Estudio de viabilidad temprana del corazón artificial total BiVACOR®
The purpose of this study is to assess the feasibility of using the BiVACOR Total Artificial Heart (TAH) System to support adult patients with severe biventricular heart failure, or univentricular heart failure in which left ventricular assist device (LVAD) support is not recommended, who require mechanical circulatory support to sustain life. The BiVACOR TAH System is intended for use as a bridge to transplant (BTT). Feasibility will be assessed by evaluating safety and performance of the BiVACOR TAH System in study subjects.
Descripción general del estudio
Estado
Estado
Condiciones
Condiciones
Intervención / Tratamiento
Intervención / Tratamiento
Tipo de estudio
Tipo de estudio
Inscripción (Estimado)
Inscripción
Fase
Fase
- No aplica
Contactos y Ubicaciones
Estudio Contacto
Estudio Contacto
- Nombre: Diane Covington
- Número de teléfono: 480-794-0232
- Correo electrónico: dcovington@crconcepts-cro.com
Copia de seguridad de contactos de estudio
- Nombre: Nicole Bartnikowski
- Número de teléfono: 657-540-9616
- Correo electrónico: nicole.bartnikowski@bivacor.com
Ubicaciones de estudio
-
-
Arizona
-
Phoenix, Arizona, Estados Unidos, 85006
- Reclutamiento
- Banner - University Medical Center Phoenix
-
Contacto:
- Francisco Arabia, MD, MBA
- Número de teléfono: 602-377-9310
- Correo electrónico: francisco.arabia@bannerhealth.com
-
-
North Carolina
-
Durham, North Carolina, Estados Unidos, 27710
- Reclutamiento
- Duke University Hospital
-
Contacto:
- Carmelo Milano, MD
- Número de teléfono: 919-684-3243
- Correo electrónico: carmelo.milano@duke.edu
-
-
Ohio
-
Cincinnati, Ohio, Estados Unidos, 45219
- Reclutamiento
- The Christ Hospital
-
Contacto:
- Robert D. Dowling, MD
- Número de teléfono: 513-585-4880
- Correo electrónico: robert.dowling@thechristhospital.com
-
-
Texas
-
Houston, Texas, Estados Unidos, 77030
- Reclutamiento
- Texas Heart Institute / Baylor St. Luke's Medical Center
-
Contacto:
- Joseph G. Rogers, MD
- Número de teléfono: 731-790-9401
- Correo electrónico: jrogers@texasheart.org
-
-
Criterios de participación
Criterio de elegibilidad
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
-
Patient has severe, irreversible biventricular heart failure and is eligible for biventricular mechanical circulatory support based on the International Society of Heart and Lung Transplantation (ISHLT) guidelines or has univentricular heart failure in which LVAD support is not recommended. ISHLT Guidelines for Biventricular Support:
-
Biventricular failure with at least two of the following hemodynamic and/or echocardiographic measurements.
- Right ventricular ejection fraction (RVEF) ≤ 30%
- Right ventricular stroke work index (RVSWI) ≤ 0.25 g/m/beat/m2
- Tricuspid annular plane systolic excursion (TAPSE) ≤ 14 mm
- Right ventricular (RV) to left ventricular (LV) end-diastolic diameter ratio > 0.72
- Central venous pressure (CVP) > 15mmHg
- CVP to pulmonary capillary wedge pressure (PCWP) ratio > 0.63
- Tricuspid insufficiency grade 4
- Pulmonary artery pressure index (PAPi) < 2
-
Treatment-refractory recurrent and sustained ventricular tachycardia or ventricular fibrillation in the presence of untreatable arrhythmogenic pathologic substrate.
-
Heart failure due to restrictive or constrictive physiology (e.g., hypertrophic cardiomyopathy, cardiac amyloidosis/senile or other infiltrative heart disease).
-
-
Patient is categorized as INTERMACS Patient Profile Classification 2 or 3.
-
Patient is classified as NYHA Class IV.
-
Patient has a left ventricular ejection fraction (LVEF) ≤ 25%, with the exception of patients with restrictive or constrictive physiology.
-
Patient is inotrope dependent, OR has a cardiac index (CI) ≤ 2.2 L/min/m2 without inotropes if inotropes are contraindicated (e.g., restrictive or constrictive heart failure), and meets one of the following criteria:
- Is on optimal medical management (OMM), based on current heart failure practice guidelines for at least 45 out of the last 60 days and is failing to respond or is not able to tolerate OMM; or
- Has advanced heart failure for at least 14 days and is dependent on an intra-aortic balloon pump (IABP) or similar temporary mechanical circulatory support device (MCSD) for at least seven days.
-
Patient is eligible for cardiac transplantation as determined by the implanting center.
-
Patient has adequate room in the chest as determined by 3-D imaging or other standard clinical assessments.
-
Patient has read and understands the informed consent form (ICF) and has voluntarily provided informed consent.
Exclusion Criteria:
-
Patient has contraindications to anticoagulation or antiplatelet therapies.
-
Patient has coagulopathy defined by a platelet count < 50 k/µl.
-
Patient has insufficient space in the chest to accommodate the BiVACOR pump.
-
Patient has a body mass index (BMI) ≥ 35 kg/m2.
-
Patient is highly pre-sensitized prior to pump implantation.
-
Patient is unconscious and unresponsive.
-
Patient is on pre-implant extracorporeal membrane oxygenation (ECMO) for > 7 days.
-
Patient is on pre-implant temporary MCSD for more than 21 days [e.g., Intra-aortic balloon pump (IABP), Impella, CentriMag, etc.] unless ambulatory and free from adverse effects associated with the MCSD.
-
Patient is implanted with durable mechanical circulatory support (LVAD or RVAD).
-
Patient experienced cerebrovascular accident (CVA) within three months of eligibility evaluation.
-
Patient has severe end-organ dysfunction as evidenced by:
- Total bilirubin > 4.0 mg/dL or cirrhosis confirmed by imaging or positive biopsy, and/or
- Glomerular filtration rate (GFR) < 30 mL/min/1.73 m2 or renal replacement therapy dependence.
-
Patient has severe chronic obstructive pulmonary disease (COPD) or restrictive lung disease requiring home oxygen.
-
Patient has primary pulmonary hypertension ≥ 8 Wood units.
-
Patient has severe systemic light-chain amyloidosis.
-
Patient has been diagnosed with severe cardiac cachexia or irreversible frailty.
-
Patient has diabetes with advanced diabetic neuropathy with accompanying skin ulceration.
-
Patient has a blood-borne infection within seven days of eligibility evaluation. Positive blood cultures reflective of contaminants (e.g., Staphylococcus epidermidis) will not be considered an exclusion.
-
Patient is pregnant or planning pregnancy.
-
Patient has a co-morbidity or illness that would limit survival to less than two years.
-
Patient has a current drug and/or alcohol addiction or known substance abuse.
-
Patient has insufficient social support or a history of non-compliance with medical instructions as determined by the Investigator.
-
Patient is participating in another clinical trial that may impact or confound the results of the BiVACOR TAH EFS.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Viabilidad del dispositivo
- Asignación: N / A
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
Número de brazos
Armas e Intervenciones
Grupo de participantes/brazoGrupo de participantes/brazo |
Intervención / TratamientoIntervención / Tratamiento |
|---|---|
|
Experimental: BiVACOR TAH
The BiVACOR TAH System will be implanted as a bridge to transplant (BTT) for adults with severe irreversible biventricular heart failure or univentricular heart failure in which LVAD support is not recommended, and who are eligible for cardiac transplantation. |
The BiVACOR pump is an implantable rotary biventricular blood pump that uses magnetic levitation technology for increased durability to replace both ventricles of a failing heart. The device is intended to replace the diseased heart in patients suffering from heart failure to bridge the time to heart transplant. |
¿Qué mide el estudio?
Medidas de resultado primarias
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Feasibility Endpoint
Periodo de tiempo: 6 months or time of heart transplant (if prior to 6 months post pump implant)
|
Survival on the original BiVACOR pump |
6 months or time of heart transplant (if prior to 6 months post pump implant)
|
|
Safety Endpoint
Periodo de tiempo: Through 6 months
|
Evaluation of the safety of the BiVACOR TAH System using INTERMACS adverse event (AE) terms and definitions (INTERMACS Manual of Operations Version 5). |
Through 6 months
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Inicio del estudio
Finalización primaria (Estimado)
Finalización primaria
Finalización del estudio (Estimado)
Finalización del estudio
Fechas de registro del estudio
Enviado por primera vez
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Publicado por primera vez
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización publicada
Última actualización enviada que cumplió con los criterios de control de calidad
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
Otros números de identificación del estudio
- CP-0001
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .