Haalbaarheid van het vergelijken van microgolfbeelden en mammogrammen bij patiënten die worden teruggeroepen voor vervolgonderzoek van de borst
Haalbaarheid van het vergelijken van microgolfbeelden en mammogrammen bij patiënten die teruggeroepen worden voor vervolgonderzoek van de borst
The goal of this clinical trial is to measure the electrical properties of breast tissues using microwave technology and to compare these collected measurements to clinical information in mammograms and/or breast ultrasound for women that have been recalled for follow-up breast imaging or further investigation. The main questions it aims to answer are:
- Can the breast area and plate separation in the microwave scan be compared to mammogram data (where available) to assess similarity in breast positioning and the feasibility of image registration?
- Can the distribution of electrical properties specifically glandular tissue distribution in the microwave image be compared to that observed in the mammogram (where available)?
- Can the characteristics of the microwave image in the localized region corresponding to an identified anomaly in the breast imaging data be examined?
Study participants will be scanned with the Wave View microwave scanner on the same day as their follow-up breast imaging appointment. Each breast will be scanned up to 3 times with the scanner oriented in both the horizontal position and oriented at 45 degrees. Study participants will also be asked to fill out three surveys, one relating to breast health and the other two relating to their scanning experience.
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Gedetailleerde beschrijving
A new approach to imaging that uses low-power microwaves to scan the breast has been developed. Since the imaging system uses low power, there is no risk of tissue heating. The imaging system places the sensors in contact with the breast tissue. The research team at the University of Calgary has previously developed microwave imaging systems that have been tested on healthy volunteers and breast cancer patients. A group of healthy volunteers were repeatedly scanned to evaluate the consistency of the imaging approach. Another study involved scanning breast cancer patients before each round of neoadjuvant chemotherapy. The information collected from experiences scanning volunteers and patients have been reviewed and translated into a next-generation imaging system design. Aspects of this system have been further improved by Wave View Imaging, which is a company formed by members of the University's research team.
The proposed study involves collecting electrical property measurements of breast tissues and comparing these measurements to clinical information in mammograms and/or breast ultrasounds. The investigators also aim to compare regions of interest in the microwave scans with regions of interest in the mammograms (where available). This includes regions dominated by fibroglandular tissues, as well as regions of interest that correspond to anomalies identified in the mammograms and/or breast ultrasounds.
The plan is to scan both breasts of up to 100 volunteers that have been recalled for follow-up breast imaging or further investigation. Each volunteer will participate in a single scanning session during their follow-up appointment. Breast tissue composition will be compared with the microwave properties and registration of mammograms (where available) with microwave images will be performed for a more detailed comparison.
Studietype
Inschrijving (Geschat)
Fase
- Niet toepasbaar
Contacten en locaties
Studie Locaties
-
-
-
Calgary, Canada
- Pureform Radiology
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Beschrijving
Inclusion Criteria:
- Female sex
- 18-74 years of age
- Able to provide informed consent
- Had a screening or diagnostic mammogram or breast ultrasound within 6 weeks
- Recalled for additional breast imaging due to BIRADS 0 or 3 or 4 or 5 findings.
Exclusion Criteria:
- Participants who have not received a mammogram or breast ultrasound
- Participants who are currently undergoing breast cancer treatment or have had treatment in the past 6 months
- Participants with breast implants
- Participants who are pregnant or breastfeeding
- Participants with active breast skin infections
- Participants with nipple piercings (unless they are removed prior to scanning)
- Participants with an implanted electronic device
- Participants with physical limitations that prevent them from placing their breasts inside the Wave View scanner.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Ander
- Toewijzing: NVT
- Interventioneel model: Opdracht voor een enkele groep
- Masker: Geen (open label)
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
|---|---|
|
Experimenteel: WVI-MIS-02-2025
A single arm study in which the collected microwave data is compared against available clinical data. |
Microwave scans |
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Measurement of electrical properties of breast tissues in microwave images
Tijdsspanne: 12 months
|
The average electrical properties of the Wave View image will be calculated (relative permittivity and attenuation). The average properties of the right and left breasts will be compared. The average properties of the craniocaudal (CC) and mediolateral oblique (MLO) views will also be compared for both breasts. |
12 months
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Examine the similarity in plate separation between mammography and Wave View scans
Tijdsspanne: 12 months
|
The separation between the plates contacting the breast will be compared between the two modalities. This comparison provides an initial indication of the volume of breast tissue in the scanner. |
12 months
|
|
Breast area
Tijdsspanne: 12 months
|
The breast area on the mammogram will be compared to the breast area on the Wave View scan. The breast area in the Wave View image will be calculated, while the breast area on the mammogram is available from automated image analysis reports. |
12 months
|
|
Examine the similarity between glandular tissue distribution on mammograms and electrical property images obtained with the Wave View scanner.
Tijdsspanne: 12 months
|
The microwave image will be overlayed on the mammogram to provide initial insight on glandular tissue distribution via visual inspection. The images will be registered based on the outline of the breast. If the breast areas are found to differ, deformable registration techniques will be applied. Higher intensity regions representing glandular tissues will be identified in both images via image filtering and segmentation. These regions will be compared with metrics such as the Dice coefficient. In addition, the area of the higher intensity region will be compared with the overall area of the breast image, similar to the method used to calculate the percentages used in breast density analysis. |
12 months
|
|
Examine how closely the presence of responses in Wave View scans correspond to anomalies detected on mammograms and/or breast ultrasounds
Tijdsspanne: 12 months
|
Localized regions in the Wave View image that correspond to locations of anomalies on the mammogram or breast ultrasounds will be identified based on the registered images. The average properties of these regions will be compared with the average properties of a region in the same breast, as well as in the contralateral breast. |
12 months
|
|
Estimate of breast composition
Tijdsspanne: 12 months
|
Pureform Radiology uses a breast density tool to automatically calculate the volume of the breast and glandular tissue, as well as the percentage density and density category (A, B, C, D). Microwave properties are expected to be greater with greater breast densities.To compare electrical properties to breast density, average electrical properties of each microwave scan will be plotted against the corresponding percent density. |
12 months
|
Medewerkers en onderzoekers
Sponsor
Publicaties en nuttige links
Algemene publicaties
- C. M. Butterworth, P. Mojabi, and E. C. Fear, "Quantifying Consistency of Microwave Breast Imaging: Laser Scanning for Assessing Breast Volume and Shape," IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology, pp. 1-9, 2025, doi: 10.1109/JERM.2025.3531693
- Mojabi P, Bourqui J, Lasemiimeni Z, Grewal B, Fear E. Microwave Imaging for Breast Cancer Detection: Performance Assessment of a Next-Generation Transmission System. IEEE Trans Biomed Eng. 2025 Jun;72(6):1787-1799. doi: 10.1109/TBME.2024.3521410.
- Wang Z, Mojabi P, Price S, Lasemiimeni Z, Besler B, Bourqui J, Deutscher D, Besler B, Shen H, Docktor B, Tsang RY, Fear E. Reproducibility of Microwave Breast Imaging: Analysis of Regular Scans of a Group of Volunteers. IEEE Trans Biomed Eng. 2025 May;72(5):1562-1571. doi: 10.1109/TBME.2024.3512572. Epub 2025 Apr 22.
- Mojabi P, Bourqui J, Fear E. Fast 3D Breast Imaging With a Transmission-Based Microwave System. IEEE Trans Med Imaging. 2025 May;44(5):2206-2217. doi: 10.1109/TMI.2025.3527916. Epub 2025 May 2.
- Smith K, Bourqui J, Wang Z, Besler B, Lesiuk M, Roumeliotis M, Quirk S, Grendarova P, Pinilla J, Price S, Docktor B, Fear E. Microwave imaging for monitoring breast cancer treatment: A pilot study. Med Phys. 2023 Nov;50(11):7118-7129. doi: 10.1002/mp.16756. Epub 2023 Oct 6.
- Smith, K., et al. "Microwave Imaging of the Breast: Consistency of Measurements Over Time." IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology, vol. 6, no. 1, 2022, pp. 61-67, https://doi.org/10.1109/JERM.2021.3099014. . : . .
Studie record data
Bestudeer belangrijke data
Studie start (Werkelijk)
Primaire voltooiing (Geschat)
Studie voltooiing (Geschat)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Andere studie-ID-nummers
- HREBA.CC-25-0255
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .