A Novel Non-invasive Technique of Cerebral Compliance and Auto-regulation Assessment

September 29, 2022 updated by: Sergio Brasil, MD, University of Sao Paulo

Introduction: Intracranial pressure (ICP) monitoring is essential in several medical situations, however, currently there is an invasive technique, costly, not widely available and sometimes contraindicated. Transcranial Doppler (TCD) pulsatility index (PI) measure can provide indirect information on the cerebrovascular resistance (CVR) augmentation, which is present concomitantly with intracranial hypertension (ICH). The hypothesis that PI measure accurately indicates cerebral compliance impairment (CCI) has been not assessed by large studies currently, and would be of value as a non invasive technique to denote earlier installing of therapeutics to prevent the effects of ICH. Likewise, a novel technique of intracranial compliance assessment by means of an external sensor has been developed, still in need of being prospectively studied.

Objective: The present study aims to assess PI accuracy indicating CCI, and dynamic cerebral auto regulation (dCAR) during internal jugular veins (IJVs) compression observed by both invasive and non-invasive techniques.

Methods: A prospective, observational controlled study, including critical neurological patients with ICP monitoring in normal range (under 20 mmHg). Initially, dCAR is monitored, then, the IJVs are compressed for 60 seconds with ultrasound guidance. We evaluate optic nerve sheath prior to intervention, and dCAR, ICP values, ICP waveforms and PI variation at different times, correlating results.

Study Overview

Detailed Description

Introduction

Multimodal monitoring in neurocritical care is essential in current practice. It aims to early Identify treats occurring in the brain of comatose patients and implement targeted therapies to avoid further complications and poorer outcomes.

Some of the methods used for monitoring require invasive procedures, such as the assessment of intracranial pressure (ICP) with intracranial catheter. However, these techniques can raise morbidity, are costly, and are not available or contraindicated1. In addition, the invasive ICP method is not perfectly accurate with cases in which, intracranial hypertension (ICH) may be present despite normal screen ICP values2.

Transcranial Doppler (TCD) is a noninvasive bedside ultrasound technique which could be implemented in the monitoring of raised ICP3-4 and in the identification of impaired intracranial compliance by the compression of the internal jugular veins (IJVs)1. However, such application of the method has not been validate with large prospective studies. Validation of this technique would be of value for screening patients prior to suffer the consequences of ICH.

Recently, a non-invasive cranial strain gauge sensor has been developed, able to provide the curves of cardiac beat by beat cranial dilation. The sensor registers have been demonstrated to be similar to those obtained by the invasive ICP catheters, although large series comparing both techniques are lacking.

Objectives

1) Evaluate the reliability of TCD during the IJVs compression in the assessment of intracranial compliance; 2) evaluate dynamic cerebral auto regulation (dCAR) during IJVs compression. 3) Evaluate the suitability of novel non-invasive intracranial compliance system.

Methods

This is a prospective observational study which will be performed at "Hospital das Clinicas" (São Paulo University School of Medicine). The sample desired is of 50 consecutive subjects, as a pilot study based in recommendations of the literature5. After ethics committee approval, neurological patients of any cause, gender or age, whose underwent ICP monitoring displaying both ICP curves and values will be included. The non-invasive system of intracranial compliance monitoring (Brain4Care Brasil, São Carlos, Brazil) will be used concomitantly. We will exclude patients without insonation window.

The control group will be of healthy subjects, to evaluate PI and cerebral auto-regulation during IJVs compression.

Initially, the study of dynamic cerebral auto regulation through the relationship of spontaneous fluctuations of cerebral blood flow velocity in the middle cerebral arteries (MCAs) and arterial blood pressure, will be done using TCD (DWL Doppler Box 2 MHz, Germany) for five minutes, plus the optic nerve sheath measure (Sonosite Micromaxx 13 MHz, USA), then, ultrasound guided (Sonosite Micromaxx 13 MHz, USA) IJVs compression will be performed for 60 seconds in the included patients, and the following variables will be registered prior and after intervention: blood pressure, mean arterial blood pressure, heart rate, breathing rate, hemoglobin, hematocrit, carbon dioxide pressure (pCO2), oxygen saturation, systolic flow velocity, diastolic flow velocity, mean velocity, pulsatility index, ICP values and changes in the ICP relation P1-P2 (waveform behavior).

One researcher (SB) will perform TCD examinations and obtain patients data, while two blinded researchers (RN and MLO) the offline analysis.

Data will be presented as mean ± standard error of the mean or median and interquartile ranges depending on the nature of the variable. The Chi-square test (χ2) will be used to assess the difference in sex distribution and etiology between groups. To evaluate the homogeneity of the sample we will use the Lèvene test. And, we will use the Kolmogorov-Smirnov test to evaluate the normality of distribution of each variable studied. According to the homogeneity and normality of distribution (Gaussian) of the studied variable will be used the appropriate statistical test (parametric or non-parametric) for each variable. In the homogeneous and Gaussian variables during rest, the analysis of variance (ANOVA) of a factor will be used to test the differences between the groups. In cases of significant F, the Scheffé post-hoc test will be performed for multiple comparisons. In non-homogeneous and / or non-Gaussian variables during rest, the Kruskal-Wallis test will be used, followed by the Dunn's test between the groups. On the other hand, in the homogeneous and Gaussian variables analyzed during the physiological maneuvers, the analysis of variance (ANOVA) of two factors will be used to test the differences between groups. In cases of significant F, the Scheffé post-hoc test will be performed for multiple comparisons. However, in non-homogeneous and / or non-Gaussian variables analyzed during physiological maneuvers, the Wilcoxon (paired data) and Mann-Whitney (unpaired data) tests, when appropriate, will be used. Will be considered as statistically significant to differences with P values <0.05.

Study Type

Observational

Enrollment (Anticipated)

133

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

  • Name: Sergio Brasil, MDMSc
  • Phone Number: +5511981210990
  • Email: sbrasil@usp.br

Study Locations

      • São Paulo, Brazil, 06630-060
        • Recruiting
        • University of Sao Paulo
        • Contact:
          • Sergio Brasil
          • Phone Number: 1136287231

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Child
  • Adult
  • Older Adult

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Neurological patients (GCS 5 and under) of any cause, gender or age.

Description

Inclusion Criteria:

Patients whose underwent ICP monitoring displaying both ICP curves and values.

Exclusion Criteria:

Patients whose display absence of insonation window

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Case-Control
  • Time Perspectives: Prospective

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Impaired cerebral compliance
Time Frame: 10 minutes
Performing ICP and TCD monitoring simultaneously during jugular compression, matching waveform behavior through computer software QL Online
10 minutes

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Impaired cerebral auto-regulation
Time Frame: 10 minutes
The same procedure detailed above, matching mean arterial pressure and CBFv in the computer software QL Online
10 minutes
Non-invasive intracranial compliance monitoring
Time Frame: 10 minutes
The same approaches described above, using the non-invasive Brain4Care system concomitantly.
10 minutes

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

May 1, 2017

Primary Completion (Anticipated)

December 15, 2023

Study Completion (Anticipated)

December 31, 2023

Study Registration Dates

First Submitted

May 4, 2017

First Submitted That Met QC Criteria

May 4, 2017

First Posted (Actual)

May 8, 2017

Study Record Updates

Last Update Posted (Actual)

September 30, 2022

Last Update Submitted That Met QC Criteria

September 29, 2022

Last Verified

September 1, 2022

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Intracranial Hypertension

Clinical Trials on Transcranial Doppler (TCD) ultrasound monitoring

Subscribe