Acute Hemodynamic Effects of Pacing the His Bundle in Heart Failure (HEPA-His)

November 16, 2023 updated by: Region Skane

Acute Hemodynamic Effects of His-bundle Pacing in Bi-Ventricular Pacing Non-responders (The HEPA-His Trial)

Cardiac resynchronization therapy (CRT) with biventricular pacing (BVP) is indicated for patients with low ejection fraction and persistent symptoms of heart failure despite medical therapy and a wide QRS-complex, preferably with a left bundle branch block. Unfortunately, up to 40 % receiving CRT do not respond with subjective improvement or improved left ventricular ejection fraction from the treatment.

There are few therapeutic alternatives in patients not responding to CRT with BVP, and the potential to improve cardiac function in these patients could make a large difference in terms of life quality and prognosis.

His-bundle pacing (HBP) can be a suitable alternative, but is likely not possible in all patients with failed CRT. Furthermore, upgrading to HBP does require a surgical procedure and therefore exposes the patients to risk of complications (e.g. infection of the device). Therefore, early identification of the patients who are likely to benefit from an upgrade to HBP would be beneficial.

This study aims to evaluate if patients with failed CRT treatment will benefit from hemodynamic improvement if they are treated with temporary HBP, without opening the device-pocket, and exposing the existing pacemaker-device to a risk of infection. A temporary pacemaker lead and pacing sheath will be used, and placed in the heart via a blood vessel on the right side of neck, using local anaesthesia. Acute effects on the pump function of the heart will be measured during rest and during supine bicycle test (the equivalent of moderate physical activity). Factors associated with a beneficial effect will be evaluated on a group basis.

If the individual patient experiences an improvement of cardiac function during the test, he/she will be offered an upgrade to HBP and a replacement of the existing device.

Study Overview

Detailed Description

Primary hypothesis:

His-bundle pacing will lead to a better stroke volume at rest compared to biventricular pacing.

Primary outcome measure:

Stroke-volume (ml) at rest assessed by invasive hemodynamic measurements with the thermodilution method using a Swan-Ganz catheter.

Recruitment:

Patients will be recruited from the dedicated CRT outpatient clinic at Skane University Hospital. This outpatient clinic evaluates all CRT recipients at 6 months after device implant. If the patient is then a "non-responder" by subjective and objective measures, it is unlikely that there will be any further benefit from the CRT device. Those patients fulfilling the inclusion criteria but not the exclusion criteria will therefore be invited to participate in the study. Written information will be provided to all patients prior to informed consent is signed.

Procedure:

All patients will undergo a temporary His-bundle stimulation and hemodynamic evaluation. The procedure takes place at the dedicated laboratory for advanced hemodynamic evaluation at Skåne University Hospital in Lund. The first option for vascular access is the right external jugular vein. Under sterile conditions, ultrasound guided venous puncture will be performed, using standard Seldinger technique. A 7F catheter (Medtronic C315) will then be placed over a long guidewire, and used to direct a dedicated pacemaker lead (Medtronic Select Secure 3830) to the His bundle. In case of technical difficulties, a steerable 9F catheter will be used instead (Medtronic C304 deflectable). If HBP is not possible using jugular access, an alternate access via the right femoral vein may be used instead. If the femoral vein is used, only resting measurements will be performed, since it is not possible to thread on the supine bike with a catheter in situ via the femoral vein. In parallel, in a short 8F introducer, a 7F multilumen Swan Ganz catheter will be placed over a long guidewire, and advanced to the left or right pulmonary artery. The catheter will then be wedged in a pulmonary segmental artery to obtain left atrial pressure curves.

The pacing lead will be connected to an electrophysiology 12-lead ECG system with capability of pacing and continuous ECG registration of external and intracardiac signals. His bundle potential will be identified probing the correct anatomical area with the electrode, and subsequently the pacing properties of the electrode will be tested. If the pacing properties result in selective (or subsidiary non-selective) his bundle capture, the electrode will be secured in place by 2-3 careful rotations of the entire electrode. The goal is to achieve a stable position with minimal risk of damaging the conduction system by trauma. Since the patients are paced using the CRT device anyway, trauma to the conduction system will not be deleterious, but it may hamper the ability to place a permanent HBP lead at a later stage, if that is indicated. When stable pacing conditions are in place, the stimulation program will be initiated. Depending on randomization results, the intrinsic CRT system or the temporary His-lead will be used for pacing. Pacing and sensing in the right atrium will be achieved using atrial electrodes on the outside of the Swan Ganz catheter.

The patients are randomized to wither Biventricular pacing first or HBP first. The pacing program involves both pacing at rest and pacing during sub maximal exercise test (supine bicycle test). Each pacing step is sustained at steady state for at least 2 minutes prior to measurement of intracardiac pressures, cardiac output and stroke volume calculations. After the pacing protocol is finished, the catheters are withdrawn and the site of venopuncture is manually compressed during 5 minutes. The patients are then observed during 2 hours, before discharge and completion of the study.

A hemodynamic optimization will be performed regarding device settings for all patients. Patients who have a positive hemodynamic response to HBP, compared to CRT pacing, will at a follow-up stage be offered an upgrade to permanent HBP.

Study Type

Interventional

Enrollment (Estimated)

30

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Contact Backup

Study Locations

      • Lund, Sweden
        • Skåne University Hospital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 99 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • Age >18 years
  • Native QRS-duration >130 ms and a left bundle branch block or previous His-ablation
  • Heart failure in New York Heart Association functional class II-IV
  • Echocardiographic non-responder to ongoing CRT defined as <15% reduction in LVESV compared to pre-CRT examination
  • Clinical non-responder to ongoing CRT, defined as lack of subjective improvement after CRT
  • Signed informed consent

Exclusion Criteria:

  • Known access site problems in vena jugularis interna dx or sin
  • Not able to perform supine cycling
  • Presence of severe tricuspid regurgitation

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: Double

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: His-bundle pacing first

AV sequential His-bundle pacing (or VVI pacing if in atrial fibrillation) via temporary right atrial and His-bundle electrodes.

Then AV sequential Biventricular pacing via the patient's already implanted CRT device.

Measurements of stroke volume, cardiac output, pressure in the right ventricle and pulmonary capillary wedge pressure, during rest and exercise. Simultaneous 12 lead ECG registration.

Temporary use of a Medtronic 3830 electrode placed in the region of His bundle, in combination with a Medtronic 5392 dual chamber external pacemaker and a temporary Edwards D200F7 Swan Ganz Pacing -catheter with atrial pacing capability.

AV synchronous pacing (DDD 60/minute during rest and 120/minute during exercise, AV delay 150/180ms) will be performed using the atrial port (pacing catheter atrial electrodes) and ventricular port (His bundle electrode), or in the case of atrial fibrillation, only pacing the the His bundle in VVI mode 60/minute during rest and 120/minute during exercise.

The Swan-Ganz will then be used for hemodynamic measurements during pacing. A Cardiotek ECG system will be used for simultaneous 12 lead ECG registration.

Active Comparator: Biventricular pacing first

AV sequential biventricular pacing (or VVI pacing if in atrial fibrillation) via the patient's already implanted CRT device. Then AV sequential or VVI His-bundle pacing via temporary right atrial and His-bundle electrodes.

Measurements of stroke volume, cardiac output, pressure in the right ventricle and pulmonary capillary wedge pressure, during rest and exercise. Simultaneous 12 lead ECG registration.

Temporary use of a Medtronic 3830 electrode placed in the region of His bundle, in combination with a Medtronic 5392 dual chamber external pacemaker and a temporary Edwards D200F7 Swan Ganz Pacing -catheter with atrial pacing capability.

AV synchronous pacing (DDD 60/minute during rest and 120/minute during exercise, AV delay 150/180ms) will be performed using the atrial port (pacing catheter atrial electrodes) and ventricular port (His bundle electrode), or in the case of atrial fibrillation, only pacing the the His bundle in VVI mode 60/minute during rest and 120/minute during exercise.

The Swan-Ganz will then be used for hemodynamic measurements during pacing. A Cardiotek ECG system will be used for simultaneous 12 lead ECG registration.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Change in stroke volume (ml) at rest, between Biventricular pacing compared to His bundle pacing
Time Frame: During procedure
Stroke volume will be measured using the thermodilution method, and paired T-test will be used to evaluate the change in stroke volume between the different pacing modes.
During procedure

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Change in stroke volume (ml) at sub-maximal exercise, between Biventricular pacing compared to His bundle pacing
Time Frame: During procedure
Stroke volume (ml) measured by the thermodilution method.
During procedure
Change in Cardiac output (l/min) at rest, between Biventricular pacing compared to His bundle pacing
Time Frame: During procedure
Cardiac output (l/min) at rest assessed by invasive hemodynamic measurements with the thermodilution method.
During procedure
Change in cardiac output (l/min) at submaximal exercise, between Biventricular pacing compared to His bundle pacing
Time Frame: During procedure
Cardiac output (l/min) at sub-maximal exercise assessed by invasive hemodynamic measurements with the thermodilution method using a Swan-Ganz catheter.
During procedure
Change in stroke work index (mmHg x ml/m2) at rest, between Biventricular pacing compared to His bundle pacing
Time Frame: During procedure
Calculation of stroke work index based on invasive measurements
During procedure
Change in stroke work index (mmHg x ml/m2) at sub-maximal exercise, between Biventricular pacing compared to His bundle pacing
Time Frame: During procedure
Calculation of stroke work index based on invasive measurements
During procedure
Change in QRS vector area, between Biventricular pacing compared to His bundle pacing
Time Frame: During procedure
Change in high resolution digital ECG recording of QRS vector area
During procedure

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Incidence of Treatment-Emergent Adverse Events, defined as pneumothorax or pericardial effusion.
Time Frame: From start of procedure to 2 hours post-procedure
If the patient develops symptoms of respiratory discomfort and/or blood pressure drop ≥10mmHg during or within 2 hours after the procedure, evaluation by chest X-ray and echocardiography will be performed. Presence of pneumothorax or pericardial effusion will be documented as adverse events.
From start of procedure to 2 hours post-procedure
Incidence of Treatment-Emergent Adverse Events, defined as local access site bleeding.
Time Frame: From start of procedure to 2 hours post-procedure
If there is bleeding that requires compression >10minutes it will be recorded as an adverse event.
From start of procedure to 2 hours post-procedure

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Oscar Braun, MD PhD, Region Skåne, Lund University
  • Principal Investigator: Rasmus Borgquist, MD PhD, Region Skåne, Lund University

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

September 1, 2021

Primary Completion (Estimated)

February 28, 2025

Study Completion (Estimated)

February 28, 2026

Study Registration Dates

First Submitted

December 18, 2020

First Submitted That Met QC Criteria

January 7, 2021

First Posted (Actual)

January 8, 2021

Study Record Updates

Last Update Posted (Estimated)

November 20, 2023

Last Update Submitted That Met QC Criteria

November 16, 2023

Last Verified

November 1, 2023

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Heart Failure, Systolic

Clinical Trials on Temporary His bundle pacing

3
Subscribe