- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT02140905
Neck-Worn Monitoring Sensor for - A Study for Monitoring Subjects With Fluid-Management Issues During Dialysis
Validation of Neck-Worn Monitoring System - A Study for Monitoring Subjects With Fluid-Management Issues Using Repeated Measurements During Hemodialysis
The study has two objectives based on the intended-use measurements and populations described above. The first objective is to show that relative changes in impedance as measured by the test device (ΔSFI) are strongly correlated with the amount of fluid removed during dialysis (ΔF) for subjects with ESRD and possibly Heart Failure.
The second objective is to demonstrate correlation between ΔSFI and relative changes in impedance (ΔZ) as measured by the reference device.
Study Overview
Status
Conditions
Detailed Description
The investigators propose a study for validating the test device's measurement of a parameter called 'sternal fluid index' (SFI) in patients with diagnosed medical conditions related to fluid management, e.g. end-stage renal disease (ESRD), that are undergoing dialysis. The reference device for the study measures the averaged amplitude of a thoracic bio-impedance signal (Z) and a parameter called 'thoracic fluid content' (TFC), which is equal to 1000/Z. The test device measures SFI from a localized region near the sternum, whereas the reference device measures Z and TFC from the entire thoracic cavity. This means Z (or TFC) and SFI, if measured at approximately the same time, will likely correlate but have different absolute values. Thus, the statistical analysis will investigate the ability of SFI to track fluids, as well as the agreement between fluid-dependent trends displayed by SFI and Z (or TFC), as described below.
During the study the investigators will collect multiple samples, each containing a value of SFI from the test device, and values of TFC and Z from the reference device. The investigators will enroll approximately 20-50 subjects, with the goal of at least 20 subjects completing the study. The statistical analysis will include data from all subjects that complete the study. All subjects will undergo dialysis in a clinic. Approximately half to a third will have both ESRD and Heart Failure (HF). The investigators will measure a sample from each subject every 15 minutes during a standard dialysis session, which typically lasts about 3 to 4 hours. Levels of extracted fluids, determined by the dialysis machine, will be recorded at the time each sample is measured.
The statistical model compares time-dependent changes in SFI and Z values collected during dialysis. It is designed is to demonstrate: 1) that the test device is effective in characterizing fluid levels when used as intended; and 2) that the test and reference devices' measurements of trends in fluids are substantially equivalent, since they sense different physiological regions which are expected to yield different values of impedance.
Study Type
Enrollment (Actual)
Contacts and Locations
Study Locations
-
-
New Jersey
-
North Brunswick, New Jersey, United States, 08902
- Dialysis Center, Inc.
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- Subject is between 21 (twenty-one) and 90 (ninety) years of age at screening.
- Subject will have one or more of the following clinical diagnoses related to fluid-management issues: ESRD, Heart Failure.
- Subject and/or legally authorized person/representative is willing to undergo the Informed Consent process prior to enrollment in the study.
- Subject is a candidate for this study based on the PI's opinion and knowledge of the subject's condition.
Exclusion Criteria:
- Subject is participating in another clinical study that may affect the results of either study.
- Subject is unable or not willing to wear electrode patches as required.
- Subject has skin sensitivity to adhesive or hydrogel materials used in electrode patches.
- Subject has a pacemaker, implanted cardioverter-defibrillator, or ventricular assist device.
- Subject is considered by the PI to be medically unsuitable for study participation
Study Plan
How is the study designed?
Design Details
- Observational Models: Cohort
- Time Perspectives: Retrospective
Cohorts and Interventions
Group / Cohort |
|---|
|
Patients undergoing hemodialysis.
Subjects participating in the study
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Comparison of fluid measurement made with the Neck-Worn Monitoring System and those with an FDA approved fluid measurement device. Both measurements will be compared to the actual fluids removed during one dialysis session.
Time Frame: Measurements will be collected during one dialysis session. Each Session will last between 2 and 5.5 hours. All measurements will be completed within one month.
|
The study's primary objective is to determine the effectiveness of SFI measurements made by the Neck-Worn Monitoring System for characterizing fluid levels compared to an FDA approved device. Data collected includes SFI from the Neck-Worn Monitoring System and a reference device, fluid extracted during a dialysis session. Data will be collected every 15 minutes. |
Measurements will be collected during one dialysis session. Each Session will last between 2 and 5.5 hours. All measurements will be completed within one month.
|
Collaborators and Investigators
Sponsor
Investigators
- Principal Investigator: Matthew Banet, PhD, Baxter Healthcare Corporation
Study record dates
Study Major Dates
Study Start
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Estimate)
Study Record Updates
Last Update Posted (Estimate)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Additional Relevant MeSH Terms
Other Study ID Numbers
- PERM-IRB-0001
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Heart Failure
-
Umeå UniversityRegion NorrbottenNot yet recruitingHeart Failure | Diastolic Heart Failure | Systolic Heart FailureSweden
-
University of Health Sciences LahoreRecruitingAcute Decompensated Heart Failure | Heart Failure, Diastolic | Heart Failure, SystolicPakistan
-
Indiana UniversityRecruitingCongestive Heart Failure | Congestive Heart Failure (CHF) | Congestive Heart Failure Chronic | Congestive Heart Failure(CHF)United States
-
Manipal UniversityUnknownHeart Failure | Decompensated Heart Failure | Acute Heart Failure | Diastolic Heart Failure | Systolic Heart FailureIndia
-
Lakeland Regional Health Systems, Inc.RecruitingHeart Failure | Heart Failure Acute | Acute Heart Failure (AHF) | Heart Failure - NYHA II - IVUnited States
-
Tufts Medical CenterMetro West Medical CenterCompletedCongestive Heart Failure | Diastolic Heart Failure | Systolic Heart FailureUnited States
-
Abbott Medical DevicesCompletedHeart Failure | Heart Failure, Diastolic | Heart Failure, Systolic | Heart Failure NYHA Class II | Heart Failure NYHA Class III | Heart Failure With Reduced Ejection Fraction | Heart Failure NYHA Class IV | Heart Failure With Normal Ejection Fraction | Heart Failure; With Decompensation | Heart Failure...United States, Canada
-
VA Eastern Colorado Health Care SystemNational Institute on Aging (NIA)CompletedHeart Failure | Heart Failure, Diastolic | Heart Failure, Systolic | Heart Failure With Reduced Ejection Fraction | Heart Failure With Preserved Ejection Fraction | Heart Failure; With Decompensation | Heart Failure,Congestive | Heart Failure AcuteUnited States
-
Eli Lilly and CompanyNot yet recruitingHeart Failure | Heart Failure, Diastolic | Heart Failure, SystolicJapan, Netherlands, United States, Moldova, Romania
-
Wake Forest UniversityNational Institute on Aging (NIA)CompletedHeart Failure, Congestive | Diastolic Heart FailureUnited States