Denna sida har översatts automatiskt och översättningens korrekthet kan inte garanteras. Vänligen se engelsk version för en källtext.

Autus Valve tidig genomförbarhetsstudie

15 oktober 2026 uppdaterad av: Edwards Lifesciences

Autus storleksjusterbar ventil för kirurgisk lungklaffbyte tidig genomförbarhetsstudie

Prospective, single-arm, multi-center study to evaluate safety and effectiveness of the Autus Size-Adjustable Valve in pediatric patients aged 18 months to 16 years requiring surgical pulmonary valve replacement. The Autus Valve may be expanded pre-implant to match the subject's body size. Subjects will be evaluated prior to the Autus Valve implant procedure, immediately post-implant, at hospital discharge, 30 days, 6 months, and annually through 10 years.

The Autus Valve may be expanded post-implant via transcatheter balloon dilation to accommodate growth of the subject. In subjects who undergo a post-implant valve expansion, follow-up will continue for a minimum of 1 year after the post-implant valve expansion procedure.

Studieöversikt

Status

Aktiv, inte rekryterande

Studietyp

Interventionell

Inskrivning (Faktisk)

62

Fas

  • Inte tillämpbar

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

    • California
      • Los Angeles, California, Förenta staterna, 90027
        • Children's Hospital Los Angeles
    • Colorado
      • Aurora, Colorado, Förenta staterna, 80045
        • Children's Hospital Colorado
    • Georgia
      • Atlanta, Georgia, Förenta staterna, 30322
        • Children's Healthcare of Atlanta
    • Illinois
      • Chicago, Illinois, Förenta staterna, 60611
        • Ann and Robert H. Lurie Children's Hospital of Chicago
    • Massachusetts
      • Boston, Massachusetts, Förenta staterna, 02115
        • Boston Children's Hospital
    • Michigan
      • Ann Arbor, Michigan, Förenta staterna, 48109
        • University of Michigan - Mott Children's Hospital
    • New York
      • New York, New York, Förenta staterna, 10032
        • Columbia University Irving Medical Center/New York Presbyterian Morgan Stanley Children's Hospital
    • North Carolina
      • Durham, North Carolina, Förenta staterna, 27703
        • Duke University
    • Ohio
      • Columbus, Ohio, Förenta staterna, 43205
        • Nationwide Children's Hospital
    • Pennsylvania
      • Philadelphia, Pennsylvania, Förenta staterna, 19104
        • Children's Hospital of Philadelphia
    • Utah
      • Salt Lake City, Utah, Förenta staterna, 84113
        • University of Utah - Intermountain Primary Children's Hospital
    • Washington
      • Seattle, Washington, Förenta staterna, 98105
        • Seattle Children's Hospital

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

1 år till 16 år (Barn)

Tar emot friska volontärer

Nej

Kön som är behöriga för studier

Allt

Beskrivning

Inclusion Criteria:

Candidates must meet all of the following inclusion criteria to be considered for enrollment in this study.

  1. Age 18 months to 16 years.

  2. Male or female.

  3. Subject has a native or repaired right ventricular outflow tract.

  4. Subject has been recommended for surgical pulmonary valve replacement by treating clinical team (cardiologist and cardiac surgeon).

  5. Subject has at least one of the following echocardiographic findings:

    1. Severe pulmonary stenosis (defined as RV to PA peak instantaneous gradient ≥60 mmHg);
    2. Moderate or greater pulmonary regurgitation;
    3. Moderate or greater pulmonary stenosis plus moderate or greater pulmonary regurgitation.
  6. Subject's body size is suitable for implantation of a study device ranging from 12.7 to 22 mm (internal diameter).

  7. Subject and parent/legal representative, where appropriate, are willing to provide informed written consent.

  8. Subject and parent/legal representative, where appropriate, and treating physician agree that the subject will return for, and comply with, all required study assessments and follow-up visits.

Exclusion Criteria:

Candidates will be excluded from the study if any of the following conditions are met:

  1. Subject requires valve replacement in a non-pulmonary position.
  2. Subject has a prosthetic valve at other valve position or will need a prosthetic valve at other valve position (i.e., anticipate additional valve replacements needed within 3 years).
  3. Subject has pulmonary arterial hypertension (defined as mean PA pressure ≥25 mmHg).
  4. Subject has pulmonary atresia and major aortopulmonary collaterals.
  5. Subject has significant peripheral pulmonary artery stenosis.
  6. Subject has an active infection requiring current systemic antibiotic therapy (if temporary illness, subject may be a candidate 4 weeks after discontinuation of antibiotics).
  7. Subject has active endocarditis or a history of infective endocarditis.
  8. Subject has renal insufficiency as determined by a serum creatinine (S-Cr) level ≥2.5 mg/dL within 60 days prior to the Screening Visit, or has end-stage renal disease.
  9. Subject has leukopenia (defined as a white blood cell (WBC) count <3.5 x 103/µL)
  10. Subject has acute or chronic anemia (defined as hemoglobin (Hgb) <10.0 g/dl or 6 mmol/L).
  11. Subject has thrombocytopenia (defined as platelet count <50 x 103/µL.
  12. Subject has a known hypersensitivity to anticoagulants and antiplatelet drugs.
  13. Subject has a known autoimmune disease or receives immunosuppressant and/or immunostimulant drugs that the Investigator or Eligibility Screening Committee (ESC) believes may negatively affect study outcomes.
  14. Subject needs emergency cardiac or vascular surgery or intervention.
  15. Positive pregnancy test prior to valve implant procedure in female subjects who have had their first menses.
  16. Subject has, in the opinion of the Investigator, a life expectancy of less than 5 years.
  17. Subject or parent/legal representative refuses blood transfusions.
  18. Subject has medical, social or psychosocial factors that, in the opinion of the Investigator or ESC, could impact safety or compliance with study procedures.
  19. Subject is participating in an investigational study of a new drug, biologic or device at the time of study screening.

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Primärt syfte: Behandling
  • Tilldelning: N/A
  • Interventionsmodell: Enskild gruppuppgift
  • Maskning: Ingen (Open Label)

Vapen och interventioner

Deltagargrupp / Arm
Intervention / Behandling
Experimentell: Autus Valve Arm

Pulmonary Valve Replacement Surgery with the Autus Valve

Pulmonary valve replacement surgery with the Autus Valve

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Valve Implantation Primary Safety Endpoint
Tidsram: 30 Days post-valve implantation

Composite Endpoint: Freedom from a device-related complication through 30 days post-valve implantation as adjudicated by an independent Clinical Events Committee (CEC), including:

  • Death;
  • Valve thrombosis requiring intervention;
  • Symptomatic thromboembolism.
30 Days post-valve implantation
Valve Implantation Primary Effectiveness Endpoint
Tidsram: 6 Months post-valve implantation

Acceptable hemodynamic performance at 6 months post-valve implantation, as assessed by the Echocardiography Core Laboratory or Clinical Events Committee, defined as:

  • A mean RVOT gradient less than or equal to 40 mmHg and
  • Less than moderate pulmonary regurgitation by transthoracic echo and
  • No Autus Valve reintervention except for valve expansion to address prosthesis-patient mismatch.
6 Months post-valve implantation

Andra resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Other Safety Outcome - Composite
Tidsram: 6 months and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following other safety outcome will be evaluated at 6 months, and annually through 10 years post-valve implantation by an independent Clinical Events Committee. The composite outcome includes device-related:

  • Death;
  • Valve thrombosis requiring intervention;
  • Symptomatic thromboembolism.
6 months and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Mortality
Tidsram: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Mortality, including: all cause, cardiac-related, procedure-related and device-related

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Valve Thrombosis
Tidsram: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Valve thrombosis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Thromboembolism
Tidsram: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Thromboembolism

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Cardiac or Valve Intervention
Tidsram: 30 days, 6 months, and annually through a minimum of 510 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Other cardiac or valve intervention (surgical or transcatheter)

30 days, 6 months, and annually through a minimum of 510 years and up to 11 years post-valve implantation
Other Safety Outcome - Major Hemorrhage
Tidsram: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Major hemorrhage

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Endocarditis
Tidsram: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Endocarditis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Hemolysis
Tidsram: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Hemolysis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Device Explantation
Tidsram: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Device explantation

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Adverse Events (AE)
Tidsram: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be reported at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Adverse events

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Valve/Stent Fracture
Tidsram: pre-discharge, 12 months, and annually through a minimum of 5 years and up to 6 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-discharge and at 12 months post-valve expansion.

An independent Medical Reviewer will review post-implant chest x-rays to assess for valve/stent fracture. Chest x-rays will be performed at the following timepoints: post-valve implantation at hospital discharge, and annually through 5 years post-valve implantation. If a post-implant valve expansion procedure is performed, chest x-rays will be performed at hospital discharge and 12 months post-procedure.

pre-discharge, 12 months, and annually through a minimum of 5 years and up to 6 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-discharge and at 12 months post-valve expansion.
Other Effectiveness Outcome - Valve Implant Acute Procedural Success
Tidsram: Immediately Post-Valve Implant Procedure

To meet the definition of success, all the parameters listed below must be achieved, as assessed by Sites using direct pressure measurements and/or transesophageal echo (TEE):

  • RV to PA peak systolic pressure gradient <40 mmHg (direct pressure measurement);
  • None or trivial pulmonary regurgitation (assessed via TEE);
  • No paravalvular leak (assessed via TEE).
Immediately Post-Valve Implant Procedure
Other Effectiveness Outcome - Pulmonary Regurgitation Grade
Tidsram: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Pulmonary regurgitation grade categorized as none, trivial, mild, moderate, severe as evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Pulmonary Stenosis Grade
Tidsram: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Pulmonary stenosis grade categorized as none/trivial, mild, moderate, severe, as evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Right Ventricular Measures
Tidsram: Baseline, 30 days, 6 months, and annually through 5 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Right ventricular (RV) measures, including RV dimension and strain will be evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 5 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Left Ventricular Measures
Tidsram: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Left ventricular (LV) measures, including LV ejection fraction and strain will be evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Modified Ross Functional Classification
Tidsram: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Modified Ross Functional Class as assessed by the Site Investigators

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - PedsQL - Cardiac Module
Tidsram: Baseline, 6 months and annually through 10 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-valve expansion, and at 6 and 12 months post-valve expansion.

The Pediatric Qualify of Life Inventory (PedsQL) Cardiac Module will be administered by Site Personnel.

Baseline, 6 months and annually through 10 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-valve expansion, and at 6 and 12 months post-valve expansion.
Other Post-Implant Valve Expansion Safety Outcome - Composite
Tidsram: 30 days after each post-implant valve expansion procedure

Composite Endpoint: Freedom from device-related complication through 30 days post-valve expansion as adjudicated by an independent CEC:

  • Death;
  • Extracorporeal membrane oxygenation (ECMO) support;
  • Unplanned intervention (surgical or transcatheter);
  • Major hemorrhage
30 days after each post-implant valve expansion procedure
Other Post-Implant Valve Expansion Effectiveness Outcome - Acute Procedural Success
Tidsram: Immediately after each post-implant valve expansion procedure

To meet the definition of success, at least one of the parameters listed below must be achieved after each post-implant valve expansion is performed, assessed during the same catheterization procedure:

  • RV to aortic systolic pressure ratio: ≤0.5.
  • Peak-to-peak RV to PA systolic pressure gradient: ≤15 mmHg;
  • Peak-to-peak RV to PA systolic pressure gradient: ≥50% reduction (pre vs post valve expansion).
Immediately after each post-implant valve expansion procedure
Other Post-Implant Valve Expansion Effectiveness Outcome - Hemodynamic Performance
Tidsram: 6 months after each post-implant valve expansion procedure

Clinically acceptable hemodynamic performance at 6 months post-valve expansion procedure, as assessed by the Echo Core Laboratory via TTE, defined as:

  • Less than moderate pulmonary regurgitation (PR) by Doppler echocardiography;
  • Transvalvular gradient (assessed by CW Doppler) at 6-months is lower than the pre-expansion gradient.
6 months after each post-implant valve expansion procedure

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Publikationer och användbara länkar

Den som ansvarar för att lägga in information om studien tillhandahåller frivilligt dessa publikationer. Dessa kan handla om allt som har med studien att göra.

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

5 februari 2024

Primärt slutförande (Faktisk)

24 augusti 2026

Avslutad studie (Beräknad)

5 februari 2036

Studieregistreringsdatum

Först inskickad

23 juli 2021

Först inskickad som uppfyllde QC-kriterierna

12 augusti 2021

Första postat (Faktisk)

16 augusti 2021

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

20 oktober 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

15 oktober 2026

Senast verifierad

1 oktober 2026

Mer information

Termer relaterade till denna studie

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

NEJ

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Ja

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .