Nachtelijke polysomnografie en ademhalingsvolumemonitor
De evaluatie van een ademhalingsvolumemonitor bij patiënten die een nachtelijke polysomnografie ondergaan
Sleep disordered breathing; specifically obstructive sleep apnea (OSA) is a disease affecting 8-12% of the general population and often more than 70% of the bariatric surgical population. OSA is characterized by the repetitive collapse of the upper airway, causing a reduction or cessation in airflow and decreases in oxygen saturation. These events are resolved by arousals from sleep, reducing sleep quality and leading to excessive daytime sleepiness. An in-laboratory polysomnography (PSG) is the gold standard for the diagnosis of OSA.
Previous studies have established obstructive sleep apnea (OSA) as a potential independent risk factor for postoperative complications, adverse surgical outcomes, and longer hospital stays.
Patients with OSA have an increase in postoperative complications, the most frequent being oxygen desaturation, postoperative atelectasis and increased postoperative pain. Despite the clear risks, OSA remains under diagnosed with an estimated 25-30% of patients at a high risk for OSA. It has been suggested that OSA events may be even more frequent post operatively because of the residual effects of anesthesia and the use of potent pain medications such as opioids. Postoperatively apneas often go undetected and untreated. The use of supplemental O2 may mask any desaturations and there is no convenient technology to noninvasively monitor ventilation to detect apnea and hypopnea in post-surgical patients.
New advances in technology and digital signal processing have led to the development of an impedance based Respiratory Volume Monitor (RVM). The RVM (ExSpiron™, Respiratory Motion, Inc.; Waltham, MA) has been shown to provide accurate real-time, continuous, non-invasive measurements of tidal volume (TV), minute ventilation (MV) and respiratory rate (RR). Our main hypotheses are that the non-invasive, impedance-based RVM monitor will accurately reflect TV, RR and MV during sleep and will detect apneas and hypopneas accurately.
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Gedetailleerde beschrijving
This is an observational pilot study to evaluate the clinical applicability of the ExSpiron Respiratory Volume Monitor (RVM) in patients with suspected sleep apnea. Previous work has demonstrated the ability of the RVM to provide non-invasive, real-time, continuous measurements of respiratory parameters (MV, TV, and RR) but those studies have not specifically looked at patients during sleep or patients with sleep disordered breathing.
Previously there was no device capable of providing, continuous, non-invasive, real time measurements of ventilatory status such as minute ventilation, tidal volume and RR. The ExSpiron system utilizing impedance based technology and proprietary algorithms (Respiratory Motion Inc., Waltham, MA) have been developed to obtain these measurements.
Respiratory Motion, Inc. has previously evaluated ExSpiron measurements of respiratory parameters (MV, TV and RR) in studies of adult volunteer subjects. Stimulus leads delivered an alternating minimal current (1.5 mA) at a frequency of 50 kHz and recording leads recorded differential voltages to calculate impedance. Subject data (height, weight, age, gender, chest circumference) were measured. The studies showed that using the optimal lead placement configuration, algorithms based on chest wall impedance and respiratory pattern analysis provided provided RVM measurements and ExSpiron curves which correlated strongly with spirometry volumes and spirometry curves.
A large percentage of patients in the general surgical population have undiagnosed OSA. This population is at increased risk of adverse perioperative respiratory complications. Assessment and management of perioperative respiratory function and early intervention when indicated is a multifaceted, complex task often complicated by the unavailability of an accurate and continuous monitoring system that can demonstrate apnea and hypopnea to guide clinical decision making. The ExSpiron system is designed for these patients and is intended to address some of the limitations of the current generation of hospital monitors as listed below:
In the, non-ventilated patient, current monitoring devices do not provide objective non-invasive, continuous real time measurements of important respiratory parameters MV, TV and RR that reflect respiratory competence, nor do they provide adequate assessment of apnea and hypopnea.
