- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT02479724
A-line/CareTaker Comparison Study in ICU Patients
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
The object of the work presented here was to validate a new approach to tracking blood pressure that is based on the pulse analysis of the peripheral arterial pressure pulse. The approach, referred to as the Pulse Decomposition Analysis (PDA) model, goes beyond traditional pulse analysis by invoking a physical model that comprehensively links the components of the peripheral pressure pulse envelope with two reflection sites in the central arteries.
In these experiments, approved by the University of Virginia Institutional Review Board, the arterial blood pressures of patients (23 m/11 f, mean age: 44.05 y, SD: 13.9 y, mean height: 173.3 cm, SD: 9.4 cm, mean weight: 95.3 kg, SD: 27.4 kg) hospitalized in University of Virginia Medical Intensive Care Units (MICUs) were monitored using radial intra-arterial catheters, while the CareTaker system collected pulse line shapes at the lower phalange of the thumb of the ipsilateral hand.
Systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements and a Bland-Altman analysis was created. Results were within the limits established for the validation of automatic arterial pressure monitoring (AAMI/ANSI/ISO 81060-1:2013).
Study Type
Enrollment (Actual)
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- male or female age 18 to 45
- presence of a radial arterial catheter
- informed consent by subject or surrogate
- available digital pulse
Exclusion Criteria:
- failure to meet all of the inclusion criteria
Study Plan
How is the study designed?
Design Details
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Performance comparison according to AAMI/ANSI/ISO 81060-1:2013
Time Frame: 40 minutes
|
Correlation and standard deviation between both blood pressure measurement systems were assessed according to 81060 standards.
|
40 minutes
|
Collaborators and Investigators
Collaborators
Investigators
- Principal Investigator: Martin C. Baruch, PhD, Empirical Technologies Corporation
Study record dates
Study Major Dates
Study Start
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Estimate)
Study Record Updates
Last Update Posted (Estimate)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
- ETC_CareTaker_01
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.