FreeStyle Libre Plus Bluetooth Transmitter Adjunct: Can This Improve Glucose Accuracy and Reduce Burden of Hypoglycaemia (FSL-M)

February 8, 2021 updated by: Dasman Diabetes Institute

Crossover Randomized Controlled Trial of FreeStyle Libre Plus Bluetooth Transmitter Compared to FreeStyle Libre Alone in People With Type 1 Diabetes Mellitus. Can This Improve Accuracy and Reduce Burden of Hypoglycaemia

Advanced glucose monitoring systems have revolutionized diabetes care and enabled people with diabetes to achieve better diabetes control with reduced risk of hypoglycaemia. Continuous glucose monitoring (CGM) systems provide real-time glucose monitoring and alarms when glucose approaches extreme readings (hypoglycaemia and hyperglycaemia) or when the change in glucose is rapid. All available CGM systems, except Dexcom G6, require daily calibrations with capillary glucose readings in order to attain accuracy of glucose readings. Decom G6 system is not widely accessible and only available in certain countries. Flash glucose monitoring systems (Flash) provide glucose readings when users actively scan their sensors. FreeStyle Libre (FSL) is the only Flash glucose monitoring system currently available in market. FSL is factory calibrated and sensors are ready to use after placement and initiation. The two main differences between Flash and CGM are user interaction and the alarm facility. While CGM provide real-time glucose readings, Flash is user-dependent for actively scanning and understanding the readings. Moreover, CGM systems provide alarms for low or high glucose and for rapid glucose changes, while Flash does not routinely provide alarms. This is particularly relevant when patients have impaired or lost hypoglycaemia awareness. CGM systems are costlier compared to Flash, which has contributed to the wider adoption of FSL.

Several Bluetooth adjuncts have been introduced to market for FSL. These devices attach to Libre sensor and connect to the user's mobile phone via Bluetooth. This enables continuous and real-time feed of glucose readings from the sensor to patient's mobile phone, which enables a wide range of customizable alarms for high and low glucose levels and for rapid glucose changes. This setup also enables calibration of Libre sensor with capillary glucose which, anecdotally, has been reported to improve sensor accuracy. None of these adjuncts have been validated clinically. FSL with Bluetooth adjunct such as MiaoMiao remain cheaper than current CGM options and could be more accessible in some countries than CGM. However, without robust evidence to support effectiveness and safety of such setup it is not possible to recommend this.

The Objective of this study is to determine whether FSL with Bluetooth Adjunct is superior to FSL alone in accuracy and reduction of hypoglycaemia burden.

Study Overview

Detailed Description

Study design:

Open-label, crossover, single center, randomized controlled trial of FreeStyle Libre with MiaoMiao Bluetooth transmitter adjunct with calibrations (FSL-M) compared to FreeStyle Libre alone (FSL-A) in people with T1DM.

Protocol summary:

Following informed written consent, subjects will be randomized to either FSL-M group using FSL with Miaomio and Tomato App premium version, or FSL-A with Libre Link app (stage 1). Subjects will spend 8 weeks in the assigned group followed by 2 weeks washout then subjects will switch over to the other group for further 8 weeks (stage 2).

While in FSL-M, subjects will be asked to calibrate FSL with daily capillary glucose values using Contour Plus, or on days 1, 2, 4, 8, 13 of sensor life. Alarms for low glucose (<3.9mmol/L, <70mg/dL) and high glucose (>13.9mmol/L, 250mg/dL) will be set in Tomato App.

7-point capillary glucose profiles will be carried out by subjects in both groups on days 2/3, 5/6, 9/10, 13/14 of each sensor life to enable paired comparison with sensor values.

Subjects will have routine labs at randomization, end of stage 1, and end of stage 2. All subjects will be asked to complete the following questionnaires: GOLD, Problem Areas in Diabetes (PAID) and Diabetes Quality of Life (DQOL) at randomization, at end of stage1 and end of stage 2. The group randomized to FSL-M at stage 1 will have qualitative interview at randomization, end of stage 1 and end of stage 2.

All adverse events will be collected including, but not limited to, all hypoglycaemia, diabetic ketoacidosis, hospital admissions for any reason, and device related skin reaction for the duration of the study in all subjects in both groups.

Data analysis will be carried out on glucose data retrieved from Libre view web-platform and Tomato App downloads in addition to Contour Plus meter downloads.

Study Type

Interventional

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Kuwait City, Kuwait, 15462
        • Ali Aldibbiat

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

21 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  1. T1DM for 12 months or more
  2. Age >21 years old
  3. Established user of FreeStyle Libre for ≥3 months
  4. HbA1C <12% (108mmol/mol)
  5. Able to check glucose with finger prick several times a day

Exclusion Criteria:

  1. Pregnancy or plans for pregnancy over the duration of the study
  2. Breast feeding
  3. Renal impairment with eGFR <30ml/minute/1.73m2
  4. Significant anaemia (males Hb<110g/L, females Hb <100g/L)
  5. Active malignancy
  6. Significant vision impairment
  7. Any significant illness- decision will be made by the PI or Co-I
  8. Not able to attend study visits
  9. Not able or not willing to participate in all study components
  10. Not able or not willing to give written informed consent

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Supportive Care
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: FSL-M
FreeStyle Libre + MiaoMiao Bluetooth adjunct with mobile application
Use of FreeStyle Libre with MiaoMiao Bluetooth adjunct and mobile device application to enable calibration and alarms
Other Names:
  • FSL-M
Active Comparator: FSL-A
FreeStyle Libre alone
Use of standard FreeStyle Libre with Libre Link app and mobile phone scanning
Other Names:
  • FSL-A

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Mean absolute relative difference (MARD)
Time Frame: 9 months
Comparing difference between glucose readings from the sensor in comparison to reference paired capillary glucose readings. This is to enable assessing accuracy of the intervention with FSL and MiaoMiao
9 months
Percentage time below 3.0mmol/L (54mg/dL)
Time Frame: 9 months
This is to enable evaluating burden of hypoglycaemia. Analysis will be done on sensor glucose data
9 months
Percentage time in range 4-10mmol/L
Time Frame: 9 months
This is to enable effect on time spent in normoglycaemia. Analysis will be done on sensor glucose data
9 months

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percentage time <4mmol/L (72mg/dL)
Time Frame: 9 months
to evaluate general hypoglycaemia
9 months
Percentage time >10mmol/L
Time Frame: 9 months
To evaluate hyperglycaemia
9 months
Change in HbA1C
Time Frame: 9 months
To evaluate for any change in overall diabetes control
9 months
Rate of hypoglycaemia
Time Frame: 9 months
number of episodes of hypoglycaemia
9 months
Rate of severe hypoglycaemia
Time Frame: 9 months
number of episodes of severe hypoglycamia
9 months
Rate of device-related complications
Time Frame: 9 months
This is specific for Libe and/or MiaoMiao
9 months
Change in dose of insulin
Time Frame: 9 months
This will compare before and at end of each phase total daily insulin dose
9 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Ali Aldibbiat, MD PhD FRCP, Dasman Diabetes Institute

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Anticipated)

October 14, 2020

Primary Completion (Anticipated)

April 30, 2021

Study Completion (Anticipated)

August 31, 2021

Study Registration Dates

First Submitted

October 30, 2019

First Submitted That Met QC Criteria

October 30, 2019

First Posted (Actual)

November 1, 2019

Study Record Updates

Last Update Posted (Actual)

February 10, 2021

Last Update Submitted That Met QC Criteria

February 8, 2021

Last Verified

October 1, 2019

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Type1 Diabetes Mellitus

Clinical Trials on FreeStyle Libre with MiaoMiao

Subscribe