Comparison of Intraocular Pressure Measurements Between Reichert Tono-Vera Tonometer and Goldmann Tonometry

December 30, 2021 updated by: Reichert, Inc.
Determine if the Tono-Vera Tonometer accurately measures intraocular pressure (IOP). The hypothesis of this test is to confirm the Tono-Vera Tonometer is equivalent to the Goldmann Applanation Tonometer (within +/- 5.0 mmHg).

Study Overview

Detailed Description

The Tono-Vera Tonometer (Reichert, Inc.) is a portable rebound tonometer intended to measure IOP. Subjects will undergo a standard ophthalmologic examination. Subjects will then be measured with four tonometers: Goldmann Applanation Tonometer (C.S.O. SRL), Ocular Response Analyzer G3 (Reichert, Inc.), iCare ic100 tonometer (iCare Finland Oy), and Tono-Vera Tonometer.

Study Type

Interventional

Enrollment (Actual)

61

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • North Carolina
      • Durham, North Carolina, United States, 27705
        • Department of Ophthalmology, Duke University

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

14 years to 86 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Subjects must be male or female, between the ages of 18 and 90 years old;
  • Be able and willing to provide signed informed consent
  • Be able to follow study instructions

Exclusion Criteria:

  • Subjects with only one functional eye;
  • Subjects with one eye having poor or eccentric fixation;
  • Subjects with central corneal thickness greater than 600 µm or less than 500 µm (about 2 standard deviations the human mean);
  • Subjects with corneal scarring or who have had corneal surgery, including corneal laser surgery;
  • Subjects with concomitant ocular diseases such as: microphthalmos, buphthalmos, nystagmus, keratoconus, severe dry eye syndrome, blepharospasm, any other corneal or conjunctival pathology or infection;
  • Contact lens wearers;
  • Known allergy to proparacaine or fluorescein as these are used to anesthetize the eye and allow IOP measurement, respectively, when used with the GAT.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: Non-Randomized
  • Interventional Model: Sequential Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Active Comparator: Goldmann Applanation Tonometer
Measurement of IOP with Goldmann Applanation Tonometer. All subjects will participate in this arm.
Measurement of intraocular pressure (IOP) with Goldmann Applanation Tonometer. Measurement will be used to categorize each subject as having Low IOP (7 to 16 mmHg), Medium IOP (>16 to <23 mmHg), or High IOP (>23 mmHg).
Active Comparator: ORA G3 and ic100
Measurement of IOP with Ocular Response Analyzer G3 and ic100 tonometers. All subjects will participate in this arm.
Measurement of IOP with Ocular Response Analyzer G3 and ic100 tonometers.
Experimental: Tono-Vera Tonometer
Measurement of IOP with Tono-Vera Tonometer. All subjects will participate in this arm.
Measurement of IOP with Tono-Vera Tonometer

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Demonstrate compliance with ANSI Z80.10-2014-Ophthalmics-Ophthalmic Instruments-Tonometers.
Time Frame: Through study completion, approximately 4 months.
Subjects will be distributed into Low IOP, Medium IOP, and High IOP groups based on Goldmann Applanation Tonometer measurements. The Tono-Vera Tonometer measurements will be within +/-5.0 mmHg of the Goldmann Applanation Tonometer measurements in all groups.
Through study completion, approximately 4 months.

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Data collection for two measurement modes.
Time Frame: Through study completion, approximately 4 months.
Tono-Vera has two measurement modes: A 6-measurement mode & "Quick" measurement mode. The 6-measurement mode will be used during this study. It will be possible to post-process the results to determine the IOP value that would have been obtained using Quick mode. As such the sponsor will be able to ensure the accuracy of both measurement modes compared to Goldmann Tonometry.
Through study completion, approximately 4 months.
Data collection for device calibration.
Time Frame: Through study completion, approximately 4 months.
The study will collect data on a large number of patients over a wide range of IOP values. The sponsor may use the collected data in a post-hoc analysis to verify that the Tono-Vera Tonometer factory calibration is optimal.
Through study completion, approximately 4 months.

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Henry Tseng, MD, PhD, Department of Opthalmology, Duke University

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

October 19, 2020

Primary Completion (Actual)

April 28, 2021

Study Completion (Actual)

April 28, 2021

Study Registration Dates

First Submitted

March 16, 2020

First Submitted That Met QC Criteria

April 21, 2020

First Posted (Actual)

April 24, 2020

Study Record Updates

Last Update Posted (Actual)

January 3, 2022

Last Update Submitted That Met QC Criteria

December 30, 2021

Last Verified

December 1, 2021

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • 16305-TPR-28

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Glaucoma

Clinical Trials on IOP with Goldmann Applanation Tonometer

Subscribe