Análisis biomecánico de la córnea mediante microscopía de Brillouin

28 de septiembre de 2026 actualizado por: James Bradley Randleman, The Cleveland Clinic

The objective of this study is to measure the Brillouin biomechanical properties in keratoconic corneas and characterize biomechanical alterations that occur after corneal procedures that inherently strengthen or weaken the cornea by evaluating the change in Brillouin metrics before and after treatments.

Descripción general del estudio

Estado

Inscripción por invitación

Intervención / Tratamiento

Descripción detallada

Surgical correction of myopia and keratoconus identification/management are separate but tightly intertwined issues of major significance. For both, there is an unmet need for direct measurements to evaluate corneal stiffness (i.e. its resistance to deformation). The prevalence of myopia is expected to double, affecting more than 50% of the US population, by 2050. Laser in situ keratomileusis (LASIK) is one of the most popular and successful surgeries in the world and compares favorably to long-term contact lens wear use for myopia correction. However, only ~10% of eligible patients undergo LASIK currently; the others cite safety concerns as a major factor in their decision. The primary risk for poor refractive surgery outcomes is biomechanical failure due to unidentified (subclinical) ectasia (i.e. keratoconus). Patients presenting for LASIK evaluation with atypical, suspicious corneal curvature but with undetermined true risk represent the leading reason for surgery screening failures. This results in good candidates being denied surgery, while up to 10% of truly poor candidates are still missed using current screening algorithms.

Keratoconus is up to 10 times more prevalent than the previously reported 1/2000 figure. Corneal cross-linking (CXL) is now FDA approved in the US for keratoconus treatment and is effective at stiffening the cornea and halting ectasia progression. Early identification of keratoconus is critical, but current tests in the clinic are morphological, not biomechanical, and therefore do not allow a definitive diagnosis at the earliest stages resulting in vision loss before CXL treatment is initiated. Thus, the need for accurate identification of subclinical ectasia has never been greater.

In the past years, newly developed technology, Brillouin microscopy, has emerged as the most promising tool to address this clinical need. This study will systemically address the critical gap in current knowledge by linking Brillouin mapping of corneal biomechanical alterations to abnormal morphological behavior and testing the findings in conditions where corneal biomechanics are abruptly altered, by: 1) weakening with refractive surgery procedures, and 2) strengthening through corneal cross-linking.

It is anticipated that a clinical tool assessing the mechanical state of the cornea will improve early diagnosis and management of keratoconus as well as refractive surgery planning. Ultimately, this will lead to predictive models where Brillouin measurements could be an accurate predictor of postoperative outcomes and thus aid in developing individualized surgical parameters.

Tipo de estudio

De observación

Inscripción (Estimado)

220

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

    • Ohio
      • Cleveland, Ohio, Estados Unidos, 44195
        • Cleveland Clinic Cole Eye Institute

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

18 años a 60 años (Adulto)

Acepta Voluntarios Saludables

No

Géneros elegibles para el estudio

Todos

Método de muestreo

Muestra no probabilística

Población de estudio

Group 1: Patients with normal corneas Group 2: Patients with keratoconus Group 3: Patients with normal corneas undergoing LASIK Group 4: Patients with normal corneas undergoing PRK Group 5: Patients with normal corneas undergoing SMILE Group 6: Patients with keratoconus undergoing CXL

Descripción

Inclusion Criteria:

  • patients aged 18-60 with keratoconus
  • patients aged 18-60 with normal corneas,
  • patients aged 18-60 undergoing refractive surgery (LASIK, PRK, SMILE)
  • patients aged 18-60 with keratoconus undergoing CXL

Exclusion Criteria:

  • outside age range
  • history of previous ocular surgeries
  • unable to cooperate for the Brillouin microscopic examination
  • unable to provide informed consent

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Modelos observacionales: CASE_CONTROL
  • Perspectivas temporales: CROSS_SECTIONAL

Cohortes e Intervenciones

Grupo / Cohorte
Intervención / Tratamiento
1: Normal Controls

Patients with normal corneas without any prior surgery to serve as the control group

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

2 Keratoconus

Patients with various stages of keratoconus

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

3: LASIK

Patients with normal corneas who are undergoing laser in situ keratomileusis (LASIK)

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

Group 4: PRK

Patients with normal corneas who are undergoing photorefractive keratectomy (PRK)

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

5: SMILE

Patients with normal corneas who are undergoing small incision lenticular extraction (SMILE)

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

6: CXL

Patients with keratoconus who are undergoing corneal cross-linking (CXL)

The Brillouin clinical instrument is comprised of three parts: a human interface, a laser-scanning confocal microscope, and an etalon-based spectrometer. The human interface is a modified ophthalmic slit-lamp instrument with chin support and headrest. The light source is a single longitudinal mode CW laser at 780 nm. A polarizing beam splitter and quarter-wave plate assembly sends the laser beam to the human interface. To focus light into the eye, a long-working distance microscope objective is used. Brillouin scattered light from the eye is collected with a single-mode optical fiber. For spectral analysis, a two-stage VIPA-etalon spectrometer configured with the cross-axis cascade principle and the spectrum is measured on a EM-CCD camera.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Change in Brillouin Metrics
Periodo de tiempo: Difference between baseline and 3 months after intervention

Brillouin metrics to be evaluated include localized Mean Brillouin modulus measure across the cornea and at each depth of the corneal stroma

Difference between baseline and 3 months after intervention

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Patrocinador

Colaboradores

Publicaciones y enlaces útiles

La persona responsable de ingresar información sobre el estudio proporciona voluntariamente estas publicaciones. Estos pueden ser sobre cualquier cosa relacionada con el estudio.

Publicaciones Generales

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

1 de junio de 2021

Finalización primaria (Estimado)

1 de julio de 2027

Finalización del estudio (Estimado)

1 de diciembre de 2027

Fechas de registro del estudio

Enviado por primera vez

19 de mayo de 2020

Primero enviado que cumplió con los criterios de control de calidad

21 de octubre de 2020

Publicado por primera vez (Actual)

22 de octubre de 2020

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

1 de octubre de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

28 de septiembre de 2026

Última verificación

1 de septiembre de 2026

Más información

Términos relacionados con este estudio

Términos MeSH relevantes adicionales

Otros números de identificación del estudio

  • 20-355

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

NO

Descripción del plan IPD

there is no plan to share data

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

Sí

producto fabricado y exportado desde los EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .