IVUS Controlled Stenting

October 14, 2016 updated by: Esther Zhou, Sydney South West Area Health Service

Expansion of Intracoronary Stents by Low Pressure Oversizing Versus High Pressure Inflation: A Randomised Intracoronary Ultrasound Controlled Study

Stent placement is now widely accepted to improve the results of angioplasty (an operation to widen the blood vessel) and decreasing the need for further surgery. Despite their worldwide acceptance, stent usage is still limited by renarrowing which occurs within the stent (restenosis) in some patients within the first six months. In addition there still exists a small risk of heart attacks soon after the stent is placed.

The purpose of this study is to test the safety and tolerability of trying to place the stents using a different strategy of using low pressure oversized balloon inflations as opposed to the usual high pressure balloon inflations. This may reduce injury to the artery and reduce the chance of renarrowing of the stent. In order to ensure this is a safe and effective way of performing the procedure the investigators intend to use an Intravascular ultrasound (IVUS) catheter to look from inside the artery at the stents. IVUS has proven to be the best way of ensuring optimal stent placement.

Primary hypothesis: Stents will be equally well expanded and apposed using a strategy of oversized stenting at normal inflation pressures ( < 10 atmospheres ) as compared to high pressure inflation (≥14 atmospheres) as guided by intravascular ultrasound imaging.

Secondary Hypothesis: There will be no difference in acute clinical endpoints (death, myocardial infarction, urgent revascularisation or stent thrombosis) using a strategy of oversized stenting at normal inflation pressure as compared to high pressure inflation.

Tertiary Hypothesis: If the above is shown to be true the investigators would hope to expand the study in order to reveal a decrease in stent restenosis using a strategy of lower pressure balloon inflations. The investigators will assess the deployment characteristics of drug eluting versus bare metal stents

Study Overview

Study Type

Interventional

Enrollment (Anticipated)

40

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • New South Wales
      • Liverpool, New South Wales, Australia, 2170
        • Department of Cardiology

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

14 years to 81 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Patients who are scheduled for a coronary angioplasty and stenting
  • Single de novo artery stenosis are eligible for this study
  • Target lesion needs to be less than 30mm long and located in a vessel more than 2.5 mm in diameter

Exclusion Criteria:

  • Contraindication to antiplatelet therapy;
  • Ostial lesion;
  • Excessive vessel tortuosity;
  • Lesion at a significant bifurcation (subbranch ≥ 2mm in diameter);
  • Suspected intracoronary thrombus.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Active Comparator: Stent oversize group

Oversized stent deployed at low pressure:

A stent premounted on a balloon with a nominal diameter halfway between the lumen diameter and the true vessel diameter (approximated by external elastic lamina) as assessed by IVUS. This will be based on the smallest of the vessel reference diameters proximal and distal to the lesion. In addition the vessel diameter must be greater than this diameter throughout the length of the lesion. This stent will be implanted at an inflation pressure of 10 atmospheres or less for at least 15 seconds and a second IVUS will be performed to assess the end point.

Stent can include both bare metal stents and drug eluting stents. Depending on availabilities, different stent brands may be used.
Active Comparator: High pressure group

Stent deployed at high pressure:

A stent premounted on a balloon with a nominal diameter approximately equal to the vessel segment lumen reference diameter as previously assessed by IVUS will be used. This stent will be implanted at an inflation pressure of 14 atmospheres or more for at least 15 seconds and a second IVUS will be performed to assess the end point

Stent can include both bare metal stents and drug eluting stents. Depending on availabilities, different stent brands may be used.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of stents which are optimally deployed as assessed against the MUSIC criteria
Time Frame: Immediately after stent is deployed until optimal deployment is achieved

The criteria for optimal stent expansion are:

1. Complete apposition of the stent over its entire length against the vessel wall.

2a. In stent minimal luminal area (MLA) >= 90% of the average reference luminal area or >=100% of lumen area of the reference segment with the lowest lumen area.

In stent lumen area of proximal stent entrance >= 90% of proximal lumen area. 2b. Where the instent luminal area is in excess of 9.0 mm2 , the following criteria apply: In stent MLA > 80% of the average reference lumen area or >=90% of lumen area of the reference segment with the lowest lumen area.

In stent lumen area of proximal stent entrance >= 90% of proximal lumen area. 3. Symmetric stent expansion defined by lumen diameter (LD)min / LD max >= 0.7

A less rigorous criteria for stent expansion will also be assessed:

  1. In stent MLA > 80% of the average reference lumen area; &
  2. Complete apposition of the stent over its entire length against the vessel wall
Immediately after stent is deployed until optimal deployment is achieved

Secondary Outcome Measures

Outcome Measure
Time Frame
Clinical endpoints including death, myocardial infarction, coronary artery bypass surgery, repeat percutaneous intervention, stent thrombosis, myonecrosis
Time Frame: 1-3 days following procedure and at 6 months
1-3 days following procedure and at 6 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Craig Juergens, Cardiologist, South West Sydney, Liverpool

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

February 1, 2014

Primary Completion (Actual)

October 1, 2016

Study Completion (Anticipated)

November 1, 2016

Study Registration Dates

First Submitted

April 28, 2014

First Submitted That Met QC Criteria

April 29, 2014

First Posted (Estimate)

May 1, 2014

Study Record Updates

Last Update Posted (Estimate)

October 17, 2016

Last Update Submitted That Met QC Criteria

October 14, 2016

Last Verified

October 1, 2016

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.