Evaluation of Minimed 670G in T1D Patients on Multiple Daily Injection

September 4, 2019 updated by: Goran Petrovski, Sidra Medical and Research Center

Evaluation of Minimed 670G Hybrid Closed Loop System On-Boarding Protocol, for Patients With Type 1 Diabetes on Multiple Daily Insulin Injection Therapy

Introduction. Sensor Augmented Pump has demonstrated superiority over insulin pump and Multiple Daily Injection (MDI) in achieving optimal glucose control and can improve quality of life in Type 1 Diabetes (T1D) patients. Hybrid closed loop (HCL) insulin pump Minimed 670G is a FDA approved device and European Conformity (CE) mark with SmartGuard technology and closed loop algorithm, which will allow the patients to improve their diabetes management. Hybrid closed loop insulin pump Minimed 670G monitors glucose in the subcutaneous tissues and automatically adjusts the delivery of rapid acting insulin as basal rate based on the user's glucose reading. SmartGuard technology in insulin pump, based on user's sensor glucose values can predict when glucose is approaching low levels, 30 minutes in advance and automatically stop insulin delivery. When user's glucose levels recover, SmartGuard will automatically resume insulin delivery. CareLink is personal software, which downloads the data from insulin pump, glucose sensor and glucometer to visualize diabetes information with charts, statistics and events that help patient and health provider to identify and understand patterns and trends The objective of this study is to assess structured group education on boarding protocol of the HCL Minimed 670 G in achieving glucose control of patients on MDI.

Methods. This study is a single-arm, single-center, clinical investigation in subjects with type 1 diabetes on HCL insulin pump (Minimed 670G) in a period of 3 months. A total of 30 subjects (age 6 - 17) will be enrolled in order to reach 26 subjects who will complete the HCL study. The investigators will start the clinical process for initiating an insulin pump, which is typically done with pre-pump classes. HbA1c, derived from CGM will be performed at baseline and 3 months during the study. The following parameters will be analyzed: % patients achieving Time in Range (TIR) > 67% from 70 mg/dl to 180 mg/dl; % patients achieving TIR <3%, below time in range (<70 mg/dl) and % patients achieving both TIR > 67% and <3% time below Range. Collection of demographics and medical history, data for diabetes devices (eg meters, sensors, pumps) and brief clinical physical exam including vital signs and skin assessment will be obtained via Hospital Electronic Medical File (Cerner Millennium, North Kansas City, US) and will be kept as electronic data on a separate research server.

Study Overview

Status

Completed

Detailed Description

This study is a single-arm, single-center, clinical investigation in subjects with Type 1 Diabetes (T1D) on Hybrid Closed Loop (HCL) insulin pump Minimed 670G in a period of 3 months. A total of 30 subjects (age 6 - 17) will be enrolled in order to reach 26 subjects who will complete the HCL study.

After reviewing the patient's eligibility and interest, the investigators will obtain informed consent and assent as appropriate. The investigators will start the clinical process for initiating an insulin pump, which is typically done with pre-pump classes. HbA1c, derived from Continuous Glucose Monitoring (CGM) device will be analyzed at baseline and 3 months during the study.

Collection of demographics and medical history, data for diabetes devices (eg meters, sensors, pumps) and brief clinical physical exam including vital signs and skin assessment will be obtained via Hospital Electronic Medical File (Cerner Millennium, North Kansas City) and will be kept as electronic data on a separate research server.

Base line assessment

The main variables checked at baseline:

  • Age
  • Duration of diabetes
  • Total daily insulin (TDI) dose
  • HbA1c, derived from CGM device

Recruitment process All patients will be recruited during the regular clinic visits at Diabetes Clinics at Sidra Medicine in Doha. The Principal investigator will evaluate the patient if meets the inclusion criteria during the regular clinic visits. A research assistant will take patient and family in other room to explain the study protocol and give patient information. Parents and patients can ask questions about the study. After the explanation of the study protocol, if parents and child agree to participate, they can sign the documents for including in the study. Parents can also take the documents home to consider the possible inclusion in the study. They can also talk to principal investigator before including the child in the study. If parents agree to be a part of the study, they can call research assistant to include the child in the study. The research assistant will arrange a separate meeting to start the process.

Run in period The 2-week run-in period will be used for education and training for the new devices.

Week 1: Pump school (Sunday to Thursday) A group session of 2-3 patients/families will be performed to allow subject to become familiar with new insulin pump. Five sessions in a row (each two hours) will be performed during a week.

Glucose sensor (Suspend before low and suspend on low will be off) will be started on the first day (Same protocol which is currently used in Sidra Medicine).

