along the vessel and induce pulsatile wall motion. As these pulse waves reach the aneurysmal segment, they generate concentrated mechanical stress at the AAA shoulder region. This stress concentration is believed to promote further expansion of the aneurysm. The A3-Shield implant is designed to counteract this mechanism by mechanically immobilizing the aneurysm neck and gradually dampening the transmission of harmful pulse wave forces into the aneurysm sac, thereby aiming to protect the aneurysm from stress-induced enlargement.

This first-in-human clinical trial evaluates the safety and feasibility of the A3-Shield implant in adult patients with small infrarenal abdominal aortic aneurysms. The primary objectives are to determine whether the implantation procedure is technically feasible and whether it is associated with suppression of pulse wave transmission at the aneurysm neck. The study also assesses the peri-procedural and safety profile of the device.

Participants undergo a catheter-based implantation of the A3-Shield device under fluoroscopic guidance. Following the procedure, they attend scheduled follow-up visits that include clinical examinations and imaging such as ultrasound and computed tomography. The study does not include randomisation or a separate control group.

The findings from this pilot phase are intended to support further clinical development and may provide early evidence for a new interventional strategy that addresses an unmet need in the management of small abdominal aortic aneurysms.

Study Type

Interventional

Enrollment (Estimated)

12

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Contact Backup

Study Locations

      • Tashkent, Uzbekistan, 100027
        • Recruiting
        • Еzgu Niyат Llc
        • Contact:

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

IC1. Male and female subjects ≥ 18 years at the time of screening. Female subjects must be of non-childbearing potential or otherwise not pregnant and not planning to become pregnant during the study follow-up period.

IC2. Subject has a documented fusiform infrarenal abdominal aortic aneurysm characterized by the following:

  • Aneurysm size of 3.0-5.0 cm in maximum diameter for males; 3.0 - 4.5 for females (diameter measured in CTA perpendicular to the line of flow)
  • Aortic neck diameter ≥17mm and ≤ 29mm
  • Adequate landing zone for a tubular device of ≥ 30 mm length

IC3. Access vessel size and morphology allows endovascular access of 14F (or larger) introducer sheaths and catheters

IC4. Subject (or legal representative) understands the study requirements and the procedures and provided written informed consent.

IC5. Subject, family member, and/or legal representative agree(s) and subject is capable of returning to the study hospital for all required scheduled follow-up visits.

Exclusion Criteria:

EC1. Subject has an aneurysm that is:

  • symptomatic
  • ruptured
  • thoracic
  • suprarenal
  • thoraco-abdominal
  • isolated ilio-femoral
  • mycotic
  • inflammatory
  • pseudoaneurysmatic
  • traumatic

EC2. Subject has any type of aortic dissection.

EC3. Subject has congenital connective tissue disease, e.g., Marfan's Syndrome or Ehlers-Danlos Syndrome.

EC4. Subject has significant (>25% of vessel circumference of aortic neck) aortic mural thrombus at the attachment site that could compromise fixation of the device.

EC5. Subject has arterial access that is not acceptable for the device delivery systems as defined in the device Instructions For Use.

EC6. Subject has had an acute myocardial infarction within 180 days prior to the index procedure (defined as Q-wave MI or non-Q-wave MI with total CK elevation ≥ twice normal in the presence of CK-MB elevation and/or troponin elevation).

EC7. Subject has untreated coronary artery disease, which in the opinion of the treating physician, is clinically significant and requires revascularization.

EC8. Subject has congestive heart failure (NYHA class III or IV) or high-grade valvular disease.

EC9. Subject has had a cerebrovascular accident or transient ischemic attack within the past 90 days prior to study enrollment.

EC10. Subject has end-stage renal disease or has a creatinine > 2.0 mg/dl or is on dialysis.

EC11. Subject has a history of bleeding diathesis or coagulopathy.

EC12. Subject has Hgb <9 g/dL, platelet count <50,000 cells/mm3 or >700,000 cells/mm3, or white blood cell count <1,000 cells/mm3.

EC13. Subject has had or plans to have an unrelated major surgical or interventional procedure (including trans-aortic interventions) within 30 days prior or post implantation of the device.

EC14. Subject has known hypersensitivity to contrast agents that cannot be adequately pre-medicated or has known hypersensitivity to the device components.

EC15. Subject is participating in another investigational drug or device study and has not completed the follow-up required for that study at least 30 days prior to signing the informed consent form in this study.

EC16. Subject is morbidly obese (body mass index ≥ 40 kg/m2) or has other documented conditions that inhibit radiographic visualization of the aorta.

EC17. Subject has an active infection or chronic systemic illness at the time of the index procedure that may interfere with study objectives.

EC18. Subject has current problems with substance abuse.Smoking is NOT an exclusion criterion.

EC19. Subject has severe incapacitating dementia.

