Performance of the Dexcom G6 Continuous Glucose Monitoring (CGM) System in Pregnant Women With Diabetes Mellitus

April 30, 2019 updated by: DexCom, Inc.
The purpose of this prospective observational study is to establish the performance and safety of the Dexcom G6 CGM System utilized for up to a 10-day wear period in pregnant women with diabetes mellitus.

Study Overview

Status

Unknown

Conditions

Intervention / Treatment

Detailed Description

Primary Study Objective:

To determine the accuracy of Dexcom G6 CGM System glucose readings in reference to arterialized venous sample YSI glucose measurements in pregnant women with diabetes mellitus.

Secondary Study Objective:

To assess the safety of the Dexcom G6 CGM System, through characterization of device-related Adverse Events (AEs), in pregnant women with diabetes mellitus.

Study Type

Observational

Enrollment (Anticipated)

60

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

Female

Sampling Method

Non-Probability Sample

Study Population

  1. Age ≥ 18 years;
  2. Currently in the 2nd or 3rd trimester of pregnancy;
  3. Diagnosis of GDM, T1DM, or T2DM;

Description

Inclusion Criteria:

  1. Age ≥ 18 years;
  2. Currently in the 2nd or 3rd trimester of pregnancy;
  3. Diagnosis of GDM, T1DM, or T2DM;
  4. Willing to avoid insulin injections or wear an insulin pump insertion set within 3 inches from the sensor site;
  5. Able to follow study procedures;
  6. Able to speak, read, and write in English or Spanish.

Exclusion Criteria:

  1. Extensive skin changes/diseases that preclude wearing the required number of devices on normal skin (e.g., extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis) at the proposed wear sites;
  2. Known allergy to medical-grade adhesives;
  3. Hematocrit of < 30%;
  4. Prescribed tocolytic drugs for treatment of pre-term labor or experiencing high-risk pregnancy complications, including, but not limited to, pre-eclampsia or HELLP syndrome (hemolysis, elevated liver enzymes, and a low platelet count) during current pregnancy;
  5. Currently receiving dialysis treatment or planning to receive dialysis during the sensor wear period;
  6. Require a Magnetic Resonance Imaging (MRI) scan, Computed Tomography (CT) scan, or diathermy during the sensor wear period;
  7. Current participation in another investigational study protocol (if a subject has recently completed participation in another drug study, the subject must have completed that study at least 30 days prior to being enrolled in this study.);
  8. Working for a competitive medical device company or having an immediate family member or person living within the same household who works for such a company (e.g., Medtronic, GlySens Inc., Abbott Laboratories, Roche, Senseonics, Waveform, and Ascencia/POCTech);
  9. Any condition that, in the opinion of the Investigator, would interfere with their participation in the trial (e.g., marked visual impairment) or pose excessive risk to study staff handling venous blood samples (e.g., known history of HIV or Hepatitis B or C).

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Cohort
  • Time Perspectives: Prospective

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Group
Device: Dexcom CGM System
Dexcom CGM System

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Dexcom G6 CGM Accuracy
Time Frame: 10 days
Accuracy of Dexcom G6 CGM System glucose readings in reference to arterialized venous sample YSI glucose measurements in pregnant women with diabetes mellitus.
10 days

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Dexcom G6 CGM Safety
Time Frame: 10 days
Assess the safety of the Dexcom G6 CGM System, through characterization of device-related Adverse Events (AEs), in pregnant women with diabetes mellitus.
10 days

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Study Director: Stayce Beck, DexCom, Inc.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Anticipated)

May 1, 2019

Primary Completion (Anticipated)

May 31, 2019

Study Completion (Anticipated)

May 31, 2019

Study Registration Dates

First Submitted

April 30, 2019

First Submitted That Met QC Criteria

April 30, 2019

First Posted (Actual)

May 2, 2019

Study Record Updates

Last Update Posted (Actual)

May 2, 2019

Last Update Submitted That Met QC Criteria

April 30, 2019

Last Verified

April 1, 2019

More Information

Terms related to this study

Other Study ID Numbers

  • PTL-903652

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Diabetes

Clinical Trials on Dexcom CGM System

3
Subscribe