Jämförelse av pulsoximetri (SpO2) med olika oximetrar och artärmättnad (SaO2): Oxygap2-studie (Oxygap 2)

3 november 2026 uppdaterad av: François Lellouche, Laval University

The oximeter is used to monitor intensive care patients undergoing oxygen therapy. It indicates pulsed oxygen saturation (SpO2), a reflection of arterial oxygen saturation (SaO2) which enables detection of hypoxemia and hyperoxia, both deleterious state. Current SpO2 recommendations aim to reduce both risk of hypoxemia and hyperoxia. SpO2 is considered the 5th vital sign.

Current recommendations for SpO2 targets do not consider the variability of oximeters used in clinical practice. This variability and lack of specification represent an obstacle to an optimal practice of oxygen therapy.

Thus, this study aims to compare the SpO2 values of different oximeters (General Electric-GE, Medtronic, Masimo and Nonin) used in clinical practice with the SaO2 reference value obtained by an arterial gas in order to specify the precision and the systematic biases of the oximeters studied. This data will also make it possible to refine the recommendations concerning optimal oxygenation

Studieöversikt

Status

Rekrytering

Betingelser

Intervention / Behandling

Studietyp

Observationell

Inskrivning (Beräknad)

250

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

      • Les Abymes, Guadeloupe, 97139
    • Quebec
      • Montreal, Quebec, Kanada, H3T 1C5
      • Montreal, Quebec, Kanada, H2X0A9
        • Avslutad
        • CHUM
      • Québec, Quebec, Kanada, G1V 4G5
        • Rekrytering
        • Institut Universitaire de Cardiologie et de Pneumologie de Québec
        • Kontakt:
      • Sousse, Tunisien, 4000
        • Har inte rekryterat ännu
        • Hôpital Universitaire Farhat Hached,
        • Kontakt:

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

1 sekund till 100 år (Barn, Vuxen, Äldre vuxen)

Tar emot friska volontärer

Nej

Kön som är behöriga för studier

Allt

Testmetod

Icke-sannolikhetsprov

Studera befolkning

All patient in ICU with arterial catheter already installed will be consider

Beskrivning

Adult population

Inclusion Criteria:

  • ≥ 18 years old (adult population) -
  • Patients admitted to the Intensive Care Unit
  • Artery catheter already installed

Pediatric population

  • Below 18 years old
  • Patients admitted to the Intensive Care Unit
  • Artery catheter already installed

Exclusion Criteria:

  • No or poor signal with the usual pulse oximeter/based on clinician judgment
  • High dose of vasopressors or inotropes (epinephrine or norepinephrine ˃ 1mcg/kg/min), shock state (lactates above 3 mmoles/L)
  • Pigmented nails or nail polish
  • Methemoglobinemia history
  • Hemoglobin below 80 g/L
  • Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-COV-2…)
  • Prone position, Extra Corporel Membrane Oxygenator

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Observationsmodeller: COHORT
  • Tidsperspektiv: PROSPECTIVE

Kohorter och interventioner

Grupp / Kohort
Intervention / Behandling
Pediatric population

Pediatric population < 18 years old (estimated 100 patients)

Several pulse oximeter will be place on patient find few minutes before arterial blood gases. During an arterial blood sample, all SpO2 will be recorded

Adult population

Adult population >or= 18 years old (estimated 150 patients)

Several pulse oximeter will be place on patient find few minutes before arterial blood gases. During an arterial blood sample, all SpO2 will be recorded

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Difference between GE oximeter SpO2 value and the SaO2 value on arterial blood gases
Tidsram: During procedure (arterial blood gas sample collection)

Difference between GE oximeter SpO2 and the SaO2 value

During procedure (arterial blood gas sample collection)
Difference between Nellcor oximeter SpO2 value and the SaO2 value on arterial blood gases
Tidsram: During procedure (arterial blood gassample collection)

Difference between Nellcor oximeter SpO2 and the SaO2 value

During procedure (arterial blood gassample collection)
Difference between Masimo oximeter SpO2 value and the SaO2 value on arterial blood gases
Tidsram: During procedure (arterial blood gas sample collection)

Difference between Masimo oximeter SpO2 and the SaO2 value

During procedure (arterial blood gas sample collection)
Difference between Nonin oximeter SpO2 value and the SaO2 value on arterial blood gases
Tidsram: During procedure (arterial blood gas sample collection)

Difference between Nonin oximeter SpO2 and the SaO2 value

During procedure (arterial blood gas sample collection)

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Overestimation and underestimation of SaO2
Tidsram: During procedure (arterial blood gas sample collection)

Percent of pairs with overestimation and underestimation of SaO2 for each oximeter.

During procedure (arterial blood gas sample collection)
Hypoxemia detection
Tidsram: During procedure (arterial blood gas sample collection)

Ability to detect hypoxemia (SpO2 below 90% and PaO2 < 60 mmHg)

During procedure (arterial blood gas sample collection)

Andra resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Hyperoxemia detection
Tidsram: During procedure (arterial blood gas sample collection)

Ability to detect hyperoxemia (SpO2 above 95% and PaO2 > 100 mmHg)

During procedure (arterial blood gas sample collection)
Gaps between the different oximeters
Tidsram: During procedure (arterial blood gas sample collection)

Average of the SpO2 value gaps between the different oximeters tested

During procedure (arterial blood gas sample collection)

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Sponsor

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

20 juni 2024

Primärt slutförande (Beräknad)

30 juni 2027

Avslutad studie (Beräknad)

30 december 2027

Studieregistreringsdatum

Först inskickad

6 oktober 2023

Först inskickad som uppfyllde QC-kriterierna

25 oktober 2023

Första postat (Faktisk)

26 oktober 2023

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

5 november 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

3 november 2026

Senast verifierad

1 november 2026

Mer information

Termer relaterade till denna studie

Andra studie-ID-nummer

  • 22373

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

NEJ

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .