Wczesne studium wykonalności firmy Autus Valve

15 października 2026 zaktualizowane przez: Edwards Lifesciences

Wczesne studium wykonalności zastawki Autus z regulacją rozmiaru do chirurgicznej wymiany zastawki płucnej

Prospective, single-arm, multi-center study to evaluate safety and effectiveness of the Autus Size-Adjustable Valve in pediatric patients aged 18 months to 16 years requiring surgical pulmonary valve replacement. The Autus Valve may be expanded pre-implant to match the subject's body size. Subjects will be evaluated prior to the Autus Valve implant procedure, immediately post-implant, at hospital discharge, 30 days, 6 months, and annually through 10 years.

The Autus Valve may be expanded post-implant via transcatheter balloon dilation to accommodate growth of the subject. In subjects who undergo a post-implant valve expansion, follow-up will continue for a minimum of 1 year after the post-implant valve expansion procedure.

Przegląd badań

Status

Aktywny, nie rekrutujący

Typ studiów

Interwencyjne

Zapisy (Rzeczywisty)

62

Faza

  • Nie dotyczy

Kontakty i lokalizacje

Ta sekcja zawiera dane kontaktowe osób prowadzących badanie oraz informacje o tym, gdzie badanie jest przeprowadzane.

Lokalizacje studiów

    • California
      • Los Angeles, California, Stany Zjednoczone, 90027
        • Children's Hospital Los Angeles
    • Colorado
      • Aurora, Colorado, Stany Zjednoczone, 80045
        • Children's Hospital Colorado
    • Georgia
      • Atlanta, Georgia, Stany Zjednoczone, 30322
        • Children's Healthcare of Atlanta
    • Illinois
      • Chicago, Illinois, Stany Zjednoczone, 60611
        • Ann and Robert H. Lurie Children's Hospital of Chicago
    • Massachusetts
      • Boston, Massachusetts, Stany Zjednoczone, 02115
        • Boston Children's Hospital
    • Michigan
      • Ann Arbor, Michigan, Stany Zjednoczone, 48109
        • University of Michigan - Mott Children's Hospital
    • New York
      • New York, New York, Stany Zjednoczone, 10032
        • Columbia University Irving Medical Center/New York Presbyterian Morgan Stanley Children's Hospital
    • North Carolina
      • Durham, North Carolina, Stany Zjednoczone, 27703
        • Duke University
    • Ohio
      • Columbus, Ohio, Stany Zjednoczone, 43205
        • Nationwide Children's Hospital
    • Pennsylvania
      • Philadelphia, Pennsylvania, Stany Zjednoczone, 19104
        • Children's Hospital of Philadelphia
    • Utah
      • Salt Lake City, Utah, Stany Zjednoczone, 84113
        • University of Utah - Intermountain Primary Children's Hospital
    • Washington
      • Seattle, Washington, Stany Zjednoczone, 98105
        • Seattle Children's Hospital

Kryteria uczestnictwa

Badacze szukają osób, które pasują do określonego opisu, zwanego kryteriami kwalifikacyjnymi. Niektóre przykłady tych kryteriów to ogólny stan zdrowia danej osoby lub wcześniejsze leczenie.

Kryteria kwalifikacji

Wiek uprawniający do nauki

1 rok do 16 lat (Dziecko)

Akceptuje zdrowych ochotników

Nie

Płeć kwalifikująca się do nauki

Wszystko

Opis

Inclusion Criteria:

Candidates must meet all of the following inclusion criteria to be considered for enrollment in this study.

  1. Age 18 months to 16 years.

  2. Male or female.

  3. Subject has a native or repaired right ventricular outflow tract.

  4. Subject has been recommended for surgical pulmonary valve replacement by treating clinical team (cardiologist and cardiac surgeon).

  5. Subject has at least one of the following echocardiographic findings:

    1. Severe pulmonary stenosis (defined as RV to PA peak instantaneous gradient ≥60 mmHg);
    2. Moderate or greater pulmonary regurgitation;
    3. Moderate or greater pulmonary stenosis plus moderate or greater pulmonary regurgitation.
  6. Subject's body size is suitable for implantation of a study device ranging from 12.7 to 22 mm (internal diameter).

  7. Subject and parent/legal representative, where appropriate, are willing to provide informed written consent.

  8. Subject and parent/legal representative, where appropriate, and treating physician agree that the subject will return for, and comply with, all required study assessments and follow-up visits.

Exclusion Criteria:

Candidates will be excluded from the study if any of the following conditions are met:

  1. Subject requires valve replacement in a non-pulmonary position.
  2. Subject has a prosthetic valve at other valve position or will need a prosthetic valve at other valve position (i.e., anticipate additional valve replacements needed within 3 years).
  3. Subject has pulmonary arterial hypertension (defined as mean PA pressure ≥25 mmHg).
  4. Subject has pulmonary atresia and major aortopulmonary collaterals.
  5. Subject has significant peripheral pulmonary artery stenosis.
  6. Subject has an active infection requiring current systemic antibiotic therapy (if temporary illness, subject may be a candidate 4 weeks after discontinuation of antibiotics).
  7. Subject has active endocarditis or a history of infective endocarditis.
  8. Subject has renal insufficiency as determined by a serum creatinine (S-Cr) level ≥2.5 mg/dL within 60 days prior to the Screening Visit, or has end-stage renal disease.
  9. Subject has leukopenia (defined as a white blood cell (WBC) count <3.5 x 103/µL)
  10. Subject has acute or chronic anemia (defined as hemoglobin (Hgb) <10.0 g/dl or 6 mmol/L).
  11. Subject has thrombocytopenia (defined as platelet count <50 x 103/µL.
  12. Subject has a known hypersensitivity to anticoagulants and antiplatelet drugs.
  13. Subject has a known autoimmune disease or receives immunosuppressant and/or immunostimulant drugs that the Investigator or Eligibility Screening Committee (ESC) believes may negatively affect study outcomes.
  14. Subject needs emergency cardiac or vascular surgery or intervention.
  15. Positive pregnancy test prior to valve implant procedure in female subjects who have had their first menses.
  16. Subject has, in the opinion of the Investigator, a life expectancy of less than 5 years.
  17. Subject or parent/legal representative refuses blood transfusions.
  18. Subject has medical, social or psychosocial factors that, in the opinion of the Investigator or ESC, could impact safety or compliance with study procedures.
  19. Subject is participating in an investigational study of a new drug, biologic or device at the time of study screening.

Plan studiów

Ta sekcja zawiera szczegółowe informacje na temat planu badania, w tym sposób zaprojektowania badania i jego pomiary.

Jak projektuje się badanie?

Szczegóły projektu

  • Główny cel: Leczenie
  • Przydział: Nie dotyczy
  • Model interwencyjny: Zadanie dla jednej grupy
  • Maskowanie: Brak (otwarta etykieta)

Broń i interwencje

Grupa uczestników / Arm
Interwencja / Leczenie
Eksperymentalny: Autus Valve Arm

Pulmonary Valve Replacement Surgery with the Autus Valve

Pulmonary valve replacement surgery with the Autus Valve

Co mierzy badanie?

Podstawowe miary wyniku

Miara wyniku
Opis środka
Ramy czasowe
Valve Implantation Primary Safety Endpoint
Ramy czasowe: 30 Days post-valve implantation

Composite Endpoint: Freedom from a device-related complication through 30 days post-valve implantation as adjudicated by an independent Clinical Events Committee (CEC), including:

  • Death;
  • Valve thrombosis requiring intervention;
  • Symptomatic thromboembolism.
30 Days post-valve implantation
Valve Implantation Primary Effectiveness Endpoint
Ramy czasowe: 6 Months post-valve implantation

Acceptable hemodynamic performance at 6 months post-valve implantation, as assessed by the Echocardiography Core Laboratory or Clinical Events Committee, defined as:

  • A mean RVOT gradient less than or equal to 40 mmHg and
  • Less than moderate pulmonary regurgitation by transthoracic echo and
  • No Autus Valve reintervention except for valve expansion to address prosthesis-patient mismatch.
6 Months post-valve implantation

Inne miary wyników

Miara wyniku
Opis środka
Ramy czasowe
Other Safety Outcome - Composite
Ramy czasowe: 6 months and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following other safety outcome will be evaluated at 6 months, and annually through 10 years post-valve implantation by an independent Clinical Events Committee. The composite outcome includes device-related:

  • Death;
  • Valve thrombosis requiring intervention;
  • Symptomatic thromboembolism.
6 months and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Mortality
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Mortality, including: all cause, cardiac-related, procedure-related and device-related

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Valve Thrombosis
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Valve thrombosis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Thromboembolism
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Thromboembolism

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Cardiac or Valve Intervention
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 510 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Other cardiac or valve intervention (surgical or transcatheter)

30 days, 6 months, and annually through a minimum of 510 years and up to 11 years post-valve implantation
Other Safety Outcome - Major Hemorrhage
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Major hemorrhage

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Endocarditis
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Endocarditis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Hemolysis
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Hemolysis

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Device Explantation
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be evaluated by an independent Clinical Events Committee at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Device explantation

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Adverse Events (AE)
Ramy czasowe: 30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation

The following measure will be reported at 30 days, 6 months, and annually through 10 years post-valve implantation. Subjects that undergo valve expansion in year 10 after device implantation will be followed for an additional year.

• Adverse events

30 days, 6 months, and annually through a minimum of 10 years and up to 11 years post-valve implantation
Other Safety Outcome - Valve/Stent Fracture
Ramy czasowe: pre-discharge, 12 months, and annually through a minimum of 5 years and up to 6 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-discharge and at 12 months post-valve expansion.

An independent Medical Reviewer will review post-implant chest x-rays to assess for valve/stent fracture. Chest x-rays will be performed at the following timepoints: post-valve implantation at hospital discharge, and annually through 5 years post-valve implantation. If a post-implant valve expansion procedure is performed, chest x-rays will be performed at hospital discharge and 12 months post-procedure.

pre-discharge, 12 months, and annually through a minimum of 5 years and up to 6 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-discharge and at 12 months post-valve expansion.
Other Effectiveness Outcome - Valve Implant Acute Procedural Success
Ramy czasowe: Immediately Post-Valve Implant Procedure

To meet the definition of success, all the parameters listed below must be achieved, as assessed by Sites using direct pressure measurements and/or transesophageal echo (TEE):

  • RV to PA peak systolic pressure gradient <40 mmHg (direct pressure measurement);
  • None or trivial pulmonary regurgitation (assessed via TEE);
  • No paravalvular leak (assessed via TEE).
Immediately Post-Valve Implant Procedure
Other Effectiveness Outcome - Pulmonary Regurgitation Grade
Ramy czasowe: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Pulmonary regurgitation grade categorized as none, trivial, mild, moderate, severe as evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Pulmonary Stenosis Grade
Ramy czasowe: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Pulmonary stenosis grade categorized as none/trivial, mild, moderate, severe, as evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Right Ventricular Measures
Ramy czasowe: Baseline, 30 days, 6 months, and annually through 5 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Right ventricular (RV) measures, including RV dimension and strain will be evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 5 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Left Ventricular Measures
Ramy czasowe: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Left ventricular (LV) measures, including LV ejection fraction and strain will be evaluated by the Echo Core Laboratory.

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - Modified Ross Functional Classification
Ramy czasowe: Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.

Modified Ross Functional Class as assessed by the Site Investigators

Baseline, 30 days, 6 months, and annually through 10 years post-valve implant and 30 days, 6 months, and 12 months post-valve expansion.
Other Effectiveness Outcome - PedsQL - Cardiac Module
Ramy czasowe: Baseline, 6 months and annually through 10 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-valve expansion, and at 6 and 12 months post-valve expansion.

The Pediatric Qualify of Life Inventory (PedsQL) Cardiac Module will be administered by Site Personnel.

Baseline, 6 months and annually through 10 years post-valve implantation. Subjects that undergo post-implant valve expansion will be evaluated pre-valve expansion, and at 6 and 12 months post-valve expansion.
Other Post-Implant Valve Expansion Safety Outcome - Composite
Ramy czasowe: 30 days after each post-implant valve expansion procedure

Composite Endpoint: Freedom from device-related complication through 30 days post-valve expansion as adjudicated by an independent CEC:

  • Death;
  • Extracorporeal membrane oxygenation (ECMO) support;
  • Unplanned intervention (surgical or transcatheter);
  • Major hemorrhage
30 days after each post-implant valve expansion procedure
Other Post-Implant Valve Expansion Effectiveness Outcome - Acute Procedural Success
Ramy czasowe: Immediately after each post-implant valve expansion procedure

To meet the definition of success, at least one of the parameters listed below must be achieved after each post-implant valve expansion is performed, assessed during the same catheterization procedure:

  • RV to aortic systolic pressure ratio: ≤0.5.
  • Peak-to-peak RV to PA systolic pressure gradient: ≤15 mmHg;
  • Peak-to-peak RV to PA systolic pressure gradient: ≥50% reduction (pre vs post valve expansion).
Immediately after each post-implant valve expansion procedure
Other Post-Implant Valve Expansion Effectiveness Outcome - Hemodynamic Performance
Ramy czasowe: 6 months after each post-implant valve expansion procedure

Clinically acceptable hemodynamic performance at 6 months post-valve expansion procedure, as assessed by the Echo Core Laboratory via TTE, defined as:

  • Less than moderate pulmonary regurgitation (PR) by Doppler echocardiography;
  • Transvalvular gradient (assessed by CW Doppler) at 6-months is lower than the pre-expansion gradient.
6 months after each post-implant valve expansion procedure

Współpracownicy i badacze

Tutaj znajdziesz osoby i organizacje zaangażowane w to badanie.

Publikacje i pomocne linki

Osoba odpowiedzialna za wprowadzenie informacji o badaniu dobrowolnie udostępnia te publikacje. Mogą one dotyczyć wszystkiego, co jest związane z badaniem.

Daty zapisu na studia

Daty te śledzą postęp w przesyłaniu rekordów badań i podsumowań wyników do ClinicalTrials.gov. Zapisy badań i zgłoszone wyniki są przeglądane przez National Library of Medicine (NLM), aby upewnić się, że spełniają określone standardy kontroli jakości, zanim zostaną opublikowane na publicznej stronie internetowej.

Główne daty studiów

Rozpoczęcie studiów (Rzeczywisty)

5 lutego 2024

Zakończenie podstawowe (Rzeczywisty)

24 sierpnia 2026

Ukończenie studiów (Szacowany)

5 lutego 2036

Daty rejestracji na studia

Pierwszy przesłany

23 lipca 2021

Pierwszy przesłany, który spełnia kryteria kontroli jakości

12 sierpnia 2021

Pierwszy wysłany (Rzeczywisty)

16 sierpnia 2021

Aktualizacje rekordów badań

Ostatnia wysłana aktualizacja (Rzeczywisty)

20 października 2026

Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości

15 października 2026

Ostatnia weryfikacja

1 października 2026

Więcej informacji

Terminy związane z tym badaniem

Plan dla danych uczestnika indywidualnego (IPD)

Planujesz udostępniać dane poszczególnych uczestników (IPD)?

NIE

Informacje o lekach i urządzeniach, dokumenty badawcze

Bada produkt leczniczy regulowany przez amerykańską FDA

Nie

Bada produkt urządzenia regulowany przez amerykańską FDA

Tak

Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .

Subskrybuj