Current monitoring of non-intubated patients mostly relies on oximetry data, subjective clinical assessment which measures respiration and not ventilation and rarely transcutaneous CO2 measurements. In appropriate clinical settings, the ExSpiron system can provide direct quantitative measure of ventilation (MV, TV and RR) that can be used in the detection of apnea and hypopnea.
Precise apnea and hypopnea assessment.
Most if not all of the gaps left by current respiratory assessment technologies may be filled by a non-invasive monitor such as the one proposed for use in this study. There are many potential applications for this technology including: use in peri-procedural environment to evaluate the effects of medication on ventilation and detection of apnea/hypopnea. In addition, it may be useful in clinical decision making regarding medication adjustments of pain management protocols, evaluation ventilatory status prior to extubation and evaluation of the need for reintubation, but also trending of respiratory patterns to avoid risky emergency airway management interventions.
New advances in technology and digital signal processing have led to the development of an impedance based Respiratory Volume Monitor (RVM). The RVM (ExSpiron™, Respiratory Motion, Inc.; Waltham, MA) has been shown to provide accurate real-time, continuous, non-invasive measurements of tidal volume (TV), minute ventilation (MV) and respiratory rate (RR). Our main hypotheses are that the non-invasive, impedance-based RVM monitor will accurately reflect TV, RR and MV during sleep and will detect apneas and hypopneas accurately.
Main Hypothesis: The ExSpiron monitor accurately reflects apnea and hypopnea events compared to standard polysomnography. The ExSpiron monitor is able to distinguish between central and obstructive apneic events.
Secondary Hypothesis: 1) Correlation of MV% as compared to % predicted with comorbidities (e.g. hypertension and metabolic syndrome), 2)Correlation of MV changes as compared to MV% predicted with oxygen desaturation index (ODI), 3) evaluation breathing patterns before and after obstruction, 4) evaluation of breathing patterns in different types of apnea (central, obstructive and mixed).
Studietype
Inschrijving (Geschat)
Contacten en locaties
Studie Locaties
-
-
Massachusetts
-
Boston, Massachusetts, Verenigde Staten, 02111
- Tufts Medical Center
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Bemonsteringsmethode
Studie Bevolking
All adult patients referred to the sleep laboratory at Tufts Medical Center for a standard polysomnogram (PSG) are eligible to be recruited.
Beschrijving
Inclusion Criteria:
- Adult Patients referred to the sleep lab for a polysomnography for suspicion of sleep disordered breathing aged ≥ 18 years. Ability to provide written informed consent.
Exclusion Criteria:
- Patients not competent to give informed consent, patients with implantable electronic devices, such as pacemakers, stimulators, etc. pregnant or lactating women.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Observatiemodellen: COHORT
- Tijdsperspectieven: PROSPECTIVE
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Tijdsspanne |
|---|---|
|
Overeenstemming tussen het aantal en de duur van apneu-gebeurtenissen gemeten als AHI (apneu-hypopneu-index) beoordeeld met standaard polysomnografie versus Exspiron-monitor
Tijdsspanne: Duur van het polysomnografieonderzoek, meestal 1 beoordeling gedurende de nacht
|
Duur van het polysomnografieonderzoek, meestal 1 beoordeling gedurende de nacht
|
Secundaire uitkomstmaten
Uitkomstmaat |
Tijdsspanne |
|---|---|
|
Mogelijkheid van de Exspiron-monitor om onderscheid te maken tussen centrale en obstructieve apneu-gebeurtenissen. Er is geen maateenheid voor deze vergelijking
Tijdsspanne: Duur van het polysomnografieonderzoek, meestal 1 beoordeling gedurende de nacht
|
Duur van het polysomnografieonderzoek, meestal 1 beoordeling gedurende de nacht
|
Medewerkers en onderzoekers
Sponsor
Medewerkers
Onderzoekers
- Hoofdonderzoeker: Roman Schumann, MD, Tufts Medical Center
Studie record data
Bestudeer belangrijke data
Studie start (Werkelijk)
Primaire voltooiing (Geschat)
Studie voltooiing (Geschat)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Geschat)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- IRB # 10880
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .