PROMISE Study: An Evaluation of an Implantable Continuous Glucose Sensor up to 180 Days (PROMISE)

July 25, 2024 updated by: Senseonics, Inc.

PROMISE Study: A Prospective, Multicenter Evaluation of Accuracy and Safety of an Implantable Continuous Glucose Sensor Lasting up to 180 Days

The purpose of this clinical investigation is to evaluate the accuracy of the Eversense® continuous Glucose Monitoring System (Eversense® 180 CGM System) measurements when compared with reference standard measurements up to 180 days of sensor use.

The investigation will also evaluate safety of the Eversense® 180 CGM System usage.

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Study Type

Interventional

Enrollment (Actual)

208

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • California
      • Concord, California, United States, 94520
        • John Muir Physician Network Clinical Research Center
      • Escondido, California, United States, 92025
        • AMCR Institute Inc.
      • Walnut Creek, California, United States, 94598
        • Diablo Clinical Research
    • Colorado
      • Aurora, Colorado, United States, 80045
        • Barbara Davis Center for Diabetes
    • Georgia
      • Atlanta, Georgia, United States, 30318
        • Atlanta Diabetes Care
    • Idaho
      • Idaho Falls, Idaho, United States, 83404
        • Rocky mountain Diabetes Center C/O Research Department
    • Texas
      • San Antonio, Texas, United States, 78229
        • Clinical Trials of Texas
    • Washington
      • Renton, Washington, United States, 98057
        • Rainier Clinical Research Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  1. Adult subjects, age ≥18 years
  2. Clinically confirmed diagnosis of diabetes mellitus for ≥1 year
  3. Subject has signed an informed consent form and is willing to comply with protocol requirements

Exclusion Criteria:

  1. History of unexplained severe hypoglycemia in the previous 6 months. Severe hypoglycemia is defined as hypoglycemia resulting in loss of consciousness or seizure
  2. History of diabetic ketoacidosis requiring emergency room visit or hospitalization in the previous 6 months
  3. Subjects with gastroparesis
  4. Female subjects of childbearing capacity (defined as not surgically sterile or not menopausal for ≥ 1 year) who are lactating or pregnant, intending to become pregnant, or not practicing birth control during the course of the study.
  5. A condition preventing or complicating the placement,operation, or removal of the Sensor or wearing of transmitter, including upper extremity deformities or skin condition.
  6. Symptomatic coronary artery disease; unstable angina; myocardial infarction, transient ischemic attack or stroke in the past 6 months; uncontrolled hypertension (systolic>160 mm Hg or diastolic >100 mm Hg at time of screening); current congestive heart failure; history of cardiac arrhythmia (benign PACs and PVCs allowed). Subjects with asymptomatic coronary artery disease (e.g. CABG, stent placement or angioplasty) may participate if negative stress test within 1 year prior to screening and written clearance from Cardiologist documented.
  7. Hematocrit <30% or >60%
  8. History of hepatitis B, hepatitis C, or HIV
  9. Current treatment for a seizure disorder unless written clearance by neurologist to participate in study
  10. History of adrenal insufficiency
  11. Currently receiving (or likely to need during the study period): immunosuppressant therapy; chemotherapy; anticoagulant/antithrombotic therapy (excluding aspirin); glucocorticoids (excluding ophthalmic or nasal). This exclusion does include the use of inhaled glucocorticoids and the use of topical glucocorticoids (over sensor site only); antibiotic for chronic infection (e.g. osteomyelitis, endocarditis)
  12. A condition requiring or likely to require magnetic resonance imaging (MRI)
  13. Known topical or local anesthetic allergy
  14. Known allergy to glucocorticoids
  15. Any condition that in the investigator's opinion would make the subject unable to complete the study or would make it not in the subject's best interest to participate in the study. Conditions include but are not limited to psychiatric conditions, known current or recent alcohol abuse or drug abuse by subject history, a condition that may increase the risk of induced hypoglycemia or risk related to repeated blood testing. Investigator will supply rationale for exclusion
  16. Participation in another clinical investigation (drug or device) within 2 weeks prior to screening or intent to participate during the study period
  17. The presence of any other active implanted device (as defined further in protocol)

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Continuous Glucose Monitoring Device
The Eversense® 180 CGM System
The Eversense® 180 CGM System
Other Names:
  • Eversense CGM system

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Effectiveness Measure - Mean Absolute Relative Difference (MARD) for Paired CGM and Reference Glucose Measurements
Time Frame: 180 days
The effectiveness endpoint is the mean absolute relative difference (MARD), calculated for paired CGM Sensor and reference glucose measurements through 180 days post-sensor insertion for reference glucose values from 40-400 mg/dL. MARD is defined as the average of absolute difference of paired Sensor and reference glucose readings divided by the reference glucose reading (reference) for reference glucose values from 40-400 mg/dL, that is: MARD = ((SUM | (Glucose)sensor - (Glucose)reference | / (Glucose)reference ) / n ) x 100%, where n is the total number of Sensor and reference glucose pairs after 180 days of sensor use (MARD is expressed as a percent). Lower MARDs indicate higher (better) accuracy.
180 days
Safety Endpoint - Incidence of Device-related or Sensor Insertion/Removal Procedure-related Serious Adverse Events
Time Frame: 180 days
Incidence of device-related or sensor insertion/removal procedure-related serious adverse events occurring up to 180 days after the insertion procedure.
180 days

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Satish Garg, MD, Barbara Davis Center

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

December 27, 2018

Primary Completion (Actual)

May 8, 2020

Study Completion (Actual)

May 8, 2020

Study Registration Dates

First Submitted

January 15, 2019

First Submitted That Met QC Criteria

January 16, 2019

First Posted (Actual)

January 17, 2019

Study Record Updates

Last Update Posted (Actual)

August 20, 2024

Last Update Submitted That Met QC Criteria

July 25, 2024

Last Verified

July 1, 2024

More Information

Terms related to this study

Other Study ID Numbers

  • CTP-0036

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Diabetes Mellitus, Type 2

Clinical Trials on Continuous Glucose Monitoring System