Observation of ImageReady™ MR Conditional Defibrillation System in China
Observation of ImageReady™ MR Conditional Defibrillation System in China (MR ICD)
Study Overview
Status
Status
Conditions
Conditions
Intervention / Treatment
Intervention / Treatment
Detailed Description
Study Type
Study Type
Enrollment (Actual)
Enrollment
Phase
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Shanghai
-
Shanghai, Shanghai, China, 200032
- Zhongshan Hospital Fudan University
-
-
Participation Criteria
Eligibility Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Subject is indicated to Class I/II indications per guidelines/consensus released by Chinese Society of Cardiac Pacing and Electrophysiology
- Subject must have the ImageReady System as their initial (de novo) defibrillation system implant
- Subject will receive an ICD or CRT-D pulse generator in the left or right pectoral region ICD
- Subject is able and willing to undergo an MR scan
- Subject is willing and capable of providing informed consent and participating in all testing/ visits associated with this clinical study at an approved clinical study center and at the intervals defined by this protocol
- Subject is age 18 or above
Exclusion Criteria:
- Subject has other active or abandoned implanted cardiac rhythm devices, components or accessories present such as pulse generators, leads, lead adaptors or extenders
- Presence of metallic objects that represent a contraindication to MR imaging at the discretion of the Radiologist and impacting the ability to conduct the study protocol
- Subject needs or will need a medically necessary MR scan, before completing the 1-month post-MR follow-up visit
Subject with:
- A history of syncope related to brady-arrhythmia
- A history of syncope of unknown etiology
- Sinus pauses (Pause > 2 s)
- Permanent or intermittent complete AV block
- Documentation of progressive AV nodal block over time
Trifascicular block (alternating bundle branch block or PR > 200 ms with LBBB or other bifascicular block)
- Note: It is required to run a 12 lead ECG and a 10s rhythm strip to document this exclusion criterion.
- Subject is not clinically capable of tolerating the absence of pacing or Resynchronization therapy support in a supine position for the duration that the pulse generator is in MRI Protection Mode, per Physician discretion
- Subject is not clinically capable of tolerating the absence of Tachycardia therapy support for the duration that the pulse generator is in MRI Protection Mode, per Physician discretion
- Subjects currently requiring dialysis
- Subject has a mechanical heart valve
- Subject has a known or suspected sensitivity to dexamethasone acetate (DXA)
- Subject is currently on the active heart transplant list
- Subject has documented life expectancy of less than 12 months
- Subject is enrolled in any other concurrent study that might interfere with this study
- Women of childbearing potential who are or might be pregnant at the time of this study
Study Plan
How is the study designed?
Design Details
- Primary Purpose: TREATMENT
- Allocation: NA
- Interventional Model: SINGLE_GROUP
- Masking: NONE
Number of Arms
Arms and Interventions
Participant Group / ArmParticipant Group / Arm |
Intervention / TreatmentIntervention / Treatment |
|---|---|
|
EXPERIMENTAL: ImageReady™ MR Conditional Defibrillation System
|
Subject will receive an ICD or CRT-D pulse generator in the left or right pectoral region ICD
|
What is the study measuring?
Primary Outcome Measures
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The Percent of Participants Who Were MR Scan-Related ImageReady System Complication -Free
Time Frame: MRI + 1 Month Visit(10-13 weeks from 0 day)
|
The primary safety endpoint of the MR ICD study will be assessed for all subjects who undergo any portion of the study-required MR scan sequences.
Safety will be confirmed by evaluating the MR scan related ImageReady System complication-free rate (CFR) between the MR Scan and the MRI Visit + 1 Month.MR scan related ImageReady System complication is relation with ImageReady System and MR Scan.
|
MRI + 1 Month Visit(10-13 weeks from 0 day)
|
|
The Percentage of Participant Whose the Average RV Shocking Impedance is >200 Ohm 1 Month Post Scan, While it is ≤ 200 Ohm Before Scan
Time Frame: MRI+1 Month visit ( 10~13 weeks from 0 day)
|
The normal RV shocking impedance measured by the system should be≤200 Ohm, with >200 Ohm considered abnormal.
The primary effectiveness endpoint 1 is defined as that the average RV shocking impedance is >200 Ohm at 1 month post scan, while it is ≤ 200 Ohm before scan.There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 1 occur.
|
MRI+1 Month visit ( 10~13 weeks from 0 day)
|
|
The Percentage of Participant That Have an Increase in Average RV Pacing Thresholds > 0.5V (at 0.5 ms) From Pre-MR Scan to MRI Visit + 1 Month Follow-up
Time Frame: MRI+1Month visit( 10~14 weeks from 0 day)
|
Subjects that have an increase in average RV pacing thresholds ≤ 0.5V (at 0.5 ms) from pre-MR Scan to MRI Visit + 1 Month follow-up will be considered a success, otherwise, it is a primary effectiveness endpoint 2 event.The performance goal of this endpoint for the ENABLE MRI study was 87%.
There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 2 occur.
|
MRI+1Month visit( 10~14 weeks from 0 day)
|
|
The Percentage of Participant That Have the Average RV Sensed Amplitude at the MRI + 1 Month Visit Remains < 5.0 mV or Less Than 50% of the Pre-MR Scan Value
Time Frame: MR+1 Month visit( 10~14 weeks from 0 day)
|
Subjects will be considered a success if the average sensed amplitude at the MRI + 1 Month Visit remains ≥ 5.0 mV and above 50% of the pre-MR scan value, otherwise, it is a primary effectiveness endpoint 3 event.The performance goal of this endpoint for the ENABLE MRI study was 85%.
There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 3 occur.
|
MR+1 Month visit( 10~14 weeks from 0 day)
|
|
The Percentage of Participant Have an Increase in Average LV Pacing Thresholds >1.0V (at 0.5 ms) From Pre-MR Scan to MRI Visit + 1 Month Follow-up
Time Frame: MR +1 Month visit( 10~14 weeks from 0 day)
|
Subjects that have an increase in average LV pacing thresholds ≤1.0V (at 0.5 ms) from pre-MR Scan to MRI Visit + 1 Month follow-up will be considered a success, otherwise, it is a primary effectiveness endpoint 4 event.The performance goal of this endpoint for the ENABLE MRI study was 87%.
There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 4 occur.
|
MR +1 Month visit( 10~14 weeks from 0 day)
|
|
the Percentage of Participant That Have the Average LV Sensed Amplitude at the MRI + 1 Month Visit Remains <5.0 mV or Above 50% of the Pre-MR Scan Value
Time Frame: MR +1 Month visit( 10~14 weeks from 0 day)
|
Subjects will be considered a success if the average sensed amplitude at the MRI + 1 Month Visit remains ≥ 5.0 mV and above 50% of the pre-MR scan value, otherwise, it is a primary effectiveness endpoint 5 event.The performance goal of this endpoint for the ENABLE MRI study was 85%.
There is no hypothesis testing in the MR ICD study, and it can be considered reasonable if up to 2 failed cases of primary effectiveness endpoint 5 occur.
|
MR +1 Month visit( 10~14 weeks from 0 day)
|
Collaborators and Investigators
Sponsor
Sponsor
Investigators
Investigators
- Principal Investigator: Yangang Su, Doctor, Fudan University
Publications and helpful links
General Publications
- Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, Caffo BS, Frick KD, Kraut MA, Kamel IR, Calkins H, Berger RD, Bluemke DA, Halperin HR. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011 Oct 4;155(7):415-24. doi: 10.7326/0003-4819-155-7-201110040-00004.
- Russo RJ, Costa HS, Silva PD, Anderson JL, Arshad A, Biederman RW, Boyle NG, Frabizzio JV, Birgersdotter-Green U, Higgins SL, Lampert R, Machado CE, Martin ET, Rivard AL, Rubenstein JC, Schaerf RH, Schwartz JD, Shah DJ, Tomassoni GF, Tominaga GT, Tonkin AE, Uretsky S, Wolff SD. Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. N Engl J Med. 2017 Feb 23;376(8):755-764. doi: 10.1056/NEJMoa1603265.
