Post Marketing Clinical Trial to Demonstrate the Performance of the MedHub AutocathFFR Device.

April 19, 2021 updated by: Medhub Ltd.

Patients with suspected coronary artery disease who are scheduled to undergo a coronary angiography procedure will be enrolled in the study. The angiographic images will be processed by the Medhub Autocath FFR device to generate the Autocath FFR measurement. Based on AMAR approval, MedHub Autocath FFR measurements may be used to determine revascularization in lesions found in the Left Anterior Descending (LAD) coronary artery. Consequently, invasive FFR (using a coronary pressure wire and hyperemic stimulus) is not mandatory for lesions in the LAD, although it is at the discretion of the physician whether or not to perform the invasive FFR procedure. Lesions in the Right Coronary Artery (RCA) and Left Circumflex Coronary (LCX) arteries, when clinically indicated, will be required to undergo an invasive FFR procedure in order to determine revascularization. In these cases, the Autocath FFR measurements will not be used for diagnostic or clinical decisions, but solely as a supportive tool. The MedHub Autocath FFR measurement per vessel will be compared to the invasive FFR measurement in the RCA and LCX lesions and in LAD lesions, for which invasive FFR measurements are available.

The dichotomously scored MedHub Autocath FFR per vessel will be compared to the invasive FFR, where an FFR ≤ 0.80 will be considered "positive", while an FFR > 0.8 will be considered "negative". The sensitivity and specificity of the MedHub Autocath FFR will be calculated.

Study Overview

Status

Unknown

Intervention / Treatment

Detailed Description

The proposed study is designed to assess performance and safety data for the MedHub Autocath FFR device.

The study is a prospective, multi-center, single-arm, post-marketing study. A total of 488 subjects will be enrolled in the study. The study population who represent the target population for this procedure consists of subjects with known or suspected coronary artery disease (CAD) who are scheduled to undergo a coronary angiography procedure and on whom invasive FFR will be measured in vessels with meaningful coronary lesions. Patients of both genders, all ethnicities, >18 years of age will be recruited to the study. Study subjects will be enrolled from numerous hospitals in Israel.

Investigators will screen subjects based on the inclusion/exclusion criteria described below. General patient demographics, medical history, concomitant medications, C-arm angiography system used, FFR pressure wire used, etc., will be obtained for study subject.

Subjects with known or suspected coronary artery disease who are scheduled for clinically indicated invasive coronary angiography will comprise the patient population. Coronary angiography will be performed in a routine fashion in patients with suspected coronary artery disease. The collected diagnostic angiography images will be processed using the Medhub Autocath FFR device to generate the Autocath FFR. Based on operator's discretion, the location of the stenosis and when clinically indicated, invasive FFR will be measured using a coronary pressure wire and hyperemic stimulus.

The patient's angiographic images will be processed using the MedHub Autocath FFR device. Based on AMAR approval, MedHub Autocath FFR measurements may be used to determine revascularization in lesions found in the Left Anterior Descending (LAD) coronary artery. Consequently, invasive FFR will not be mandatory for lesions in the LAD, and it will be at the discretion of the physician whether or not to perform the invasive FFR procedure. Lesions in the Right Coronary Anterior (RCA) and Left Circumflex (LCX) arteries, when clinically indicated, will be required to undergo an invasive FFR procedure in order to determine revascularization. That is, in these cases the MedHub Autocath FFR measurements will not be used for diagnostic or clinical decisions, but solely as a supportive tool. The MedHub Autocath FFR measurement per vessel will be compared to the invasive FFR measurement in the RCA and LCX lesions and in LAD lesions, for which invasive FFR measurements are available.

The dichotomously scored MedHub Autocath FFR per vessel will be compared to the invasive FFR, where an FFR ≤ 0.80 will be considered "positive", while an FFR > 0.8 will be considered "negative". The sensitivity and specificity of the MedHub Autocath FFR will be calculated. Furthermore, the MedHub Autocath FFR accuracy, positive predictive value and negative predictive value per vessel and per lesion will be determined.

Usability of the MedHub Autocath FFR device will also be assessed including User Manual comprehension and usability of the device in the hands of the users (i.e., interventional cardiologists).

Details on the timing of all study procedures are given in the Time and Events Schedule in Attachment B of the Protocol. The anticipated enrollment period is approximately 6 months and therefore, the complete duration of the study is expected to last 8-9 months, during which time 488 patients will be enrolled at 3-5 sites.

Study Type

Interventional

Enrollment (Anticipated)

488

Phase

  • Not Applicable

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Age > 18 years.
  • Subjects with suspected coronary artery disease who are scheduled to undergo a coronary angiography procedure.
  • Subject is planned to undergo an invasive FFR with IV or intracoronary Adenosine, Adenosine Triphosphate (ATP) or Papaverine used as hyperemic stimulus*.

    • Not mandatory for lesions in the LAD

Exclusion Criteria:

  • Vessel size less than 2 mm.
  • CTO in target vessel.
  • Prior CABG, heart transplant or valve surgery, prior TAVI/TAVR or severe left sided valvular disease.
  • TIMI Grade 2 or lower.
  • Target lesion involves Left Main (stenosis >50%).
  • Severe, diffuse disease defined as the presence of diffuse, serial gross luminal irregularities present in the majority of the coronary tree.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: OTHER
  • Allocation: NA
  • Interventional Model: SINGLE_GROUP
  • Masking: NONE

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Sensitivity and specificity
Time Frame: Immediate after procedure
Co-primary endpoints are the sensitivity and specificity of the dichotomously scored MedHub Autocath FFR per vessel compared to invasive FFR, where an FFR ≤ 0.80 is scored "positive", while an FFR > 0.8 is considered "negative".
Immediate after procedure

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (ANTICIPATED)

June 1, 2021

Primary Completion (ANTICIPATED)

June 1, 2021

Study Completion (ANTICIPATED)

August 1, 2021

Study Registration Dates

First Submitted

January 21, 2021

First Submitted That Met QC Criteria

January 25, 2021

First Posted (ACTUAL)

January 26, 2021

Study Record Updates

Last Update Posted (ACTUAL)

April 20, 2021

Last Update Submitted That Met QC Criteria

April 19, 2021

Last Verified

January 1, 2021

More Information

Terms related to this study

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Coronary Artery Disease

Subscribe