EC20. Subject has a life expectancy of less than 24 months due to comorbid conditions based on the assessment of the investigator at the time of enrollment.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Device implantation
A3-Shield Implantation
Implantation of the A3-Shield Device in the AAA Neck

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Primary safety endpoint: Absence of Major Adverse Events
Time Frame: Within 30 days of the index procedure
Absence of Major Adverse Events (MAEs). MAE is defined as a composite of death, stroke, myocardial infarction, new onset renal failure (requiring dialysis), respiratory failure (requiring mechanical ventilation), paralysis or paraparesis, bowel ischemia requiring surgical intervention or procedural blood loss ≥ 1,000 cc.
Within 30 days of the index procedure
Primary effectiveness endpoint: Successful aneurysm treatment
Time Frame: Conclusion of the index procedure

Successful aneurysm treatment defined as the composite of

  1. Technical success, defined as successful insertion of the delivery system through the vasculature, successful deployment of the device at the intended location, and successful removal of the delivery system
  2. Immobilization of the distal AAA neck, defined as aortic circumferential cyclic strain < 10% as measured via vascular ultrasound at the conclusion of the index procedure and assessed by an independent core laboratory
Conclusion of the index procedure

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Secondary safety endpoint: Absence of Major Adverse Events
Time Frame: 180 days, 12 months,vand 24 months post index procedure.
Absence of Major Adverse Events (MAEs)
180 days, 12 months,vand 24 months post index procedure.
Secondary safety endpoint: Procedure related complications
Time Frame: 30 days, 6 months, 12 months, and 24 months post index procedure
Procedure related complications
30 days, 6 months, 12 months, and 24 months post index procedure
Secondary safety endpoint: Device related complications
Time Frame: 30 days, 6 months, 12 months, and 24 months post index procedure
Device related complications. Device related complications are defined as device fracture(s), device migration, and device occlusion requiring intervention.
30 days, 6 months, 12 months, and 24 months post index procedure
Secondary effectiveness endpoint: Freedom of aneurysmal sac enlargement
Time Frame: Day 30, 12 months, and 24 months post index procedure
Freedom of aneurysmal sac enlargement. Enlargement is defined as a maximum diameter increase of > 5 mm per year, or aneurysm sac volume increase of > 10% per year, as assessed by CT scans and an independent core laboratory
Day 30, 12 months, and 24 months post index procedure
Secondary effectiveness endpoint: Aneurysmal sac shrinkage
Time Frame: Day 30, 12 months, and 24 months post index procedure
Aneurysmal sac shrinkage. Shrinkage is defined as a maximum diameter decrease of > 5 mm per year, or aneurysm sac volume decrease of > 10% per year, as assessed by CT scans and an independent core laboratory
Day 30, 12 months, and 24 months post index procedure
Secondary effectiveness endpoint: Technical Success
Time Frame: Day 30
Technical Success, defined as the presence of the device in the intended location and absence of device occlusion as assessed by CT scans and an independent core laboratory
Day 30
Secondary effectiveness endpoint: Immobilization of the distal AAA neck
Time Frame: Day 30, 6 months, 12 months and 24 months
Immobilization of the distal AAA neck, defined as aortic circumferential cyclic strain < 10% as measured via vascular ultrasound and assessed by an independent core laboratory
Day 30, 6 months, 12 months and 24 months
Secondary effectiveness endpoint: Absence of requirement of AAA repair
Time Frame: 30 days, 6 months, 12 months, and 24 months post index procedure.
Absence of requirement of AAA repair (OR or EVAR)
30 days, 6 months, 12 months, and 24 months post index procedure.
Secondary effectiveness endpoint: Absence of aneurysm related mortality
Time Frame: 6 months, 12 months, and 24 months post index procedure
Absence of aneurysm related mortality
6 months, 12 months, and 24 months post index procedure

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Quality of Life (QoL)
Time Frame: 6 months, 12 months, 4 months
Short Form-36 (SF-36) Health Survey will be used to assess quality of life. The scale ranges from 0 to 100, where higher scores represent better quality of life. Subscale scores (each 0-100, higher is better) will also be reported.
6 months, 12 months, 4 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Study Director: Uwe Raaz, MD, Angiolutions GmbH

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

November 28, 2025

Primary Completion (Estimated)

July 1, 2026

Study Completion (Estimated)

July 1, 2028

Study Registration Dates

First Submitted

November 28, 2025

First Submitted That Met QC Criteria

January 5, 2026

First Posted (Actual)

January 9, 2026

Study Record Updates

Last Update Posted (Actual)

January 9, 2026

Last Update Submitted That Met QC Criteria

January 5, 2026

Last Verified

January 1, 2026

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Aneurysm, Abdominal Aortic

Clinical Trials on A3-Shield Implantation