Cardiovascular Magnetic Resonance-Guided Radiofrequency-Ablation for Atrial Flutter II (2016-IMR)

February 18, 2019 updated by: University of Leipzig

Cardiovascular Magnetic Resonance-Guided Radiofrequency-Ablation for Atrial Flutter

The purpose of this clinical study is to assess the safety and performance of the Vision-MR Ablation Catheter, the primary safety endpoint will measure the rate of serious adverse events related to the device or procedure; the primary performance endpoint will measure the acute success rate defined as the demonstration of bidirectional cavo-tricuspid isthmus block after radiofrequency application in the cavo-tricuspid isthmus; the secondary performance endpoint will measure the chronic success rate defined as the freedom of recurrence of type 1 atrial flutter at three months post-ablation procedure. The study will be a single center study conducted at the Leipzig University Hospital Heart Center in Germany. The study population will consist of adult patients requiring ablation for type I atrial flutter.

This is the follow-up study to NCT02699034.

Study Overview

Detailed Description

The study will be a single center study conducted at the Leipzig University Hospital Heart Center in Germany. 35 subjects subjects meeting inclusion/exclusion criteria will participate in the study. This study requires the use of investigational products from two independent manufacturers: Imricor and Philips. Imricor has developed the ablation catheter with related accessory cables and electrophysiology (EP) recorder/stimulator system. Philips has developed an image guidance and mapping system that is compatible with the Imricor products.

The Vision-MR Ablation Catheter will be used in conjunction with the following investigational products: Vision-MR Ablation Cable Set, Vision-MR Diagnostic Cable, the interventional MRI Suite (iSuite) image guidance and mapping system, and a dStream Interface (dSIF-FE). With the exception of iSuite and dSIF-FE, which are manufactured by Philips, Imricor will provide all investigational devices used in the study.

Each procedure will require two single-use Vision-MR Ablation Catheters and one each of the two single-use accessory catheter cables.

Ablation for the treatment of arrhythmia is an inherently complex procedure. Use of the investigational products listed above under MR guidance is an emerging approach to performing the interventional electrophysiology procedure.

The study population will consist of adult patients requiring ablation for type I atrial flutter.

Study subjects will require a follow-up visit or telephone call at seven days post procedure. Accordingly, the expected total study duration is approximately 9 months with study start planned for February 2017.

This is the follow-up study to NCT02699034 which was terminated early based on a recommendation by the oversight authority. The study device (Vision-MR ablation catheter) has been improved and the EP recorder/stimulator is now CE-marked.

Study Type

Interventional

Enrollment (Actual)

36

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Saxony
      • Leipzig, Saxony, Germany, 04289
        • Heart Center Leipzig

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

16 years to 73 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • First time indication for ablation for type I atrial flutter.
  • Patients willing and able (mentally and physically capable per physician's discretion) to understand the investigational nature, potential risks and benefits of the study and able to provide written informed consent to participate in the study and agree to comply with follow-up visits and evaluation
  • Patients able to receive anticoagulation therapy to achieve adequate anticoagulation

Exclusion Criteria:

  • Contraindication for MRI diagnostic exam
  • A cardiac ablation or cardiac surgery within 180 days prior to enrollment
  • Documented intracardiac thrombus, tumor, bleeding, clotting or other abnormality that precludes catheter introduction and placement
  • Myocardial infarction within 60 days prior to enrollment
  • Current unstable angina
  • History of cerebrovascular event (within 180 days prior to enrollment)
  • Patients with an ejection fraction less than or equal to 35% within 90 day prior to enrollment
  • Permanent leads in or through the right atrium
  • Clinically significant structural heart disease (including tricuspid valve regurgitation, tricuspid valve stenosis or other congenital heart disease) that would preclude catheter introduction and placement, as determined by the Investigator
  • Uncompensated congestive heart failure (NYHA Class III or IV)
  • Arrhythmia is secondary to electrolyte imbalance, thyroid disease, or other reversible or non-cardiovascular cause
  • Known sensitivity to heparin or warfarin
  • Active or systemic infection
  • Any other significant uncontrolled or unstable medical condition (including but not limited to hypertension and diabetes)
  • contraindication for conventional ablation procedure know allergy against radiocontrast agents renal insufficiency with glomerular filtration rate < 30ml/min/1,73m2
  • Women who are pregnant or plan to become pregnant within the course of their participation in the investigation or who are breastfeeding
  • Life expectancy of less than 12 months
  • Patients with prosthetic valves
  • Contraindicated for transfemoral venous access
  • Older than 75 years
  • Current enrollment in any other clinical investigation

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: ablation for typical atrial flutter
This group receives an MR-guide ablation for atrial flutter with the study device ( Vision-MR ablation catheter )
Typical atrial flutter is treated by catheter ablation.
The study device is used inside the MRI environment to localize the cavotricuspid isthmus (CTI). The CTI is then treated with radiofrequency energy to achieve complete isthmus block.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of patients with bidirectional cavo-tricuspid isthmus block after intervention
Time Frame: 9 months
Acute success defined as the demonstration of the established electrophysiological endpoint (e.g. bidirectional cavo-tricuspid isthmus block after radiofrequency application in the cavo-tricuspid isthmus) with the investigational device. Historically, the acute success rate for RF ablation of type I atrial flutter is 85%. Analysis will be based on a binomial proportion and expressed as a percentage. For a total of N subjects with S achieving success, the percentage, represented as P, will be calculated as P = 100*S/N.
9 months
Number of patients with periinterventional serious adverse events
Time Frame: 9 months
The primary safety endpoint is the rate of serious adverse events (SAEs) related to the device or procedure assessed at the 7-day follow-up. Analysis will be based on a binomial proportion and expressed as a percentage. For a total of N subjects with S experiencing an SAE related to the device or procedure, the percentage, represented as P, will be calculated as P = 100*S/N.
9 months
Number of patients with freedom from recurrence from type I atrial flutter 3 months (chronic success rate) after ablation (study procedure).
Time Frame: 9 month
The primary chronic performance endpoint is the chronic success rate defined as freedom from recurrence of type 1 atrial flutter at 3 months post procedure. Freedom of recurrence of type I atrial flutter is assessed using symptom survey, ECG and echocardiogram data. This will be evaluated separately for all subjects treated with the device and for all subjects that achieved acute success (defined by the primary acute performance endpoint).
9 month

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of patients with freedom from recurrence from type I atrial flutter 6 months after ablation (study procedure).
Time Frame: 9 month
The secondary performance endpoint is the chronic success rate defined as freedom from recurrence of type 1 atrial flutter at 6 months post procedure. Freedom of recurrence of type I atrial flutter is assessed using symptom survey, ECG and echocardiogram data. This will be evaluated separately for all subjects treated with the device and for all subjects that achieved chronic success (defined by the primary chronic acute performance endpoint).
9 month

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Helpful Links

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

March 30, 2017

Primary Completion (Actual)

December 11, 2017

Study Completion (Actual)

January 15, 2019

Study Registration Dates

First Submitted

December 10, 2016

First Submitted That Met QC Criteria

March 8, 2017

First Posted (Actual)

March 14, 2017

Study Record Updates

Last Update Posted (Actual)

February 19, 2019

Last Update Submitted That Met QC Criteria

February 18, 2019

Last Verified

February 1, 2019

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Atrial Flutter

Clinical Trials on ablation for typical atrial flutter

3
Subscribe