Evaluating Safety and Efficacy of Umbulizer in Patients Requiring Intermittent Positive Pressure Ventilation

August 5, 2020 updated by: Umbulizer LLC
Randomized, control crossover study comparing Umbulizer's efficacy to traditional mechanical ventilators

Study Overview

Detailed Description

Mechanical ventilators cost tens of thousands of dollars, rendering them difficult for low and middle income countries to procure. Given chronic shortages of these devices in places like Pakistan, patients are often ventilated manually with Ambu Bags for hours or days. This method is not suitable for long term ventilation and is extremely life threatening.

Umbulizer seeks to reduce preventable deaths due to ventilator scarcity by providing a reliable alternative. The device is accurate, low-cost, and portable, and is an ideal solution for treating patients with respiratory diseases in low-resource settings. The primary objective of this study is to test efficacy and determine non-inferiority of this device in comparison to ventilation provided by traditional ventilators. Secondary objectives of this study include:

  1. Testing safety of the device for use over a predetermined period of time
  2. Evaluating patient comfort during ventilation with this device
  3. Assessing ease of use of this device for doctors and medical staff

Study Type

Interventional

Enrollment (Actual)

24

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Punjab
      • Lahore, Punjab, Pakistan
        • Services Hospital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 85 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

1) Patients needing mechanical ventilation

Exclusion Criteria:

  1. Patients with significant airway resistance (i.e. severe chronic obstructive pulmonary disease or asthma) where clinical investigator is concerned about gas trapping or barotrauma
  2. Patients with very low lung compliance (i.e. severe acute respiratory distress syndrome or restrictive disease) where clinical investigator is concerned about problems delivering sufficient tidal volume or end expiratory pressure
  3. Patients who experienced a myocardial infarction within the last 6 weeks
  4. Patients who are determined by a clinical investigator to be too hemodynamically unstable to be enrolled

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Supportive Care
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: Single

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Other: Sequence A
If patient is randomly assigned to Sequence A, they will be placed on a traditional ventilator for the first period. This period will be 1 hour in duration. Delivered tidal volume, respiratory rate (RR), and peak pressure will be recorded throughout the period using the ventilator. Heart rate (HR), Blood Pressure (BP), and oxygen (O2) saturation measurements will be recorded every 5 minutes. Arterial blood gas (ABG) data will be taken at t = 1 hour if the patient is stable on the traditional ventilator. ABGs may be taken more frequently if the patient is destabilizing. Next, the study participant will be transitioned to the Umbulizer to begin Period 2. This period will also be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using Umbulizer. HR, BP, O2 saturation measurements will be taken every 5 minutes. ABG data will be taken at t = 30 minutes and 1 hour.
Umbulizer is a reliable, user-friendly, and portable system that provides consistent air to the patient by quantitatively delivering a pre-determined tidal volume, inhale/exhale (I/E) ratio, peak pressure, and breathing rate (BPM). It can be configured to deliver either controlled ventilation or assist-controlled ventilation.
A medical device designed to artificially perform breathing functions for patients who are unable to adequately perform these functions on their own.
Other: Sequence B
If patient is randomly assigned to Sequence B, they will be shifted to the Umbulizer for the first period. This period will be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using Umbulizer. HR, BP, O2 saturation measurements will be recorded every 5 minutes. ABG data will be taken at t = 30 minutes and 1 hour. Next, the study participant will be transitioned to a traditional ventilator to begin Period 2. This period will also be 1 hour in duration. Delivered tidal volume, RR, and peak pressure will be recorded throughout the period using the ventilator. HR, BP, O2 saturation measurements will be recorded every 5 minutes. ABG data will be taken at t = 1 hour if the patient is stable on the traditional ventilator. ABGs may be taken more frequently if the patient is destabilizing.
Umbulizer is a reliable, user-friendly, and portable system that provides consistent air to the patient by quantitatively delivering a pre-determined tidal volume, inhale/exhale (I/E) ratio, peak pressure, and breathing rate (BPM). It can be configured to deliver either controlled ventilation or assist-controlled ventilation.
A medical device designed to artificially perform breathing functions for patients who are unable to adequately perform these functions on their own.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Partial Pressure of Carbon Dioxide (PaCO2)
Time Frame: 2 hours
PaCO2 in mmHg at t = 1 hr (from Arterial Blood Gas or ABG readings) when patient is on Umbulizer compared to PaCO2 at t = 1 hr when patient is on a traditional mechanical ventilator
2 hours
pH
Time Frame: 2 hours
pH at t = 1 hr (from ABG readings) when patient is on Umbulizer compared to pH at t = 1 hr when patient is on a traditional mechanical ventilator
2 hours

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of Adverse Events
Time Frame: 2 hours
Number of adverse events on Umbulizer compared to on a traditional ventilator
2 hours

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Level of Comfort
Time Frame: 2 hours
Level of comfort on Umbulizer will be determined based on a 4-point scale assessing a combination of indicators (1 point will be assigned for the absence of each of the following: substantial deviation of heart rate from baseline, substantial deviation of blood pressure from baseline, grimacing, diaphoresis).
2 hours
Ease of Use
Time Frame: 2 hours
Umbulizer's ease of use will be determined based on survey responses by doctors/medical staff at the end of each experiment
2 hours
Partial Pressure of Oxygen (PaO2)
Time Frame: 2 hours
PaO2 in mmHg at t = 1 hr (from ABG readings) when patient is on Umbulizer compared to PaO2 at t = 1 hr when patient is on a traditional mechanical ventilator
2 hours
O2 Saturation
Time Frame: 2 hours
O2 saturation (from pulse oximeter) when patient is on Umbulizer compared to when patient is on a traditional mechanical ventilator
2 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Kamran Chima, MBBS, Services Hospital

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

September 4, 2018

Primary Completion (Actual)

November 30, 2019

Study Completion (Actual)

December 31, 2019

Study Registration Dates

First Submitted

July 9, 2019

First Submitted That Met QC Criteria

July 21, 2019

First Posted (Actual)

July 23, 2019

Study Record Updates

Last Update Posted (Actual)

August 6, 2020

Last Update Submitted That Met QC Criteria

August 5, 2020

Last Verified

August 1, 2020

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

Undecided

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Respiratory Failure

Subscribe