Functionality and Accuracy of the smART System in Real-Life ICU Settings

January 17, 2023 updated by: ART Medical Ltd.

Total participants in the "Functionality and Accuracy of the smART System in Real-Life ICU Settings" are 10 participants - in the current Jefferson study site there were 8 participants and in the Sheba study site (identifier NCT03689985) there were 2 participants.

This study is conducted in order to validate the functionality and accuracy of the smART™ System in a real-life ICU setting. The smART™ Feeding Tube System is a novel system with nasogastric tube developed by ART Medical (Healthcare) Ltd, based on sensor-lined tubes that transmit real-time information to an external console. The smART™ feeding tube is equipped with reflux sensors which alert when gastric contents regurgitating into the esophagus. In addition, the smART™ feeding tube is equipped with sensors designed to provide information about the location of the tube thus assisting in reducing the incident of misplacement during first positioning. The smART™ feeding tube is also automatically and in real-time stops feeding if the feeding tube moves out of position during ongoing use or detect gastric content in esophagus. Furthermore, smART™ Feeding Tube System can guide operator to correctly re-position the tube. Lastly the smART™ Feeding Tube System is equipped with an anti-reflux mechanism and automatic Gastric Residual Volume (GRV).

Study Overview

Status

Terminated

Intervention / Treatment

Study Type

Interventional

Enrollment (Actual)

10

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Pennsylvania
      • Philadelphia, Pennsylvania, United States, 19107
        • Thomas Jefferson University Hospital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Males and females 18 years or older
  • Patient has already been admitted to ICU
  • Patient requires enteral feeding
  • Patients receiving Proton Pump Inhibitors (PPI) therapy
  • Informed consent by independent physician and next of kin
  • ICU ventilated patients

Exclusion Criteria:

  • Patients with anomalies or diseases of the esophagus and or stomach.
  • Patients with known sensitivities or allergies to any of the feeding tube materials
  • Inability to place patient in semi-Fowler's position.
  • Any clinically significant abnormality upon physical examination which may, in the opinion of the investigator, pose difficulty in inserting the feeding tube (e.g. cervical spine disorder)
  • Pregnancy
  • Recent abdominal surgery (less than 30 days)

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: PREVENTION
  • Allocation: NA
  • Interventional Model: SINGLE_GROUP
  • Masking: NONE

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
EXPERIMENTAL: General
One arm study: smART Feeding Tube System.
Replacing generic feeding tube with smART Feeding Tube System.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of Participants for Whom the System is Able to Accurately Guide and Ensure Correct Initial Feeding Tube Placement , Verified by X-ray
Time Frame: Enrollment visit
Enrollment visit
Number of Tube "Out of Position" Alerts During Ongoing Use
Time Frame: Enrollment day up to 7 days
Any patient , in which the system alerted of tube out of position , the staff verified this according to the tube markings and if mispositioning occurred, this was documented the event in the CRF "out of position"
Enrollment day up to 7 days
Number of Automatic System Generated Feeding Stops in Response to Tube Displacement.
Time Frame: From insertion of feeding tube up to its removal , an average of 7 days.
In order to prevent aspiration, the system software automatically pinches the feeding tube. Verification of the functionality of automatic feeding cessation when tube displacement is detected was done as part of the formal verification of the system
From insertion of feeding tube up to its removal , an average of 7 days.
Number of Major Reflux Event That Lead to Automatically Stops Feeding
Time Frame: From insertion of feeding tube up to its removal , an average of 7 days.

Overfeeding is the intolerance of the stomach to feeding material that causes reflux. There are two types of reflux events as defined by the system; minor-reflux and massive-reflux. Both types of reflux events are detected by the level of fluids interacting with the sensors, the severity is defined by gastric content reaching above the lower esophageal sphincter (LES):

  1. Minor reflux is defined as gastric content that did not pass the balloon located in the center of the esophagus.
  2. Massive reflux is defined as gastric content that reaches immediately above the balloon.
From insertion of feeding tube up to its removal , an average of 7 days.
Average Duration of Balloon Inflation in Response to a Reflux Episode.
Time Frame: From insertion of feeding tube up to its removal , an average of 7 days.
Massive reflux of gastric content can aspirate into the lungs and cause pneumonia. The system detects the reflux event and will (according to an algorithm) stop the feeding and or inflate a balloon and evacuate gastric residuals from the stomach to lower the risk of aspiration.
From insertion of feeding tube up to its removal , an average of 7 days.

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Recording of Impedance Detected by the System and Correlation to Patient Positioning.
Time Frame: From insertion of feeding tube up to its removal , an average of 7 days.
Positions changes were correlated to reflux events captured by the system console. Any position change of the patient was captured by the video camera located above the patient's bed.
From insertion of feeding tube up to its removal , an average of 7 days.
Number of Participant With Change in Ventilator Associated Pneumonia (VAP) and/or Acute Respiratory Distress Syndrome (ARDS) as a Measure of Effectiveness in Reducing Aspiration of Gastric Contents
Time Frame: 56 - 168 hours per patient
The first and the last X-ray images obtained for patients during the study were used to evaluate Ventilator Associated Pneumonia (VAP) and/or Acute respiratory distress syndrome (ARDS)
56 - 168 hours per patient

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (ACTUAL)

October 26, 2017

Primary Completion (ACTUAL)

April 28, 2020

Study Completion (ACTUAL)

April 28, 2020

Study Registration Dates

First Submitted

June 12, 2017

First Submitted That Met QC Criteria

June 21, 2017

First Posted (ACTUAL)

June 26, 2017

Study Record Updates

Last Update Posted (ACTUAL)

February 13, 2023

Last Update Submitted That Met QC Criteria

January 17, 2023

Last Verified

July 1, 2021

More Information

Terms related to this study

Other Study ID Numbers

  • CLT3_SY_P001

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.