Corneal biomekanisk analys med hjälp av Brillouin-mikroskopi

28 september 2026 uppdaterad av: James Bradley Randleman, The Cleveland Clinic

The objective of this study is to measure the Brillouin biomechanical properties in keratoconic corneas and characterize biomechanical alterations that occur after corneal procedures that inherently strengthen or weaken the cornea by evaluating the change in Brillouin metrics before and after treatments.

Studieöversikt

Status

Anmälan via inbjudan

Intervention / Behandling

Detaljerad beskrivning

Surgical correction of myopia and keratoconus identification/management are separate but tightly intertwined issues of major significance. For both, there is an unmet need for direct measurements to evaluate corneal stiffness (i.e. its resistance to deformation). The prevalence of myopia is expected to double, affecting more than 50% of the US population, by 2050. Laser in situ keratomileusis (LASIK) is one of the most popular and successful surgeries in the world and compares favorably to long-term contact lens wear use for myopia correction. However, only ~10% of eligible patients undergo LASIK currently; the others cite safety concerns as a major factor in their decision. The primary risk for poor refractive surgery outcomes is biomechanical failure due to unidentified (subclinical) ectasia (i.e. keratoconus). Patients presenting for LASIK evaluation with atypical, suspicious corneal curvature but with undetermined true risk represent the leading reason for surgery screening failures. This results in good candidates being denied surgery, while up to 10% of truly poor candidates are still missed using current screening algorithms.

Keratoconus is up to 10 times more prevalent than the previously reported 1/2000 figure. Corneal cross-linking (CXL) is now FDA approved in the US for keratoconus treatment and is effective at stiffening the cornea and halting ectasia progression. Early identification of keratoconus is critical, but current tests in the clinic are morphological, not biomechanical, and therefore do not allow a definitive diagnosis at the earliest stages resulting in vision loss before CXL treatment is initiated. Thus, the need for accurate identification of subclinical ectasia has never been greater.

In the past years, newly developed technology, Brillouin microscopy, has emerged as the most promising tool to address this clinical need. This study will systemically address the critical gap in current knowledge by linking Brillouin mapping of corneal biomechanical alterations to abnormal morphological behavior and testing the findings in conditions where corneal biomechanics are abruptly altered, by: 1) weakening with refractive surgery procedures, and 2) strengthening through corneal cross-linking.

It is anticipated that a clinical tool assessing the mechanical state of the cornea will improve early diagnosis and management of keratoconus as well as refractive surgery planning. Ultimately, this will lead to predictive models where Brillouin measurements could be an accurate predictor of postoperative outcomes and thus aid in developing individualized surgical parameters.

Studietyp

Observationell

Inskrivning (Beräknad)

220

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

    • Ohio
      • Cleveland, Ohio, Förenta staterna, 44195
        • Cleveland Clinic Cole Eye Institute

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

18 år till 60 år (Vuxen)

Tar emot friska volontärer

Nej

Kön som är behöriga för studier

Allt

Testmetod

Icke-sannolikhetsprov

Studera befolkning

Group 1: Patients with normal corneas Group 2: Patients with keratoconus Group 3: Patients with normal corneas undergoing LASIK Group 4: Patients with normal corneas undergoing PRK Group 5: Patients with normal corneas undergoing SMILE Group 6: Patients with keratoconus undergoing CXL

Beskrivning

Inclusion Criteria:

  • patients aged 18-60 with keratoconus
  • patients aged 18-60 with normal corneas,
  • patients aged 18-60 undergoing refractive surgery (LASIK, PRK, SMILE)
  • patients aged 18-60 with keratoconus undergoing CXL

Exclusion Criteria:

  • outside age range
  • history of previous ocular surgeries
  • unable to cooperate for the Brillouin microscopic examination
  • unable to provide informed consent

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Observationsmodeller: CASE_CONTROL
  • Tidsperspektiv: CROSS_SECTIONAL

Kohorter och interventioner

Grupp / Kohort
Intervention / Behandling
1: Normal Controls

Patients with normal corneas without any prior surgery to serve as the control group

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

2 Keratoconus

Patients with various stages of keratoconus

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

3: LASIK

Patients with normal corneas who are undergoing laser in situ keratomileusis (LASIK)

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

Group 4: PRK

Patients with normal corneas who are undergoing photorefractive keratectomy (PRK)

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

5: SMILE

Patients with normal corneas who are undergoing small incision lenticular extraction (SMILE)

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

6: CXL

Patients with keratoconus who are undergoing corneal cross-linking (CXL)

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Change in Brillouin Metrics
Tidsram: Difference between baseline and 3 months after intervention

Brillouin metrics to be evaluated include localized Mean Brillouin modulus measure across the cornea and at each depth of the corneal stroma

Difference between baseline and 3 months after intervention

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Samarbetspartners

Publikationer och användbara länkar

Den som ansvarar för att lägga in information om studien tillhandahåller frivilligt dessa publikationer. Dessa kan handla om allt som har med studien att göra.

Allmänna publikationer

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

1 juni 2021

Primärt slutförande (Beräknad)

1 juli 2027

Avslutad studie (Beräknad)

1 december 2027

Studieregistreringsdatum

Först inskickad

19 maj 2020

Först inskickad som uppfyllde QC-kriterierna

21 oktober 2020

Första postat (Faktisk)

22 oktober 2020

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

1 oktober 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

28 september 2026

Senast verifierad

1 september 2026

Mer information

Termer relaterade till denna studie

Ytterligare relevanta MeSH-villkor

Andra studie-ID-nummer

  • 20-355

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

NEJ

IPD-planbeskrivning

there is no plan to share data

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Ja

produkt tillverkad i och exporterad från U.S.A.

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .