Revised Algorithm for Next Generation DeVilbiss Healthcare AutoAdjust Continuous Positive Airway Pressure (CPAP)

May 7, 2015 updated by: DeVilbiss Healthcare LLC

Revised Algorithm for Next Generation DeVilbiss AutoAdjust CPAP

The purpose of this protocol is to improve the current FDA-approved continuous positive pressure auto-titration algorithm, including treatment precision, residual disease detection characteristics and phenotyping with the DeVilbiss brand AutoAdjust CPAP device.

Study Overview

Status

Completed

Study Type

Interventional

Enrollment (Actual)

33

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Massachusetts
      • Newton, Massachusetts, United States, 02459
        • Neurocare, Inc.

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

21 years to 75 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Apnea hypopnea index ≥ 15 / hour of sleep on the diagnostic polysomnogram, or diagnostic component of a split night polysomnogram
  • Sleep efficiency reported on CPAP titration night PSG should be > 78%
  • Diagnosis of moderate to severe obstructive sleep apnea (AHI ≥15) or central/complex sleep apnea
  • On stable CPAP or APAP [Automatic Positive Airway Pressure] therapy, with compliant use averaging 4 or more hours nightly
  • Age 21-75 years
  • Predominately central sleep apnea for up to 10 subjects only
  • Willingness to give written informed consent and ability to adhere to visit schedule

Exclusion Criteria:

Diagnosis of mild sleep apnea

  • CPAP naïve patients
  • Co-morbid conditions which render participation to be at risk. These may include patients with congestive heart failure, COPD [Chronic Obstructive Pulmonary Disease] or psychiatric illness
  • Allergies to mask materials
  • Evidence of another primary sleep disorder
  • Evidence of arousing periodic limb movements during titration
  • Contraindications as listed on product labeling
  • Pregnant
  • Currently diagnosed with depression if symptomatic
  • Deemed medically unsuitable by investigator
  • Evidence of any type of infection or treatment of an infectious condition during the period of research participation
  • Have a bi-level requirement, or prescribed a bi-level therapy device
  • CPAP pressure >15cmH2O
  • Subjects with tracheotomy
  • Uncontrolled hypertension
  • Require supplemental oxygen
  • Stimulants, major tranquillizers or antipsychotics
  • Restless Leg Syndrome or Periodic Leg Movement Index of > 5 per hour

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: All subjects
All subjects underwent a sleep study with DeVilbiss AutoAdjust CPAP with revised algorithm simultaneously with hand-scored PSG.
Subjects attended a one-night polysomnogram (PSG) titration study. DeVilbiss AutoAdjust CPAP with a revised algorithm was compared to hand-scored PSG to determine whether the machine results were equivalent to hand scores. Each subject acted as their own control for assessing accuracy of sleep event analysis and prescribed CPAP pressure.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
% Agreement Between Reviewer and Machine Pressure Settings During Sleep Study
Time Frame: 1 night
The primary objective for the study was demonstrating effectiveness of therapy provided by the DeVilbiss AutoAdjust device as reported by the expert human reviewer. The primary endpoint was that expert human reviewer should agree with pressure changes made by the machine during each 20-minute epoch at least 80% of the time. Expert human reviewer reviewed each study and made a determination at 20-minute time points, in context of REM/NON-REM sleep and supine/non-supine positions, marking the machine pressure response with an "agree" or "disagree." Pressure changes might show no change, increase or decrease during the 20-minute epoch.
1 night

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

January 1, 2013

Primary Completion (Actual)

May 1, 2014

Study Completion (Actual)

May 1, 2014

Study Registration Dates

First Submitted

January 8, 2013

First Submitted That Met QC Criteria

February 21, 2013

First Posted (Estimate)

February 22, 2013

Study Record Updates

Last Update Posted (Estimate)

May 29, 2015

Last Update Submitted That Met QC Criteria

May 7, 2015

Last Verified

May 1, 2015

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.