Topics and education Sunday: Basal Bolus Concept; Operational modes of the pump; Understanding the pump; Pump buttons; Glucose readings; Reading the display; Sensor alerts and alarms; How to insert a sensor; Pump menus Monday: Carelink personal; Battery change; Setting date and time; Basal rates; Pump suspend; Bolus wizard, Bayer Contour Next Link 2.4 Tuesday: Infusion set change; Alarms and alerts; Suspend before low and Suspend of low; Advanced carbs counting Wednesday: Temporary basal; Emergency kit; Low blood sugar; High Blood Sugar; Sick day's management for pump; Diabetic Ketoacidosis (DKA)' Blood glucose Blood ketones; Exercise; Travel Thursday: Auto Mode; Using Auto Mode for the first time Checking Auto Mode Readiness; sensor graph in Auto Mode' Using pump in Auto Mode; Entering a Blood Glucose value in Auto Mode; Suspending and Resuming Delivery in Auto Mode Evaluation Check List; Pump Care; Setting the pump; Sensor start by patient The sensor will be started the first day for teaching and observational purposes only. CareLink account creation and sharing of access is mandatory in the first week.

Week 2: Pump start (Sunday to Sunday) Sensor Augmented Pump will be initiated with both Smart Guard and HCL turned off, to allow the investigators, subjects and device to get more information before starting the HCL. This week will be used as a warm up period for Auto Mode.

Study Period 3 Months Following the two week run-in period using the Minimed 670G, Auto Mode will be activated and all subjects will participate in a 3-month study period.

The patient/family will upload data into the CareLink system and the data will be reviewed by a member of the clinical team. Adjustments to system settings will be suggested to the family as clinically appropriate.

Visit 1- Day 1 In-Clinic visit. Insulin pump will be started in Auto mode. Carelink download for initial pump settings. HbA1c will be obtained.

Visit 2- Day 3 Remote visit. Download pump at home, phone call to patient. Check the Auto Mode, pump settings and fine tune.

Visit 3- Day 7 In-Clinic visit. Download pump. Check the Auto Mode, pump settings and fine tune. Visit 4- Day 14 Remote visit. Download pump at home, phone call the patient. Check the Auto Mode, pump settings and fine tune.

Visit 5- Day 28 In-Clinic visit. Month 1 Download pump. Check the Auto Mode, pump settings and fine tune. Visit 6- Day 42 Remote visit. Download pump at home, phone call to patient. Check the Auto Mode, pump settings and fine tune.

Visit 7- Day 84 In-Clinic visit. Month 3 Download pump. HbA1c will be obtained.

Data Analysis The analysis of the main endpoint 3 month Time in Range, post intervention will be tested by the paired student t-test or paired Wilcoxon test, in case of non-normality of the end point. Evident outliers will be excluded from the main analysis but data on the complete analysis will be provided as well.

The different continuous baseline values will be checked for normality of distribution using the mean-median comparison, skewness and kurtosis (-3 to +3), spread, Kolmogorov-Smirnow and Shapiro-Wild tests, outliers, histograms and Q-Q-plots.

Study Type

Observational

Enrollment (Actual)

30

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Qa
      • Doha, Qa, Qatar, 26999
        • Sidra Medicine

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

6 years to 17 years (Child)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Patients with Type 1 Diabetes on Multiple Daily Injections, motivated and willing to switch Hybrid Closed Loop Minimed 670G.

Description

Inclusion Criteria:

  1. Clinical diagnosis of Type 1 Diabetes. Diagnosis of Type 1 Diabetes is based on the investigator's judgment; C peptide level and antibody determinations are not required.
  2. HbA1c < 12.5%
  3. Age 6-17 years at the initiation of the 670G system
  4. Multiple Daily Injections (Basal Bolus therapy) with Total Daily Insulin use of great than 8.0 units per day over a 1 week period
  5. Willing and able (access to internet from home) to download information into the Medtronic CareLink software
  6. Clinically planning to and be able to start the Medtronic 670G HCL system
  7. History of 3 clinic visits in the last year

Exclusion Criteria:

1. DKA in the 6 months prior to screening visit

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Cohort
  • Time Perspectives: Prospective

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Single Arm Group
Patient on multiple daily injections will start Hybrid Closed Loop System Insulin Pump Minimed 670G
Five Day Group Education on Minimed 670G

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Achieving glucose values more than 67% in Time in Range (70-180 mg/dl)
Time Frame: 3 months
Primary outcome is to achieve glucose levels more than 67% in range between 70mg and 180 mg/dl using Minimed 670G
3 months

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Change in HbA1c
Time Frame: 3 months
Change in HbA1c levels from previous treatment (multiple daily injections) compared to Minimed 670G treatment
3 months
Glucose values above Range (>180 mg/dl)
Time Frame: 3 months
Percentage of glucose values more than 180 mg/dl on Minimed 670G
3 months
Time spend in Auto Mode
Time Frame: 3 months
Percentage of time in Auto Mode, where Minimed 670G automatic adjusts the basal insulin dose according glucose levels
3 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Goran Petrovski, MD, PhD, Sidra Medicine

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

February 1, 2019

Primary Completion (Actual)

August 8, 2019

Study Completion (Actual)

August 11, 2019

Study Registration Dates

First Submitted

November 25, 2018

First Submitted That Met QC Criteria

November 25, 2018

First Posted (Actual)

November 28, 2018

Study Record Updates

Last Update Posted (Actual)

September 6, 2019

Last Update Submitted That Met QC Criteria

September 4, 2019

Last Verified

September 1, 2019

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.