Estudio de validación clínica del sistema de citología cervical Techcyte SureView
Estudio clínico multicéntrico de dos brazos que evalúa el rendimiento del sistema de citología cervical Techcyte Sureview y sus subsistemas para la evaluación de pruebas de Papanicolaou
The clinical study is a multi-center, retrospective, blinded, matched-pair, two-arm clinical study in which one test assessment and a control assessment of liquid-based cytology (LBC) pap tests are performed and compared to a reference.
The control method is defined as the assessment of LBC pap tests performed by a cytotechnologist (CT) and/or cytopathologist (CP) using standard laboratory cervical cytology practices (i.e. a manual assessment) and following The Bethesda System for Reporting Cervical Cytology (TBS). The control method is referred to as the "standard method".
The test method is referred to as the AI-assisted method. The AI-assisted method is defined as the assessment of the same LBC pap test as in the standard method, but with the Techcyte SureView Cervical Cytology System, which has pap tests being scanned on the Pramana SpectralHT Cubiq whole slide scanner, analyzed with the Techcyte Cervical Cytology Algorithm (TCCA) and presented on the Techcyte Viewer, using the Dell U3223QE Monitor for assessment by a CT/CP following the TBS.
In addition, the time spent reviewing the samples with both methods will be measured such that the CT workload limit for AI-assisted assessment can be established.
The LBC pap tests to be assessed in the clinical study must have been prepared with either ThinPrep® Pap Test (Hologic) or BD SurePath™ (Beckton, Dickinson and Company). The study aims to enroll for each preparation method an equal number of samples.
Descripción general del estudio
Estado
Condiciones
Tipo de estudio
Inscripción (Actual)
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Utah
-
Orem, Utah, Estados Unidos, 84057
- Techcyte
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Método de muestreo
Población de estudio
Samples collected in the United States
Descripción
Inclusion Criteria:
- Glass slides with cervical cytology specimens prepared with Hologic ThinPrep according to the pap test manufacturers Instructions for Use (IFU) using any of Hologic's processors (i.e. ThinPrep2000, ThinPrep 3000, ThinPrep 5000, ThinPrep Genesis). OR Glass slides with cervical cytology specimens prepared with BD SurePath according to the pap test manufacturers IFU using any of BD's processors (i.e. BD Totalys MultiProcessor, BD Totalys SlidePrep, BD PrepStain).
- Slide has the original diagnosis available.
- Slides that fit within the required sample size strata using the sample collection procedure.
- Glass slides with QR code, barcodes, or plain text labels.
- Glass slides with coverslips.
- Slides that are 5 years old or less, except where rare specimens require it. Those specimens will have to be reviewed for degradation per the sample collection procedure.
- Only one sample per patient and all are to be collected, processed and scanned.
Exclusion Criteria:
- Broken or cracked slides.
- Slides with air bubbles or scratched, cracked or broken/missing cover slips.
- Slides with markings (e.g., handwritten pen marking) that cannot be removed according to standard laboratory procedures.
- Degraded slides (ex: slides whose stains have lost color or specimens that are not intact).
- Glass slides with cervical cytology specimens prepared with methods other than Hologic ThinPrep or BD SurePath.
- Glass slides with specimens other than cervical cytology specimens.
- Slides from the same patient.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Modelos observacionales: OTHER
- Perspectivas temporales: RETROSPECTIVE
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Sensitivity and specificity when comparing pap tests using the Techcyte SureView Cervical Cytology System
Periodo de tiempo: 1 day
|
The primary objective of the study is to show that the assessment of LBC pap tests using the Techcyte SureView Cervical Cytology System, i.e. the AI-assisted method, is non-inferior to current standard practice, i.e., the standard method, at the LSIL+ threshold used for cytologic diagnosis (specimen adequacy and descriptive diagnosis) as defined by The Bethesda System for Reporting Cervical Cytology. |
1 day
|
Colaboradores e Investigadores
Patrocinador
Publicaciones y enlaces útiles
Publicaciones Generales
- Nayar R, Wilbur DC. (eds), The Bethesda System for Reporting Cervical Cytology: Definitions, Criteria, and Explanatory Notes. 3rd ed. Cham, Switzerland: Springer: 2015.
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Otros números de identificación del estudio
- 1-VAL-3427
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
producto fabricado y exportado desde los EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .