A 7 Day Adhesive Device Wear Study to Evaluate BD StatLock™ Devices

September 7, 2021 updated by: Becton, Dickinson and Company
This study is a prospective healthy human volunteer wear study to evaluate the safety and performance of the StatLock™ Catheter Stabilization Device and the StatLock™ stabilization device Accessory (Foam Strip) and to meet the design input requirements.

Study Overview

Detailed Description

This is a healthy human wear study of a minimum of 100 study participants who will wear the StatLock™ Catheter Stabilization Device and the StatLock™ stabilization device accessory (Foam Strip). Participants will report to the research center, on Day 0 to have their skin assessed by the investigator, clinician, and/or designee. On Day 0 the participants will have the StatLock™ Catheter Stabilization Devices and the foam strip applied. There will be 4 separate StatLock™ devices being tested, which will be randomly applied to the participants on their inner (ventral) forearms. Each participant will have 2 separate devices applied, one to each arm. Those participants who have the StatLock ™ Arterial Plus, StatLock™ Dialysis II, or the StatLock IV Select , will also have the Foam Strip applied.

The participant will also wear a dressing over top of a StatLock™ compatible catheter, which will be connected to an extension set. For this study, the catheter will not be inserted into the study participant but instead will be shortened or modified and placed on the skin and secured with the StatLock™ Stabilization Device. The catheter tip will be marked with indelible ink, to assess for any catheter movement. The use of the catheters, extension set, and the dressing represents how the device would be used in a clinical setting.

Once the devices are applied the participant will have the skin assessed (using a modified Draize Scale) by the investigator, clinician, and/or designee on Day 0. The participant will wear the devices at home for 1 week and will return on Day 7 for a skin assessment, evaluation of pad and foam strip lift, and catheter movement, evaluation of pain, and removal of the devices by the investigator, clinician, and/or designee. On Day 8, there may be a phone call and/or video call with the participant to review the skin condition of the site where the device was removed from.

Study Type

Observational

Enrollment (Actual)

120

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • New Jersey
      • Fair Lawn, New Jersey, United States, 07410
        • TKL Research, Inc
    • Pennsylvania
      • Broomall, Pennsylvania, United States, 19008
        • Dermico, LLC

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

14 years and older (Child, Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Potential participants will be recruited from healthy volunteer participant pools identified and/or held by the study site(s).

Description

Inclusion Criteria:

  • Any healthy human, 14 years of age or older, regardless of gender (Note: Consent of guardian or parent may be required for patients who are under 18 years of age).
  • If they provide written and informed consent.
  • Expected to be available for observation from consent, to the application on Day 0, and then removal of the StatLock™ Catheter Stabilization Device on Day 7, and for a follow up phone and/or video call on Day 8 for a skin assessment.
  • Participant agrees to maintain a diary from device application, Day 0, through device removal, Day 7.
  • Agree to keep the StatLock™ Catheter Stabilization Device, ancillary products, and site dry throughout the study
  • Agrees to not use moisturizers, lotions, or any product labeled moisturizing for the duration of the study, at or around the site where the devices will be applied (inner (ventral) forearm).

Exclusion Criteria:

  • If their skin was burned or scarred at the insertion site, or has any other skin condition which may affect the adherence and/or interpretation of the study results.
  • If they have a known allergy to the study products (known tape or adhesive allergy).
  • Has a medical condition which may prevent him or her from completing the study or place the study participant at undue risk.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Other
  • Time Perspectives: Prospective

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Healthy volunteers
There will be 4 separate StatLock™ devices being tested, which will be randomly applied to the participants on their inner (ventral) forearms. Each participant will have 2 separate devices applied, one to each arm. Those participants who have the StatLock ™ Arterial Plus, StatLock™ Dialysis II, or the StatLock IV Select, will also have the foam strip applied.
Each participant will have 2 separate devices applied, one to each arm, with or without a Foam Strip.The participant will also wear a dressing over top of a StatLock™ compatible catheter, which will be connected to an extension set. For this study, the catheter will not be inserted into the study participant but instead will be shortened or modified and placed on the skin and secured with the StatLock™ Stabilization Device. The use of the catheters, extension set, and the dressing represents how the device would be used in a clinical setting.
Other Names:
  • StatLock™ Pro- SLP001
  • StatLock™ IV Select- IV0521CE
  • StatLock™ Arterial Plus- ART0222CE
  • StatLock™ Dialysis II- VDS2
  • Foam Strip- SL0100 (StatLock™ Stabilization Device Accessory)

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percent lift of the StatLock™ Catheter Stabilization Device Pad at Day 7
Time Frame: Day 7
Device pad lift will be assessed using a Visual Analog Scale, ranging from 0 - Fully adhered to 100 - Device absent. When the percent lift at day 7 is >50%, it is considered a failure.
Day 7
Percent of Device Movement of the StatLock™ Catheter Stabilization Device at Day 7
Time Frame: Day 7
The percent of device movement of the catheter will be assessed using a Visual Analog Scale from 0 (no movement) to 100 (completely displaced).
Day 7
Ease of Removal of the StatLock™ Catheter Stabilization Device
Time Frame: Day 7
Ease of removal will be assessed using a 100 mm Visual Analog Scale where 0 is very easy to remove and 100 is very difficult to remove.
Day 7
Percent of participants with erythema score ≥ 2
Time Frame: Day 7
Erythema score will be assessed using a modified Draize scale with scores ranging from 0 to 3, as follows: 0 - no visible reaction; 0.5 - Doubtful or negligible erythema reaction; 1.0 - mild or just perceptible macular erythema reaction in a speckled/follicular, patchy or confluent pattern (slight pinking); 2.0 - Moderate erythema reaction in a confluent pattern (definite redness); 3 - Strong or brisk erythema reaction that may spread beyond the test site).
Day 7

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percent lift of the Foam Strip (StatLock™ Stabilization Device Accessory) at Day 7
Time Frame: Day 7
The percent lift of the foam strip will be assessed using a Visual Analog Scale ranging from 0 (fully adhered) to 100 (device accessory absent).
Day 7
Ease of Removal of the StatLock™ Stabilization Device Foam Strip
Time Frame: Day 7
The ease of removal will be assessed using a 100 mm Visual Analog Scale where 0 is very easy to remove and 100 is very difficult to remove.
Day 7
Level of pain upon removal of the StatLock™ Catheter Stabilization Device
Time Frame: Day 7
Pain upon removal of the device will be assessed using a 100 mm Visual Analog Scale where 0 is no pain and 100 is unbearable pain.
Day 7

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Rate of StatLock™ Catheter Stabilization device-related adverse events
Time Frame: Day 0 to Day 8
Rate of device-related adverse events.
Day 0 to Day 8

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Michael Casser, MD, TKL Research, Inc.
  • Principal Investigator: Gary Grove, PhD, Dermico, LLC

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

July 13, 2021

Primary Completion (Actual)

August 23, 2021

Study Completion (Actual)

August 23, 2021

Study Registration Dates

First Submitted

April 9, 2021

First Submitted That Met QC Criteria

April 13, 2021

First Posted (Actual)

April 15, 2021

Study Record Updates

Last Update Posted (Actual)

September 8, 2021

Last Update Submitted That Met QC Criteria

September 7, 2021

Last Verified

September 1, 2021

More Information

Terms related to this study

Other Study ID Numbers

  • MDS-21STLOCK001

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Peripheral Intravenous Catheter Stabilization

3
Subscribe