Dexcom Hybrid Closed Loop Insulin Pump Study in Type 1 Diabetes
An Evaluation of the Tandem IQ Insulin Pump and DEXCOM G6 Continuous Glucose Monitoring Hybrid Closed Loop Insulin Delivery System (Control-IQ) on Patient Wellbeing and Diabetes Control in Adults With Type 1 Diabetes
Treatment of Type 1 diabetes requires daily insulin therapy given by either multiple daily injections or by continuous infusion via a pump. The insulin dose is calculated taking into consideration blood glucose levels, food intake and activity levels, aiming to avoid high and low readings.
The Tandem t:slim X2™ insulin pump has recently become available in NHS Scotland and can link with the Dexcom G6 continuous glucose monitoring (CGM) system.
The Tandem Control IQ hybrid closed loop system is made up of the Tandem insulin pump communicating wirelessly with the Dexcom G6 CGM. The algorithm within the pump then automatically controls the background (basal) insulin infusion rate. Manually delivered bolus insulin is still required for meals. This system is designed to finely tune insulin delivery and increase the amount of time that glucose is in the target range, reducing hypoglycaemia and hyperglycaemia.
Insulin pumps are funded by NHS Scotland, but the Dexcom G6 continuous glucose monitoring system is not, so this hybrid closed loop system is not currently routinely available in NHS Scotland.
The purpose of this study is to gain experience of this system and investigate how effective and acceptable the hybrid closed loop system is for people in Scotland. Participants will be provided with the Dexcom G6 CGM device to link to their current pump to create the hybrid closed loop system for the duration of the study.
30 people with Type 1 diabetes who have been using the Tandem pump for at least one month will be invited to participate.
The study will include a screening visit, a 30 day run-in period, 52 week treatment period and 12 week follow-up period. Diabetes control will be measured using HbA1c and downloaded information from the hybrid closed loop system. Participants will also complete questionnaires and a reflective diary.
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
In those without diabetes insulin is automatically secreted according to requirement. The background insulin secretion rate will vary depending on need, and is affected by many factors e.g. activity or stress levels. When a person eats the rise of glucose is sensed and insulin automatically secreted to keep the glucose within a narrow range.
People with Type 1 diabetes cannot make insulin to control their blood glucose. Insulin is therefore required as treatment. This is given subcutaneously by injection or using an infusion pump. Both background insulin is required and boluses of insulin given when carbohydrate is eaten.
It is difficult to accurately calculate and deliver the correct amount of insulin. More recently standard treatment involved giving one of two injections of long acting insulin and more frequent injections of fast acting insulin when carbohydrate is ingested. The dose of insulin is determined by measuring the ambient glucose level, knowing the amount of carbohydrate in any meal and predicting activity during the subsequent four or five hours.
Insulin pumps can be used to provide a pre-tuned variable background insulin infusion with manual intervention to deliver boluses of insulin with food.
In Scotland many people have now replaced finger prick glucose testing with flash glucose monitoring. This uses a small subcutaneous needle that senses subcutaneous glucose levels and, when scanned this sensor will provide a measure of subcutaneous glucose and information on the last 8 hours of glucose values (Flash Glucose monitoring). This system does not currently link to an insulin pump.
The Tandem insulin pump has recently become available in NHS Scotland for those already established on pump therapy.
The Dexcom G6 continuous glucose monitoring (CGM) system provides continuous glucose monitoring (without requiring any scanning by the patient) but is not currently funded within NHS Scotland, except for special circumstances. This system does communicate via bluetooth with the T-slim insulin pump.
The Tandem Control IQ insulin delivery system uses the link between the dexcom glucose sensor and pump to control the background insulin delivery and has been shown to work well in those studied. Patients still have to administer insulin when ingesting carbohydrate. This system is described as a hybrid closed loop system.
Problems that can potentially occur using this system include loss of communication between sensor and pump (bluetooth or wired connection), the algorithm not working or glucose readings out with the set ranges resulting in the automatic system turning off.
It is also unknown how acceptable the 'loss of control' is to a more general diabetes population as individuals with diabetes are used to making all the insulin dose decisions very frequently across each day. Confidence in these systems, both by patients and clinicians will be vital if these are to be introduced into more general use.
Therefore, the potential for these systems to have a major impact for people with diabetes is clear, but further understanding of the effectiveness and acceptability of such systems in the Scottish population is required.
The Investigator's aim is to investigate the effect of this system compared to current 'best practice' in Scotland, to gain experience in using this system and to investigate the acceptability of this system in the Scottish population.
Study Type
Enrollment (Actual)
Contacts and Locations
Study Locations
-
-
Lothian
-
Edinburgh, Lothian, United Kingdom, EH4 2XU
- Professor John McKnight
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Sampling Method
Study Population
Description
Inclusion Criteria:
- Clinical diagnosis of Type 1 diabetes
- Established on the Tandem pump for at least 1 month, using novorapid or humalog insulin
- HbA1c 40-90 mmol/mol inclusive within the previous 30 days or taken at screening
- Total daily dose of insulin 10-100 units
- Willing to attend study visits and complete all study assessments
- Able to give informed consent
Exclusion Criteria:
- Pregnant or planning pregnancy
- Any disorder which, in the investigators' opinion might jeopardise the patient's safety or compliance with the protocol
- Life expectancy less than one year
- Severe renal impairment (eGFR <30ml/min)
- Weight <25kg
- Patients prescribed hydroxycarbamide
Study Plan
How is the study designed?
Design Details
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
HbA1c
Time Frame: Change from baseline to 20 days after initiation of the sensor, and at 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system
|
Measurement of HbA1c
|
Change from baseline to 20 days after initiation of the sensor, and at 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system
|
|
Severe hypoglycaemia
Time Frame: Change from the 12 weeks before baseline measures to 20 days after initiation of the sensor, and at 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Self reported severe hypoglycaemia
|
Change from the 12 weeks before baseline measures to 20 days after initiation of the sensor, and at 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
|
Time in range
Time Frame: Change from baseline to 4, 12, 26 and 52 weeks after initiation of the Control-IQ hybrid closed loop system
|
Percentage of time glucose is in range (download of Dexcom results)
|
Change from baseline to 4, 12, 26 and 52 weeks after initiation of the Control-IQ hybrid closed loop system
|
|
Time in hypoglycaemia
Time Frame: Change from baseline to 4, 12, 26 and 52 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Percentage of time hypoglycaemic
|
Change from baseline to 4, 12, 26 and 52 weeks after initiation of the Control-IQ hybrid closed loop system.
|
|
Time Control IQ system not in use
Time Frame: From baseline to 4, 12, 26 and 52 weeks after initiation of the system.
|
Time when the Control-IQ hybrid closed loop system is not in use
|
From baseline to 4, 12, 26 and 52 weeks after initiation of the system.
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Diabetes distress score
Time Frame: Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Diabetes distress score.
The Diabetes Distress Score yields a total diabetes distress scale score plus 4 sub scale scores, each addressing a different kind of distress.
The score is the sum of the patient's responses to the appropriate items divided by the number of items in that scale.
A mean question score of 3 or higher (moderate distress) as a level of distress is worthy of clinical attention.
|
Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
|
Problem Areas In Diabetes questionnaire
Time Frame: Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Problem Areas In Diabetes questionnaire. Each question has five possible answers with a value from 0 to 4, with 0 representing "no problem" and 4 "a serious problem". The scores are added up and multiplied by 1.25, generating a total score between 0 - 100. Patients scoring 40 or higher may be at the level of "emotional burnout" and warrant special attention. PAID scores in these patients may drop 10-15 points in response to educational and medical interventions. An extremely low score (0-10) combined with poor glycaemic control may be indicative for denial |
Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
|
Patient experience diary
Time Frame: Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Patient written report of experience using hybrid closed loop system
|
Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
|
Hospital Anxiety and Depression questionnaire
Time Frame: Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Hospital Anxiety and Depression questionnaire.
Minimum score is 0, maximum score is 21.
0-7 = Normal, 8-10 = Borderline abnormal (borderline case), 11-21 = Abnormal (case)
|
Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Other Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Gold Hypoglycaemia questionnaire
Time Frame: Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Gold hypoglycaemia questionnaire.
The Gold score measures hypoglycaemia awareness.
Minimum score is 1 - always aware, Maximum score is 7 - never aware.
The higher the score the greater the risk of severe hypoglycaemia
|
Change from baseline to 4, 12, 26, 52 and 64 weeks after initiation of the Control-IQ hybrid closed loop system.
|
Collaborators and Investigators
Sponsor
Collaborators
Investigators
- Principal Investigator: John McKnight, NHS Lothian
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Estimated)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- 299847
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Type 1 Diabetes Mellitus
-
COUR Pharmaceutical Development Company, Inc.RecruitingType 1 Diabetes | Type 1 Diabetes Mellitus | T1DM | T1D | Type 1 Diabetes in Adolescence | Type 1 Diabetes in Children | Type 1 Diabetes Patients | Type 1 Diabetes Mellitis | T1DM - Type 1 Diabetes Mellitus | Type 1 Diabetes (Juvenile Onset)United States
-
Sultan Qaboos UniversityUniversity of Mosul; University of Child Health Sciences and Children's Hospital...Not yet recruitingType 1 Diabetes Mellitus | T1DM | Type 1 Diabetes Mellitus (T1DM) | T1DM - Type 1 Diabetes Mellitus
-
Lund UniversityEnrolling by invitationType 1 Diabetes Mellitus | Stage 2 Type 1 Diabetes | Stage 1 Type 1 Diabetes | Stage 3 Type 1 DiabetesSweden
-
Superior UniversityActive, not recruitingType 2 Diabetes Mellitus 1Pakistan
-
SanofiCompletedType 1 Diabetes Mellitus-Type 2 Diabetes MellitusHungary, Russian Federation, Germany, Poland, Japan, United States, Finland
-
Abdullah KarsNot yet recruitingType 1 Diabetes Mellitus | Autoimmune Diabetes | Type 1 Diabetes Mellitus (T1DM)Turkey (Türkiye)
-
Immunocore LtdNot yet recruitingType 1 Diabetes | Diabetes Type 1 | Type 1 Diabetes (T1D)
-
University of Colorado, DenverMassachusetts General Hospital; Beta Bionics, Inc.CompletedDiabetes Mellitus, Type 1 | Type 1 Diabetes | Diabetes type1 | Type 1 Diabetes Mellitus | Autoimmune Diabetes | Diabetes Mellitus, Insulin-Dependent | Juvenile-Onset Diabetes | Diabetes, Autoimmune | Insulin-Dependent Diabetes Mellitus 1 | Diabetes Mellitus, Insulin-Dependent, 1 | Diabetes Mellitus, Brittle | Diabetes Mellitus, Juvenile-Onset and other conditionsUnited States
-
University of California, San FranciscoJuvenile Diabetes Research FoundationCompletedType 1 Diabetes Mellitus | Diabetes Mellitus, Type I | Insulin-Dependent Diabetes Mellitus 1 | Diabetes Mellitus, Insulin-Dependent, 1 | IDDMUnited States, Australia
-
Al-Zaytoonah University of JordanActive, not recruitingType 1 Diabetes | Type 1 Diabetes MellitusEgypt
Clinical Trials on Tandem Control IQ Hybrid Closed Loop Insulin Delivery System
-
University of Colorado, DenverNational Cancer Institute (NCI); Children's Hospital Colorado; DexCom, Inc.; Tandem...TerminatedHigh Risk Acute Lymphoblastic LeukemiaUnited States
-
Masaryk Hospital Usti nad LabemRecruitingDiabetes Mellitus, Type 1Czechia
-
Jaeb Center for Health ResearchThe Leona M. and Harry B. Helmsley Charitable TrustRecruitingDiabetes Mellitus, Type 1 | Pregnancy in Diabetics | Insulin Dependent Diabetes | Pregnancy, High-RiskUnited States
-
Centre Hospitalier Sud FrancilienCompletedSecondary Pancreatic DiabetesFrance
-
Rigshospitalet, DenmarkAbbott; The Novo Nordic Foundation; mylife Diabetes Care AGRecruitingPregnancy Complications | Diabete Type 1Denmark
-
University of VirginiaDexCom, Inc.; Tandem Diabetes Care, Inc.CompletedType 1 Diabetes MellitusUnited States
-
King Fahad Medical CityActive, not recruitingInsulin Pump TherapySaudi Arabia
-
Rabin Medical CenterTandemNot yet recruiting
-
Centre Hospitalier Sud FrancilienCompletedType 1 DiabetesFrance
-
Second Xiangya Hospital of Central South UniversityRecruitingType 2 DiabetesChina