The eValuation of Image Guidance of Stereotactic Body Radiotherapy Using Ultrasound: Assessment in Liver (VISUAL)
연구 개요
상세 설명
PURPOSE: This observational study of healthy volunteers will test the use of a safe, non-ionising and non-invasive imaging system for the purpose of improving radiotherapy treatment of the liver. This study investigates an image guidance method which may be used to guide liver radiotherapy and enable more accurate treatment on a conventional radiotherapy machine.
RECRUITMENT: Volunteers will need to provide their National Health Service (UK) numbers and be registered on the hospital information system. Volunteers need to be aware that if they are patients at the Royal Marsden Hospital this may raise confidentiality issues. Volunteers will be made aware of this in the recruiting email and the Volunteer Information sheet.
USE OF HEALTHY VOLUNTEERS: This is an observational study, which tests the use of an ultrasound guidance system. The accuracy of the system will be evaluated by comparison of proprietary methods of liver position estimation against a manually derived ground truth derived from the images of the liver acquired in participants. It is not expected that there are any differences in the ultrasound guidance images of healthy volunteers and patients with liver cancer. Ultrasound is a non-invasive and non-ionizing imaging modality with no known side-effects or complications and therefore, safe to use in a volunteer study. Furthermore, this study requires that participants lie on a hard couch for up to 45 minutes and up to 4 times. This would be challenging for the vast majority of patients with liver cancer who are most likely to be frail due to their disease.
REVIEW OF HEALTHY VOLUNTEERS: The ultrasound guidance system provides 3D ultrasound images of the liver for providing positional information only. It is very unlikely that abnormalities will be detectable on 3D US guidance images. These images are not of diagnostic quality, being lower resolution than routine diagnostic 2D ultrasound images. Furthermore, diagnostic ultrasound of the liver often incorporates ultrasound contrast agent, which will not be used in this study. If a radiographer does observe any abnormality, this will be reported to a Consultant Radiologist and the images will be made available for review.
RISKS, BURDENS and BENEFITS: There is very little risk associated with ultrasound and the Automatic Breathing Control (ABC) device, both pieces of the equipment are CE-marked and we are not using outside of their respective intended purposes. The burden to the volunteer is (1) the time spent, (2) the uncomfortable nature of the couch (it has a hard flat surface) and (3) discomfort using the ABC device. The participant will made aware of (1), (2) and (3) by providing details in the volunteer information sheet. Participants who are uncomfortable using the ABC can opt out and still participate in the study. Participants will be able to get up from the couch and walk around to relive any discomfort at any point during the imaging sessions.
There are no benefits for the participants. It is expected that this study will be of benefit to future patients with liver cancer. To ensure there is no perceived benefit of the acquisition of ultrasound images it is stated clearly in the volunteer information sheet that images are not of diagnostic quality and that persons obtaining the images are trained in therapy guidance and not diagnostics, and that images will only be reviewed by a radiologist in the unlikely event that an abnormality is observed.
CONFLICTS OF INTEREST: There are no conflicts of interest. No study investigator will benefit financially from the outcome of this study. The manufacturer of the Clarity ultrasound system will be informed of the results and all results will be published in a peer-reviewed scientific journal.
ACCESS to DATA: The investigators wish to make anonymised image data available to other researchers on request for the purpose of medical research. This is intended to maximise the value of the data and may provide greater benefit to patients. Volunteers will be asked to consent to their anonymised data being made available to other researchers.
AIMS: This study aims to develop and evaluate the use of 3 ultrasound (3DUS) to estimate the liver position for radiotherapy guidance. This is an exploratory study, where Clarity US of the abdomen will be acquired in healthy volunteers. 3D US-sim, 3D US-guide, 3D US-monitor and 4D US-monitor will be used to meet the following study objectives:
- The accuracy of B-mode 3D US-guide, to estimate the position of the liver in breath-hold for verification of (1) daily set-up position (set-up US-guide) and (2) repeat breath hold relevant to radiotherapy delivery (BH 3D US-monitor), will be determined. Accuracy will be assessed by comparison of US-guide and US-monitor determined liver position with a manually determined 'gold standard' liver position.
- Comparison of the accuracy of power Doppler and B-mode US-guide/US-monitor for the estimation of liver position during breath hold.
- Assessment of the accuracy of 4D US-monitor to estimate liver motion during free breathing (FB).
- Comparison of liver position stability using spirometry and voluntary breath-hold (vBH) using US-monitoring.
- Assessment of the availability of a second or third ultrasound field of view for increasing the accuracy of US-guide based estimates of liver motion.
연구 유형
등록 (예상)
연락처 및 위치
연구 연락처
- 이름: Emma J Harris, PhD
- 전화번호: +44 208 6613320
- 이메일: eharris@icr.ac.uk
연구 연락처 백업
- 이름: Jeff C Bamber, PhD
- 전화번호: +44 915 4562
- 이메일: jbamber@icr.ac.uk
참여기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
샘플링 방법
연구 인구
설명
Inclusion Criteria:
- Healthy adult volunteers.
Exclusion Criteria:
- Healthy volunteers who are unable to hold their breath for 30 seconds in exhalation.
- Healthy volunteers who are unable to lie for extended period of time (up to 45 minutes including short breaks of a few minutes) on a hard surface (treatment couch).
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
코호트 및 개입
그룹/코호트 |
개입 / 치료 |
---|---|
Group A
Volunteers will be asked to receive ultrasound imaging in repeat voluntary breath hold and during free-breathing.
|
Abdominal ultrasound imaging of the liver
|
Group B
Volunteers will be asked to receive ultrasound imaging in repeat voluntary breath hold and during free-breathing and in repeat spirometer assisted breath hold.
|
Abdominal ultrasound imaging of the liver
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
---|---|---|
Difference in liver position after set-up
기간: Within 2 months of image acquisition
|
The mean difference in liver position measured by US-guide and using the gold standard method, measured in millimetres in the superior-inferior (SI), anterior-posterior (AP) and left-right (LR) directions over 5 set-up sessions.
|
Within 2 months of image acquisition
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
---|---|---|
Difference in liver position during repeat breath hold
기간: Within 2 months of image acquisition
|
The mean difference in liver position measured by US-guide and using the gold standard method, measured in millimetres in the superior-inferior (SI), anterior-posterior (AP) and left-right (LR) directions over 3 breath holds.
|
Within 2 months of image acquisition
|
Difference in liver position after set-up using Doppler ultrasound and US-guide
기간: Within 2 months of image acquisition
|
The mean difference in liver position measured by B-mode US-guide and using the Doppler US-guide method, measured in millimetres in the superior-inferior (SI), anterior-posterior (AP) and left-right (LR) directions over 5 set-ups.
|
Within 2 months of image acquisition
|
Liver position during free-breathing
기간: Within 2 months of image acquisition
|
The liver position measured in the SI, AP and LR directions, measured by 4D US-monitor during free breathing.
|
Within 2 months of image acquisition
|
Liver motion between fractions and repeat thresholds
기간: Within 2 months of image acquisition
|
The mean motion of the liver during voluntary breath-hold and assisted breath-hold measured using the gold standard method.
|
Within 2 months of image acquisition
|
Number of available imaging positions
기간: Within 2 months of image acquisition
|
The number of imaging positions (1 2 or 3), which may be used by 4D US-guide to accurately measure liver position during free-breathing.
|
Within 2 months of image acquisition
|
공동 작업자 및 조사자
수사관
- 수석 연구원: Emma J Harris, PhD, Institute of Cancer Research, United Kingdom
연구 기록 날짜
연구 주요 날짜
연구 시작 (예상)
기본 완료 (예상)
연구 완료 (예상)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .
Ultrasound imaging에 대한 임상 시험
-
French Cardiology SocietyInstitut National de la Santé Et de la Recherche Médicale, France완전한
-
University of Pennsylvania종료됨
-
Nigde Omer Halisdemir University완전한
-
Academisch Medisch Centrum - Universiteit van Amsterdam...완전한
-
Piazza della Vittoria 14 Studio Medico - Ginecologia...모병
-
Ramsay Générale de SantéDr François Lintz완전한