- ASTM Standard F2503-08, "Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment" ASTM International, West Conshohocken, PA, 2008, DOI: 10.1520/F2503-08, www.astm.org..
- Chinese Medical Association ECG Physiology and Pacing Branch, Chinese Medical Association Cardiology Branch, Chinese Medical Association Cardiology Professional Committee implanted cardioverter defibrillator treatment. Implanted cardioverter defibrillator treatment of China Expert consensus. Chinese Journal of Cardiac Arrhythmias 2014;18:242-53.
- Shu Zhang, Dejia Huang, Wei Hua, et al. Advice on cardiac resynchronization therapy for chronic heart failure (revised in 2013). Chinese Journal of Cardiac Arrhythmias 2013;17:247-61.
- Wolf SM, Lawrenz FP, Nelson CA, Kahn JP, Cho MK, Clayton EW, Fletcher JG, Georgieff MK, Hammerschmidt D, Hudson K, Illes J, Kapur V, Keane MA, Koenig BA, Leroy BS, McFarland EG, Paradise J, Parker LS, Terry SF, Van Ness B, Wilfond BS. Managing incidental findings in human subjects research: analysis and recommendations. J Law Med Ethics. 2008 Summer;36(2):219-48, 211. doi: 10.1111/j.1748-720X.2008.00266.x.
- Kanal E, Barkovich AJ, Bell C, Borgstede JP, Bradley WG Jr, Froelich JW, Gilk T, Gimbel JR, Gosbee J, Kuhni-Kaminski E, Lester JW Jr, Nyenhuis J, Parag Y, Schaefer DJ, Sebek-Scoumis EA, Weinreb J, Zaremba LA, Wilcox P, Lucey L, Sass N; ACR Blue Ribbon Panel on MR Safety. ACR guidance document for safe MR practices: 2007. AJR Am J Roentgenol. 2007 Jun;188(6):1447-74. doi: 10.2214/AJR.06.1616. No abstract available.
Study record dates
Study Major Dates
Study Start (ACTUAL)
Study Start
Primary Completion (ACTUAL)
Primary Completion
Study Completion (ACTUAL)
Study Completion
Study Registration Dates
First Submitted
First Submitted
First Submitted That Met QC Criteria
First Submitted That Met QC Criteria
First Posted (ACTUAL)
First Posted
Study Record Updates
Last Update Posted (ACTUAL)
Last Update Posted
Last Update Submitted That Met QC Criteria
Last Update Submitted That Met QC Criteria
Last Verified
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
Other Study ID Numbers
- C2082
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Arrhythmia
-
NCT06082388Not yet recruitingCardiac Arrhythmia | Supraventricular Arrhythmia
-
NCT00993382CompletedVentricular Arrhythmia | Arrhythmia Prophylaxis
-
NCT07353710Not yet recruitingAtrial Arrhythmia | Intracardiac Echocardiography
-
NCT00000476CompletedHeart Diseases | Cardiovascular Diseases | Heart Failure | Arrhythmia | Heart Failure, Congestive | Sinus Arrhythmia
-
NCT06384963RecruitingHeart Failure With Reduced Ejection Fraction
-
NCT06359938RecruitingHeart Failure With Reduced Ejection Fraction
-
NCT05247320CompletedAtrial Arrhythmia
-
NCT02245009UnknownAtrial Arrhythmia
Clinical Trials on ImageReady™ MR Conditional Defibrillation System
-
NCT02652481CompletedMagnetic Resonance Imaging
-
NCT02636712CompletedVasovagal Syncope | Arhythmia | Atrioventricular Heart-block | Chronic Bundle Branch or Branch Block | Carotid Sinus Hypersensitivity Reaction Syndrome
-
NCT03884231CompletedParkinson Disease | Movement Disorders | Dystonia | Essential Tremor | Tremor
-
NCT02849769Completed
-
NCT01781078CompletedBradycardia | Sinus Node Dysfunction
-
NCT06675448RecruitingHeart Failure | Mitral Regurgitation | Valvular Heart Disease
-
NCT02043535TerminatedCoronary Artery Disease