Lifelight Non-Invasive Blood Pressure and Heart Rate Validation Study

August 1, 2024 updated by: Xim Limited

CLINIMARK Clinical Investigation Plan Lifelight Non-Invasive Blood Pressure Validation Study PR 2021-443

Lifelight is a novel, calibration-free, software-only medical device that contactlessly measures pulse rate and blood pressure by measuring remote photoplethysmography (rPPG) signals of a patient's face using a standard smartphone or tablet camera.

By conducting this validation study we aimed to demonstrate the accuracy of Lifelight for classifying the full range of clinically relevant blood pressures as hypertensive or non-hypertensive and demonstrate its accuracy for measuring pulse rate and the blood pressure of people with blood pressures relevant to Stage 1 hypertension.

Study Overview

Status

Completed

Conditions

Detailed Description

Lifelight was investigated in a study that followed the data collection and data analysis methodology outlined in ISO 81060-2:2018/AMD 1:2020 "Non-invasive Sphygmomanometers - Part 2: Clinical investigation of automated measurement type". This validation study was conducted by independent laboratory Element Materials Technology Boulder (formerly Clinimark), a global leader for clinical testing of vital sign data for medical devices and consumer wearable products. The study generated data from 85 people aged 18-85 with the wide-ranging distribution of blood pressures specified in ISO 81060-2:2018/AMD 1:2020.The accuracy of Lifelight's blood pressure measurements was assessed by comparing Lifelight's measurements with measurements made by dual-observer manual auscultation using the same-arm sequential method specified in ISO 81060-2:2018/AMD 1:2020. The accuracy of Lifelight's pulse rate measurements was assessed by comparing Lifelight's measurements with concurrent ECG-derived heart rate values. Data was collected in an Independent Review Board-approved study conducted by independent laboratory Clinimark (now Element) in the USA and repurposed for the testing of the new improved model in the latest version of the Lifelight® EA device. The data collected consists of 85 participants, with three readings each. The participants were resting before their measurements were taken to ensure stable blood pressure. For each Lifelight measurement, two sets of blood pressure measurements were taken using dual-observer manual auscultation. The first measurement was taken before the Lifelight measurement and the second just after. If there was too much variation between the before and after measurement, or between the two independent observers, the measurement was discarded and repeated. For this validation procedure, the measurements taken during the original study were re-processed in the new updated algorithm.

Study Type

Interventional

Enrollment (Actual)

85

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Colorado
      • Louisville, Colorado, United States, 80027
        • Element Materials Technology

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  • Healthy volunteers aged 18 to 85 years old

Exclusion Criteria:

  • Participants were excluded from the study if they were medically unsuitable for participation at time of visit, any heart dysrhythmias (except respiratory sinus arrhythmia) as confirmed with a 3 lead ECG, compromised circulation or peripheral vascular disease, clotting disorders, female participants who were pregnant or trying to get pregnant, excessive facial hair, and conditions that affect the skin, such as anaemia, jaundice, rosacea, psoriasis, acute acne, and erythropoietic protoporphyria.

Participants could choose to withdraw themselves from the study without prejudice or they could be withdrawn by study investigators for predetermined reasons. Data excluded from the analysis was documented with justifications.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Lifelight and BP measurement
The accuracy of Lifelight's blood pressure measurements was assessed by comparing Lifelight's measurements with measurements made by dual-observer manual auscultation using the same-arm sequential method specified in ISO 81060-2:2018/AMD 1:2020. The accuracy of Lifelight's pulse rate measurements was assessed by comparing Lifelight's measurements with concurrent ECG-derived heart rate values. The participants were resting before their measurements were taken to ensure stable blood pressure. For each Lifelight measurement, two sets of blood pressure measurements were taken using dual-observer manual auscultation. The first measurement was taken before the Lifelight measurement and the second just after. If there was too much variation between the before and after measurement, or between the 2 independent observers, the measurement was discarded and repeated. For this validation procedure, the measurements taken during the original study were re-processed in the new updated algorithm.
Lifelight is a artificial-intelligence software application that enables completely contactless spot measurements (i.e., one-off measurements without prior calibration) of PR and BP. Based on the science of remote photoplethysmography (rPPG), Lifelight works by detecting tiny changes in the colour of facial skin that occur every time the heart beats. All that is required to generate PR and BP measurements is that the software application is downloaded onto a computer device (e.g., smartphone or tablet) with a standard camera, and that the patient's face stays in the line of sight of the camera for up to 60 seconds.
Blood Pressure Measurement

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Blood Pressure Classification Accuracy
Time Frame: Up to 3 Months
Demonstrate the accuracy of Lifelight for classifying the full range of clinically relevant blood pressures as hypertensive or non-hypertensive.
Up to 3 Months
Pulse Rate Measurement Accuracy (relevant to Stage 1 hypertension
Time Frame: Up to 3 Months
Demonstrate its accuracy for measuring pulse rate (bpm)
Up to 3 Months
Blood Pressure Measurement Accuracy (relevant to Stage 1 hypertension
Time Frame: Up to 3 Months
the blood pressure (mmHg) of people with blood pressures relevant to Stage 1 hypertension
Up to 3 Months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Arthur Cabrera, MD, Investigator has left Element Materials Technology

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

May 18, 2023

Primary Completion (Actual)

August 3, 2023

Study Completion (Actual)

August 3, 2023

Study Registration Dates

First Submitted

July 18, 2024

First Submitted That Met QC Criteria

August 1, 2024

First Posted (Actual)

August 2, 2024

Study Record Updates

Last Update Posted (Actual)

August 2, 2024

Last Update Submitted That Met QC Criteria

August 1, 2024

Last Verified

July 1, 2024

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • PR 2021-443

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Hypertension

Clinical Trials on Lifelight

Subscribe