Clinical Decision Support System (CDSS) in Neurostimulation Therapy (CDSS)

February 2, 2021 updated by: Jose De Andres, General University Hospital of Valencia

Development of a Predictive Model of Effectiveness for the Implantation of Electrical Neurostimulators in Patients With Chronic Pain Using Imaging Biomarkers Extracted From Magnetic Resonance

Chronic pain is correlated with alterations in the structure and function of the brain, developed according to the phenotype of pain. Still today, the data on functional connectivity (FC), on chronic back pain, in patients with failed back surgery syndrome (FBSS), is limited. The selection process for the ideal candidate for spinal cord stimulation (SCS) is based on results from test and functional variables analysis as well as pain evaluation. In addition to the difficulties in the initial selection of patients and the predictive analysis of the test phase, which undoubtedly impact on the results in the middle and long term, the rate of explants is one of the most important concerns, in the analysis of suitability of implanted candidates. The hypothesis is that the structural and functional quantitative information provided by imaging biomarkers will improve the characterization of the patients compared to the characterization with the current clinical variables alone and this will allow establishing a CDSS that improve the effectiveness of the SCS implantation, optimizing human, economic and psychological resources.

A prospective, consecutive and observational, open-label, single-center study conducted at the Multidisciplinary Pain Management Department of our University Hospital. A total of 69 subjects were initially included in the study. The population split in 3 groups:

  • Interventional Group-SCS, included 35 patients with failed back surgery syndrome (FBSS) who were treated with SCS implants.
  • Comparator group included 23 patients with patients with chronic low-back pain who were treated with conventional medication (CM) for their pain.
  • Control Group included 11 subjects as health controls who volunteered to participate in the study.

MR images were obtained in a 1.5T MR system (Ingenia, Philips, Best, The Netherlands) using an 8-channel head coil.Clinical variables were evaluated at two different time points baseline and 12 months after SCS implantation or conventional medication. An ad hoc database was created to evaluate the different variables involved in pain , including sociodemographic variables (age, gender, level of studies and marital status), clinical variables (anxiety, depression, sleeping hours, resilience, NRS, the Pain Detect Questionnaire (PD-Q)) , and the images obtained from the fMRI.

Study Overview

Status

Completed

Detailed Description

Primary objective:To develop a predictive model integrated in a clinical decision support system (CDSS) feed by neuroimaging quantitative information objectively extracted from Magnetic Resonance (MR) images, which maximizes the appropriate use and effectiveness of electrical stimulation devices surgically implanted in selected patients with chronic pain.

Exploratory Objectives:

  1. -Analyze functional and anatomical brain connectivity patterns in patients with chronic pain , to develop a predictive model based on quantitative magnetic resonance neuroimaging which maximizes the effectiveness of neurostimulation devices surgically implanted in patients with chronic pain.
  2. -Analyze the relationship between neuroimaging biomarkers and the different clinical scales and variables captured from each patient (VAS, Oswestry Disability Index, DN4, Pain Detect, Moss, SF12, coping scale, optimism, resilience and HAD).

Test Device:1.5 Tesla MR system (Philips Healthcare, Best, The Netherlands) Boston Scientific Neuromodulation (BSN) Precision Spectra™ Spinal Cord Stimulation System with Illumina 3D™ Software and 32 contacts.

Device Description: Precision Spectra™ system IPG is a multiple independent current controlled pulse generator, capable of delivering current through 32 contacts. It is powered by a 3D programming software that considers the anatomical position of the leads. Two models of SCS leads will be provided, featuring 8 or 16 contacts with 1.3 mm diameter, 3 mm contact length, and contact spacing of 1, 4 or 6 mm. The use of SCS extensions will be optional to connect the IPG.

fMR description: The MR experiment will be consistent with on-label requirements. The MR procedure will be performed prior to the device implantation to avoid bias. Even more, the Food and Drug Administration (FDA) does not recommend the examination of patients with this kind of devices for security reasons.

Examinations will be performed in a 1'5 Tesla MR system (Philips Healthcare, Best, The Netherlands) at the Quiron Hospital. Decision of magnetic field is based on the quality of the examinations and must rely on label products and approval of company for interaction with implanted system.

A head coil with 8 reception channels will be used. Once the patient has been positioned in the system, initial and fast localization images will be acquired in order to properly plan the MR sequences of the research study.

After planning, a resting-state functional MR (rs-fMR) imaging sequence will be acquired, asking to the patient to be quiet with the eyes closed and thinking in a blue sky. The acquisition parameters will consist of an Echo Planar (EPI) dynamic T2* sequence, full brain coverage with the following parameters: TR=2000 ms; TE=30 ms; voxel size, 1.8 × 1.8 x 3.5 mm; flip angle, 90º; 40 axial slices; acquisition time 5:20 min.

A DTI MR sequence will be acquired in order to analyze white matter microstructure and connectivity by tractography techniques with the following parameters: Spin-Echo Echo Planar Imaging (SE-EPI) sequence, single shot; full brain coverage; 64 gradient directions; b-value, 1300 s/mm2; TR=6200 ms; TE=67 ms; voxel size, 2 x 2 x 2 mm; 60 axial slices; acquisition time 9:40 min.

An additional anatomic sequence will allow overlying structural and functional results and, in addition, obtaining the volumetry values of each brain region. The sequence parameters are: T1-weighted 3D gradient echo sequence (GRE), full brain coverage; TR=11.6 ms; TE=5.69; voxel size, 0.48 x 0.48 x 0.50 mm; flip angle, 8º; 280 axial slices; acquisition time 5:36 min.

After image acquisition, all data sets will be sent to the Imaging Biomarkers Platform of the Biomedical Imaging Research Group (GIBI230) of the La Fe Research Institute.

The fMR images will be aligned in order to correct possible small patient's head movements during examination. For that, the open source SPM8 (Statistical Parametric Mapping, http://www.fil.ion.ucl.ac.uk/spm/) software tool will be used. After movement correction, a temporal correction will be applied optimizing the slice timing. Images will be then normalized to a standardized brain template in order to allow for the study of the oscillations between individuals. After such processes, data will be filtered by a3D-Gaussian kernel in order to increase signal-to-noise ratio (SNR) while minimizing inter-subject differences. Finally, the application of independent component analysis (ICA) algorithms will allow for the extraction of brain activation maps in the subject during the acquisition.

The analysis of the DTI MR data for extracting white matter tracts connectivity will be performed with the open-source FSL software tool (http://www.fmrib.ox.ac.uk/fsl/). An initial Eddy currents correction will be applied to the images, in order to minimize slight images displacements and geometry distortions. The brain will be then segmented using the BET algorithm and brain data of all patients will be normalized to a common template for the group-based analysis. After this process, fractional anisotropy (FA), diffusivity (D) and orientation maps will be obtained .

Once the MR images have been processed, the structural and functional connectivity properties have to be extracted from the regions of interest (ROI). The positioning of these regions will be obtained from the zones involved in the Default Mode Network (DMN). The DMN might take an important role in pain perception and shows a high correlation with the symptoms described by patients, which makes this network useful for the prediction of patient response after the implantation of electrical stimulation, either at a functional or at a structural level.

Since images will be normalized to a common template, the automated area labeling AAL-tool will be used to define the regions of interest of the study, that compound the DMN and are formed by the medial temporal lobe, prefrontal cortex, posterior cingulate, precuneus and the parietal cortex. After measurement the connectivity parameters in these regions, a predictive model will be developed by combining clinical variables (scales and symptoms of each patient) and neuroimaging information. These models will be initially adjusted with a total of 30 patients (training data) and later validated in a group of 30 patients (validation data), obtaining thereafter results of specificity, sensitivity and models precision.

Study Type

Interventional

Enrollment (Actual)

60

Phase

  • Not Applicable

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

20 years to 80 years (Adult, Older Adult)

Accepts Healthy Volunteers

Yes

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Patients presenting pain of more than 6 months in duration
  • VAS Score at baseline ≥ 5
  • Patients with degenerative spine pain. Non specific low-back pain, nociceptive pain / mixed neuropathic
  • Post-operative spine pain, failed back surgery syndrome, mixed pain
  • Low consumption of analgesic and adjuvant drugs.
  • Pure radiculopathy
  • No suffering other serious chronic diseases.
  • No history of drug or alcohol.

Exclusion Criteria:

  • Having implanted pacemakers, stimulators or hearing aids incompatible with MR imaging.
  • Patients presenting psychiatric illness or significant cognitive deficits.
  • Psychological instability.
  • History of alcohol and drugs.
  • Severe coagulopathy.
  • Pending Surgery.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Screening
  • Allocation: Non-Randomized
  • Interventional Model: Sequential Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Active Comparator: Study group-SCS Impanted
30 patients with chronic pain (at least 6 months) with pre- and post-surgery evaluation (imaging and clinical evaluation) implanted with Precision SpectraTM for the validation study and for the construction of the predictive model.
Analyze the relationship between neuroimaging biomarkers and the different clinical scales and variables captured from each patient
Other Names:
  • Functional Magnetic Resonance performed after implating SCS
No Intervention: Comparator-chronic pain
20 patients with chronic pain (at least 6 months) with degenerative spine pain. Non-specific low-back pain, nociceptive pain / mixed neuropathic. This will be the comparator group
No Intervention: Control-healthy volunteers
10 volunteers without pain or related disease, age less than 25 years, to establish a control group whose pattern is used as a comparator with chronic pain groups. This is the control group

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Clinical decision support system (CDSS) for selection of patients candidates for SCS implant
Time Frame: 12 months
to analyze the neuronal circuits involved in FBSS patients in order to extract predictive imaging biomarkers capable of determining the characteristics of patients that predict the success of SCS implants. This information might be used to develop a CDSS to maximize the effectiveness of electrical stimulation devices surgically implanted in patients with chronic pain.
12 months

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
neuronal circuits involved in chronic pain
Time Frame: 12 months
to describe neuronal circuits involved in chronic pain by comparation between patients and control subjects; identify differences in the neural circuits between patients who have successfully undergone the implantation of the SCS and who failed the trial phase attending to the current criteria
12 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

November 20, 2016

Primary Completion (Actual)

June 30, 2018

Study Completion (Actual)

June 30, 2019

Study Registration Dates

First Submitted

January 26, 2021

First Submitted That Met QC Criteria

January 30, 2021

First Posted (Actual)

February 3, 2021

Study Record Updates

Last Update Posted (Actual)

February 4, 2021

Last Update Submitted That Met QC Criteria

February 2, 2021

Last Verified

February 1, 2021

More Information

Terms related to this study

Other Study ID Numbers

  • Biomarcadores-RM

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.