50/50% Tilt and Tuned Defibrillation Waveform

February 1, 2019 updated by: Abbott Medical Devices

50/50% Tilt and Tune Defibrillation Waveforms Utilizing High Voltage Lead Integrity Check Impedance

Study Description The purpose of this study is to compare the defibrillation efficacy between the 50/50% tilt biphasic waveform and the Tuned biphasic waveform utilizing High Voltage Lead Integrity Check (HVLIC) impedance.

Hypothesis The Tuned waveform reduces defibrillation thresholds (DFTs) when compared to the 50/50% tilt waveform.

Study Methods

  • This is an acute, paired-sample, randomized (waveform testing sequence) study
  • Patients are implanted with an Food and Drug Administration (FDA) approved St. Jude Medical (SJM) EpicTM+ DR/VR Implantable Cardioverter Defibrillator (or any other standard output ICD after Epic+) and defibrillation lead system.
  • All study testing is completed at implant.
  • Patients are randomized to begin DFT testing with either the 50/50% tilt or Tuned waveform.
  • The HVLIC feature is used to measure the high voltage lead impedance and to program the Tuned and 50/50% tilt waveform pulse widths.
  • The DFT is determined by an optimized binary search method that is designed to reduce the overall number of VF inductions to an average of 5.13 while maintaining very high resolution and accuracy.
  • Total # of centers - 15 centers
  • Sample size - 60 patients

Study Overview

Status

Completed

Conditions

Study Type

Interventional

Phase

  • Not Applicable

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

N/A

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Patient is a candidate for ICD implantation.
  • Patient is able to tolerate DFT testing.

Exclusion Criteria:

  • Patient has a mechanical valve in the tricuspid position.
  • Patient is pregnant.
  • Patient is less than 18 years old.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: OTHER
  • Allocation: RANDOMIZED
  • Interventional Model: PARALLEL

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
OTHER: 50/50% Tilt

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

January 1, 2005

Primary Completion (ACTUAL)

November 1, 2005

Study Completion (ACTUAL)

November 1, 2005

Study Registration Dates

First Submitted

May 16, 2007

First Submitted That Met QC Criteria

March 28, 2011

First Posted (ESTIMATE)

March 30, 2011

Study Record Updates

Last Update Posted (ACTUAL)

February 4, 2019

Last Update Submitted That Met QC Criteria

February 1, 2019

Last Verified

February 1, 2019

More Information

Terms related to this study

Other Study ID Numbers

  • CRD 274

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on DFT Testing

Clinical Trials on DFT Testing with a 50/50% tilt

Subscribe