5 Years Long Term Results After Standalone CyPass-Implantation
Retro- and Prospective Monocentric Study to Evaluate the Safety and Effectiveness of the CyPass Stent 5 Years After Implantation
Descripción general del estudio
Estado
Condiciones
Intervención / Tratamiento
Descripción detallada
Tipo de estudio
Inscripción (Actual)
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Mecklenburg/Vorpommern
-
Neubrandenburg, Mecklenburg/Vorpommern, Alemania, 17036
- Dietrich-Bonhoeffer-Klinikum
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
- Niño
- Adulto
- Adulto Mayor
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Método de muestreo
Población de estudio
Descripción
Inclusion Criteria:
- patients who have undergone a standalone cypass implantation at least 3 years ago
- signed inform consent
Exclusion Criteria:
- Patients with too much travel distance between study center and home
- Patients who cannot visit the studycenter due to health problems
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
Cohortes e Intervenciones
Grupo / Cohorte |
Intervención / Tratamiento |
|---|---|
|
standalone CyPass implantation
patients who have undergone a standalone cypass implantation without cataract surgery more than 3 years ago
|
Heidelberg Retinal Topography, Visual field, intraocular pressure, medication anamnesis, Retinal Nerve Fibre Layer Thickness, Pachymetry, Sonography, full ophthalmological examination
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Number of Patients That Achieve Target Intraocular Pressure Measured in mmHg
Periodo de tiempo: up to 7 years
|
The Measurement of the intraocular Pressure is a criterion for the Efficacy of CyPass Stent Procedure (success rate).
The postoperative target pressure of 15 mmHg, 18 mmHg or 21 mmHg (depends on Glaucoma Status) has to be achieved.
|
up to 7 years
|
|
Change of Intraocular Pressure in mmHg Preoperatively (at Screening) vs Postoperatively (at Final Examination)
Periodo de tiempo: up to 7 years
|
The Measurement of the intraocular Pressure change is a criterion for the Efficacy of CyPass Stent Procedure (success rate). The postoperative pressure after 5 to 7 years has to be compared with the preoperative intraocular pressure (in mmHg) The change was calculated only from two time points. (value at the final examination time point minus the value at the preoperative time point |
up to 7 years
|
|
Number of Additional Medication
Periodo de tiempo: up to 7 years
|
The number of additional Glaucoma medication is another criterion to determine the Efficacy of CyPass Stent Procedure (success rate). Up to 4 different active substance can be used to reduce the pressure. The more substances the patient uses postoperatively the lower is the success after implantation of the CyPass-Stent. The postoperative number of medications after 5 years has to be compared with the preoperative number |
up to 7 years
|
|
Number of Participants With Secondary Glaucoma Procedures
Periodo de tiempo: up to 7 years
|
To determine the efficacy of CyPass Stent procedure completely the number of secondary glaucoma interventions (surgery/ laser) has to be documented
|
up to 7 years
|
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Number of Participants With Stability of Optic Nerve
Periodo de tiempo: up to 7 years
|
The Measurement of the optic nerve (HRT) in comparison to the baseline is needed to evaluate the glaucoma damage and to determine the Longterm success rate and the safety of the CyPass- Stent-Procedure. With papillary tomography you can see the changes in the optic nerve head concerning the permanent failures of retinal vessels , measure them very precisely and compare them with the previous findings. |
up to 7 years
|
|
Number of Participants With Stability of the Retinal Nerve Fiber Layer Thickness
Periodo de tiempo: up to 7 years
|
The comparison of the preoperative and postoperative Status of the Retinal Nerve Fiber Layer Thickness and retinal fundus thickness is another criterion to determine the absolute and relative success.
|
up to 7 years
|
|
Number of Participants With Stability of Visual Field
Periodo de tiempo: up to 7 years
|
The parameter of the Visual Field (Mean Deviation) will be measured to evaluate the Safety of the CyPass-Procedure.
The visual Field shows the progression of glaucoma defects concerning failures in different parts for example in the center or in the periphery of the visual field.
|
up to 7 years
|
|
Number of Participants With Stability of Stent Position
Periodo de tiempo: up to 7 years
|
Determination of Stent Position in the angle will be performed to exclude moving of the Stent.
The stent could change it's position forward into the anterior chamber or backwards in the suprachorioidal space
|
up to 7 years
|
|
Number of Participants With Stability of Stent Tissue Interaction
Periodo de tiempo: up to 7 years
|
Determination of iris tissue reaction to the stent will be performed.
We document slight reaction without involving the opening, iris reaction with involving the opening and complete closure by tissue.
The iris Tissue could overgrow the opening of the stent and therfore the function of the stent is disturbed or even failed.
|
up to 7 years
|
|
Number of Participants With Intra-and Postoperative Complications
Periodo de tiempo: up to 7 years
|
The number of intra-and postoperative complications are needed to evaluate the safety of the CyPass-procedure
|
up to 7 years
|
|
Number of Participants With Suprachoroidal Bleb
Periodo de tiempo: up to 7 years
|
The Determination of the suprachoroidal bleb is needed to evaluate the function of the stent and the efficacy of the procedure.
The bigger the bleb, the lower the pressure and the better the efficacy of the stent
|
up to 7 years
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: Helmut Hoeh, MD,FEBO, Department of Ophthalmology, Neubrandenburg, Germany
Publicaciones y enlaces útiles
Publicaciones Generales
- Garcia-Feijoo J, Rau M, Grisanti S, Grisanti S, Hoh H, Erb C, Guguchkova P, Ahmed I, Grabner G, Reitsamer H, Shaarawy T, Ianchulev T. Supraciliary Micro-stent Implantation for Open-Angle Glaucoma Failing Topical Therapy: 1-Year Results of a Multicenter Study. Am J Ophthalmol. 2015 Jun;159(6):1075-1081.e1. doi: 10.1016/j.ajo.2015.02.018. Epub 2015 Mar 3.
- Hoeh H, Vold SD, Ahmed IK, Anton A, Rau M, Singh K, Chang DF, Shingleton BJ, Ianchulev T. Initial Clinical Experience With the CyPass Micro-Stent: Safety and Surgical Outcomes of a Novel Supraciliary Microstent. J Glaucoma. 2016 Jan;25(1):106-12. doi: 10.1097/IJG.0000000000000134.
- Grisanti S, Margolina E, Hoeh H, Rau M, Erb C, Kersten-Gomez I, Dick HB, Grisanti S. [Supraciliary microstent for open-angle glaucoma: clinical results of a prospective multicenter study]. Ophthalmologe. 2014 Jun;111(6):548-52. doi: 10.1007/s00347-013-2927-6. German.
- Hoh H, Grisanti S, Grisanti S, Rau M, Ianchulev S. Two-year clinical experience with the CyPass micro-stent: safety and surgical outcomes of a novel supraciliary micro-stent. Klin Monbl Augenheilkd. 2014 Apr;231(4):377-81. doi: 10.1055/s-0034-1368214. Epub 2014 Apr 25.
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- STI 06/16
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Descripción del plan IPD
Marco de tiempo para compartir IPD
Criterios de acceso compartido de IPD
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .
Ensayos clínicos sobre Glaucoma diagnostics
-
Mansoura UniversityTerminadoSatisfacción del paciente | Resultados clínicos | Complicación protésicaEgipto
-
North Bristol NHS TrustEKF DiagnosticsRetiradoMedición de lactato en sangre fetalReino Unido
-
Medfield DiagnosticsTerminado
-
National Institute for Tuberculosis and Lung Diseases...Healthup Sp. z o.o.Terminado
-
Stanford UniversitySanta Clara Valley Medical CenterRetiradoSífilis congénita | Sífilis materna durante el embarazo: el bebé aún no ha nacidoEstados Unidos
-
Haukeland University HospitalSiemens Corporation, Corporate TechnologyTerminado
-
Center for Eye Research AustraliaNovartis PharmaceuticalsTerminadoDegeneración macular | Neovascularización coroideaAustralia
-
Polish Society of Disaster MedicineActivo, no reclutando
-
McMaster UniversityMerck Frosst Canada Ltd.; Public Health Agency of Canada (PHAC); Juravinski Cancer...DesconocidoNeoplasia intraepitelial cervical | Infección por el virus del papiloma humanoCanadá
-
Steno Diabetes Center CopenhagenHvidovre University HospitalTerminado