Comparación de oximetría de pulso (SpO2) con diferentes oxímetros y saturación arterial (SaO2): estudio Oxygap2 (Oxygap 2)
The oximeter is used to monitor intensive care patients undergoing oxygen therapy. It indicates pulsed oxygen saturation (SpO2), a reflection of arterial oxygen saturation (SaO2) which enables detection of hypoxemia and hyperoxia, both deleterious state. Current SpO2 recommendations aim to reduce both risk of hypoxemia and hyperoxia. SpO2 is considered the 5th vital sign.
Current recommendations for SpO2 targets do not consider the variability of oximeters used in clinical practice. This variability and lack of specification represent an obstacle to an optimal practice of oxygen therapy.
Thus, this study aims to compare the SpO2 values of different oximeters (General Electric-GE, Medtronic, Masimo and Nonin) used in clinical practice with the SaO2 reference value obtained by an arterial gas in order to specify the precision and the systematic biases of the oximeters studied. This data will also make it possible to refine the recommendations concerning optimal oxygenation
Descripción general del estudio
Estado
Estado
Condiciones
Condiciones
Intervención / Tratamiento
Intervención / Tratamiento
Tipo de estudio
Tipo de estudio
Inscripción (Estimado)
Inscripción
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Quebec
-
Montreal, Quebec, Canadá, H3T 1C5
- Reclutamiento
- CHU Ste-Justine
-
Contacto:
- Michael Sauthier
- Número de teléfono: 514 345-4931
- Correo electrónico: Michael.Sauthier.med@ssss.gouv.qc.ca
-
Montreal, Quebec, Canadá, H2X0A9
- Terminado
- CHUM
-
Québec, Quebec, Canadá, G1V 4G5
- Reclutamiento
- Institut Universitaire de Cardiologie et de Pneumologie de Québec
-
Contacto:
- Francois Lellouche
- Número de teléfono: 3572 1-418-656-8711
- Correo electrónico: francois.lellouche@criucpq.ulaval.ca
-
-
-
-
-
Les Abymes, Guadalupe, 97139
- Aún no reclutando
- CHU de la Guadeloupe
-
Contacto:
- Frederic Martino
- Número de teléfono: : +590 590 89 11 32
- Correo electrónico: frederic.martino@chu-guadeloupe.fr
-
-
-
-
-
Sousse, Túnez, 4000
- Aún no reclutando
- Hôpital Universitaire Farhat Hached,
-
Contacto:
- Mohamed Boussarsar
- Número de teléfono: +216 73 102 500
- Correo electrónico: hamadi.boussarsar@gmail.com
-
-
Criterios de participación
Criterio de elegibilidad
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Método de muestreo
Población de estudio
All patient in ICU with arterial catheter already installed will be consider
Descripción
Adult population
Inclusion Criteria:
- ≥ 18 years old (adult population) -
- Patients admitted to the Intensive Care Unit
- Artery catheter already installed
Pediatric population
- Below 18 years old
- Patients admitted to the Intensive Care Unit
- Artery catheter already installed
Exclusion Criteria:
- No or poor signal with the usual pulse oximeter/based on clinician judgment
- High dose of vasopressors or inotropes (epinephrine or norepinephrine ˃ 1mcg/kg/min), shock state (lactates above 3 mmoles/L)
- Pigmented nails or nail polish
- Methemoglobinemia history
- Hemoglobin below 80 g/L
- Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-COV-2…)
- Prone position, Extra Corporel Membrane Oxygenator
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Modelos observacionales: COHORT
- Perspectivas temporales: PROSPECTIVE
Número de grupos/cohortes
Cohortes e Intervenciones
Grupo / CohorteGrupo / Cohorte |
Intervención / TratamientoIntervención / Tratamiento |
|---|---|
|
Pediatric population
Pediatric population < 18 years old (estimated 100 patients) |
Several pulse oximeter will be place on patient find few minutes before arterial blood gases. During an arterial blood sample, all SpO2 will be recorded |
|
Adult population
Adult population >or= 18 years old (estimated 150 patients) |
Several pulse oximeter will be place on patient find few minutes before arterial blood gases. During an arterial blood sample, all SpO2 will be recorded |
¿Qué mide el estudio?
Medidas de resultado primarias
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Difference between GE oximeter SpO2 value and the SaO2 value on arterial blood gases
Periodo de tiempo: During procedure (arterial blood gas sample collection)
|
Difference between GE oximeter SpO2 and the SaO2 value |
During procedure (arterial blood gas sample collection)
|
|
Difference between Nellcor oximeter SpO2 value and the SaO2 value on arterial blood gases
Periodo de tiempo: During procedure (arterial blood gassample collection)
|
Difference between Nellcor oximeter SpO2 and the SaO2 value |
During procedure (arterial blood gassample collection)
|
|
Difference between Masimo oximeter SpO2 value and the SaO2 value on arterial blood gases
Periodo de tiempo: During procedure (arterial blood gas sample collection)
|
Difference between Masimo oximeter SpO2 and the SaO2 value |
During procedure (arterial blood gas sample collection)
|
|
Difference between Nonin oximeter SpO2 value and the SaO2 value on arterial blood gases
Periodo de tiempo: During procedure (arterial blood gas sample collection)
|
Difference between Nonin oximeter SpO2 and the SaO2 value |
During procedure (arterial blood gas sample collection)
|
Medidas de resultado secundarias
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Overestimation and underestimation of SaO2
Periodo de tiempo: During procedure (arterial blood gas sample collection)
|
Percent of pairs with overestimation and underestimation of SaO2 for each oximeter. |
During procedure (arterial blood gas sample collection)
|
|
Hypoxemia detection
Periodo de tiempo: During procedure (arterial blood gas sample collection)
|
Ability to detect hypoxemia (SpO2 below 90% and PaO2 < 60 mmHg) |
During procedure (arterial blood gas sample collection)
|
Otras medidas de resultado
Otras medidas de resultado
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Hyperoxemia detection
Periodo de tiempo: During procedure (arterial blood gas sample collection)
|
Ability to detect hyperoxemia (SpO2 above 95% and PaO2 > 100 mmHg) |
During procedure (arterial blood gas sample collection)
|
|
Gaps between the different oximeters
Periodo de tiempo: During procedure (arterial blood gas sample collection)
|
Average of the SpO2 value gaps between the different oximeters tested |
During procedure (arterial blood gas sample collection)
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Inicio del estudio
Finalización primaria (Estimado)
Finalización primaria
Finalización del estudio (Estimado)
Finalización del estudio
Fechas de registro del estudio
Enviado por primera vez
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Publicado por primera vez
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización publicada
Última actualización enviada que cumplió con los criterios de control de calidad
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
Otros números de identificación del estudio
- 22373
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .