Эта страница была переведена автоматически, точность перевода не гарантируется. Пожалуйста, обратитесь к английской версии для исходного текста.

CorPath™ 200: Robotically-Assisted Percutaneous Coronary Intervention (PCI)

6 августа 2015 г. обновлено: Corindus Inc.

CorPath™ 200 System: Coronary Remote Catheterization Feasibility Study

An open-label, prospective, single-arm study is designed to evaluate safety, clinical and technical efficacy of the CorPath 200 System in delivery and manipulation of the coronary guide wires and balloon/stent systems for use in robotically-assisted, percutaneous coronary intervention (PCI) procedures.

Обзор исследования

Статус

Завершенный

Подробное описание

Objective The study objective was to evaluate the safety and technical efficacy of the CorPath 200 System in delivery and manipulation of commercially available coronary guidewires and rapid exchange balloon/stent catheter for use in percutaneous coronary interventions (PCI).

I. Primary Endpoints A. Efficacy Device Clinical Success: Achievement of a < 30% final diameter stenosis after utilizing the CorPathTM 200 System to deliver a PTCA balloon and then stent to the target lesion, and successfully retract the devices without the occurrence of any In-Hospital MACE (Major Adverse Cardiac Event) evaluated at 48 hours post procedure or hospital discharge, whichever occurs first.

B. Safety In-Hospital MACE: Defined as the composite of death (all cause mortality), recurrent MI, and target vessel revascularization (emergent coronary artery bypass surgery [CABG] or PCI) evaluated at 48 hours post procedure or hospital discharge, whichever occurs first.

II. Secondary Endpoints A. Efficacy CorPath Success: Ability to navigate the guide wire and balloon/stent system utilizing the CorPath™ device to the target site without the occurrence of any In-Hospital MACE, evaluated at 48 hours post index procedure or hospital discharge, whichever occurs first.

Subject Clinical Success: Achievement of a <30% final diameter stenosis of target lesion after treatment with PTCA/coronary stent without the occurrence of any In-Hospital MACE, evaluated at 48 hours post-index procedure or hospital discharge, whichever occurs first.

B. Safety:

MACE at 30 Days: Defined as the composite of death, recurrent MI and/or target vessel revascularization (emergent coronary artery bypass surgery [CABG] or PCI) evaluated at 30 days post index procedure.

Adverse Events:

A summary of all adverse events observed in the trial summarized as either serious or non-serious and summarized by attribution.

C. Clinical Utility

Procedure Attributes:

  • Reduction in radiation exposure to the primary operator based on radiation dosimetry measurements at the procedure table and on a primary operator.
  • Total Procedure time based on the time of insertion of hemostasis sheath through the time of final guide catheter withdrawal.

CorPath Device Attributes:

Subjective assessment by the operator of the following device performance attributes:

  • Movement of the guide wire (rotating and advancing) to the target.
  • Maneuvering and position of the guide wire to the target.
  • Movement of the stent/balloon (advancing) to the target.
  • Ability to deliver the PTCA/stent system to the intended target lesion.
  • Ease of use of the system by the interventional team.

Material & Methods Corindus obtained approval (from the CORBIC Ethics Committee and INVIMA - Colombian Ministry of Health) in 2009 to conduct the clinical study "CorPath™ 200 System: Coronary Remote Catheterization Feasibility Study" at CORBIC Cardio-Neuro-Vascular Institute, Envigado, Colombia.

The study is a single arm, open label, prospective registry to treat a de novo lesion (2.5 mm - 4.0 mm in diameter, 25.0 mm in length with >50% stenosis) utilizing the CorPath 200 System to deliver, manipulate and retrieve a guidewire and a balloon/stent catheter system.

Тип исследования

Интервенционный

Регистрация (Действительный)

8

Фаза

  • Фаза 2

Контакты и местонахождение

В этом разделе приведены контактные данные лиц, проводящих исследование, и информация о том, где проводится это исследование.

Места учебы

      • Medellin, Колумбия
        • CORBIC Cardio-Neuro-Vascular Institute

Критерии участия

Исследователи ищут людей, которые соответствуют определенному описанию, называемому критериям приемлемости. Некоторыми примерами этих критериев являются общее состояние здоровья человека или предшествующее лечение.

Критерии приемлемости

Возраст, подходящий для обучения

18 лет и старше (Взрослый, Пожилой взрослый)

Принимает здоровых добровольцев

Нет

Полы, имеющие право на обучение

Все

Описание

General Inclusion Criteria:

  • The subject is an acceptable candidate for PCI.
  • The subject must have clinical evidence of ischemic heart disease or a positive functional study.
  • Female subjects must be of non-child bearing potential, or if able to bear children, have a negative pregnancy test within seven (7) days prior to index procedure.
  • The subject or the legal representative has been informed of the nature of the study and agrees to its provisions and has provided written informed consent.

Angiographic Inclusion Criteria:

  • Study lesion is de novo native coronary artery lesion (i.e., a coronary lesion not previously treated)
  • The study lesion is intended to be treated with one stent. Maximum stent length allowed is 20 mm.
  • The study reference vessel diameter is between 2.50 mm and 4.5 mm by visual estimate.
  • Study lesion diameter showing significant stenosis of at least 50% by visual estimate.

General Exclusion Criteria:

  • Subject requires planned PCI or CABG within 30 days following the index procedure.
  • Subject has evolving ST elevation myocardial infarction (STEMI) (i.e., beginning of MI symptoms within 72 hours prior to the planned index procedure).
  • Subject has documented left ventricular ejection fraction < 30%.
  • Subject has known hypersensitivity or contraindication to aspirin, heparin, ticlopidine, clopidogrel, stainless steel, or a sensitivity to contrast media, which cannot be adequately pre-medicated.
  • Subject has a platelet count < 100,000 cell/mm3 or > 700,000 cell/ mm3, a WBC of < 3000 cell/ mm3 (e.g. thrombocytopenia, thrombocythemia, neutropenia or leucopenia).
  • Subject has a history of a stroke (CVA) or TIA within 30-days prior to planned index procedure.
  • Subject has an active peptic ulcer or upper GI bleeding within the 6 months prior to planned index procedure.
  • Subject has a history of bleeding diathesis or coagulopathy or will refuse blood transfusions.
  • Subject is currently participating in another investigational drug l drug or device trial that has not completed the entire follow up period.

Angiographic Exclusion Criteria:

  • Study lesion that cannot be fully covered by a single stent of maximal length
  • Subject requires treatment of multiple lesions in the study vessel at the time of index procedure.
  • The study lesion requires planned treatment with DCA, laser, rotational atherectomy, or any device except for balloon dilatation prior to stent placement.
  • The study vessel has evidence of intraluminal thrombus or moderate to severe tortuosity (> 90o) proximal to the target lesion.
  • The study lesion has any of the following characteristics:

    1. Total occlusion
    2. Ostial location
    3. Involves a side branch > 2.0 mm vessel diameter
    4. Is located at >45° bend in the vessel
    5. Is moderately to severely calcified
    6. Moderate-to-severe calcification at the part of the vessel prior to target lesion
  • Unprotected left main coronary artery disease (an obstruction greater than 50% diameter stenosis in the left main coronary artery).

Учебный план

В этом разделе представлена ​​подробная информация о плане исследования, в том числе о том, как планируется исследование и что оно измеряет.

Как устроено исследование?

Детали дизайна

  • Основная цель: Уход
  • Распределение: Н/Д
  • Интервенционная модель: Одногрупповое задание
  • Маскировка: Нет (открытая этикетка)

Оружие и интервенции

Группа участников / Армия
Вмешательство/лечение
Экспериментальный: CorPath 200 System
Robotic-assisted PCI with the CorPath 200
CorPath™ 200 robotically-assisted percutaneous coronary intervention
CorPath PCI - robotic-assisted percutaneous coronary intervention

Что измеряет исследование?

Первичные показатели результатов

Мера результата
Мера Описание
Временное ограничение
Clinical Procedural Success
Временное ограничение: 48-hrs or hospital discharge, whichever occurs first
The percentage of Participants with <30% final diameter stenosis of the target lesion without in-hospital major adverse cardiovascular events (MACE) (defined as the composite of death, recurrent MI, and target vessel revascularization)
48-hrs or hospital discharge, whichever occurs first
Procedural Technical Success
Временное ограничение: Intervention
Successful robotic delivery and retraction of all PCI devices during CorPath PCI procedure.
Intervention

Соавторы и исследователи

Здесь вы найдете людей и организации, участвующие в этом исследовании.

Спонсор

Следователи

  • Главный следователь: Juan F Granada, MD, CORBIC and CRF

Даты записи исследования

Эти даты отслеживают ход отправки отчетов об исследованиях и сводных результатов на сайт ClinicalTrials.gov. Записи исследований и сообщаемые результаты проверяются Национальной медицинской библиотекой (NLM), чтобы убедиться, что они соответствуют определенным стандартам контроля качества, прежде чем публиковать их на общедоступном веб-сайте.

Изучение основных дат

Начало исследования

1 февраля 2010 г.

Первичное завершение (Действительный)

1 мая 2010 г.

Завершение исследования (Действительный)

1 мая 2010 г.

Даты регистрации исследования

Первый отправленный

24 февраля 2010 г.

Впервые представлено, что соответствует критериям контроля качества

24 февраля 2010 г.

Первый опубликованный (Оценивать)

25 февраля 2010 г.

Обновления учебных записей

Последнее опубликованное обновление (Оценивать)

7 сентября 2015 г.

Последнее отправленное обновление, отвечающее критериям контроля качества

6 августа 2015 г.

Последняя проверка

1 августа 2015 г.

Дополнительная информация

Термины, связанные с этим исследованием

Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .