Clinical Outcomes of IVUS-guided Stenting Using the Boston Scientific Vici® Venous Stent System in Treating Patients With Chronic Non-malignant Iliofemoral Vein Obstruction (APAC VICI)

October 25, 2021 updated by: Boston Scientific Corporation

Clinical Outcomes of Intravascular Ultrasound-guided Stenting Using the Boston Scientific Vici® Venous Stent System in Treating Patients With Chronic Non-malignant Iliofemoral Vein Obstruction

The APAC VICI Study is a prospective, non-randomized, multicenter, single-arm study. The purpose of this study is to evaluate the clinical results of using the Boston Scientifc's Vici Venous Stent (Vici stent) to treat significant blockage in the iliofemoral veins under the guidance of intravascular ultrasound (IVUS) in ethnic Chinese patient population. The study will assess the safety and effectiveness of Vici stent implanted under the guidance of the Boston Scientific's IVUS technology (by using Opticross 35 catheter) for treating adult ethnic Chinese patients with symptomatic chronic venous obstruction in the iliofemoral veins.

During the trial, anticoagulant or antiplatlet therapy will be administered for 6 - 12 months post index procedure. Ongoing dynamic data safety monitoring will be performed throughout the trial to minimize subject risk. All enrolled subjects receiving the Vici stent treatment will be followed for 24 months post index procedure.

Study Overview

Study Type

Interventional

Phase

  • Not Applicable

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

14 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Key Inclusion Criteria:

  1. Ethnic Chinese subjects of age ≥ 18 years (or meet age requirements per local law)
  2. Subjects willing to provide written informed consent prior to the study and AND willing to comply with all follow-up evaluations at the specified times
  3. Subjects with presence of unilateral, clinically significant, chronic non-malignant obstruction of the common femoral vein, external iliac vein, common iliac vein, or any combination thereof, defined as a > 50% reduction in target vessel lumen diameter measured by venogram and/or IVUS during procedure
  4. Clinically significant venous obstruction defined as meeting at least one of the following clinical indicators:

    • Clinical severity class of CEAP classification ≥ 3
    • rVCSS Pain Score ≥ 2
  5. Negative pregnancy test in female subjects of potential child-bearing
  6. Intention to stent the target lesion only with the Vici Stent(s)

Key Exclusion Criteria:

Pre-Procedural EC--

  1. Subjects with presence or history of clinically significant pulmonary emboli within 6 months prior to enrollment.
  2. Subjects with venous obstruction that extends into the inferior vena cava
  3. Subjects with contralateral disease of the common femoral vein, external iliac vein, common iliac vein, or any combination thereof with planned treatment within 30 days after enrollment
  4. Subjects with a life expectancy < 12 months
  5. Female of childbearing potential who is pregnant or plans to become pregnant during the duration of the clinical study (Note: subjects who plan to become pregnant after 12-month follow-up visit may be included in the trial at the discretion of treating physician)
  6. A. Subjects with uncontrolled or active coagulopathy OR

    B. Subjects with known uncorrectable bleeding diathesis with the following definitions:

    • Uncorrected INR ≥ 2.0 or aPTT ≥ 1.5X normal local lab value
    • Platelet count < 80,000
  7. Subjects with uncorrected hemoglobin of ≤ 9 g/dL
  8. Subjects with an estimated glomerular filtration rate (eGFR) < 30 mL/min. In patients with diabetes mellitus, eGFR < 45 mL/min.
  9. Subjects have known hypersensitivity to nickel or titanium
  10. Subjects have contrast agent allergy that cannot be managed adequately with pre-medication
  11. Subjects will have intended concurrent thrombolysis or thrombectomy procedure OR intended or planned (within 30 days) adjuvant procedure such as creation of temporary AV fistula, placement of IVC filter, endovenectomy or saphenous vein ablation
  12. Subjects currently participating in another drug or device clinical trial (participation in observational studies is acceptable)
  13. Subjects judged to be a poor candidate by the primary investigator
  14. Subjects who have had any prior surgical or endovascular intervention of the target vessel

Intra-Procedural:

  1. Subjects in whom the lesions cannot be traversed with a guide wire
  2. Subjects where the obstruction extends into the inferior vena cava or below the level of the lesser trochanter
  3. Subjects whose vein diameters are not within limits stated in the Instructions for Use of study device as determined by venogram and/or IVUS
  4. Subjects who do not meet the venogram or IVUS binary stenosis by the definition above (IC3), as determined by the treating physician
  5. Subjects who are judged by treating physician to have a lesion that prevents completed inflation of a balloon dilation catheter or proper placement of the stent or the stent delivery system

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: VICI Stent
The Vici Venous Stent System comprises two components: the implantable endoprosthesis and its stent delivery system, a 9F over-the-wire system available in a 100cm length and compatible with 0.035in (0.89mm) guidewires
The Opticross 35 catheter is designed for use with a guidewire measuring ≤0.035 in (0.89 mm) in diameter and an introducer sheath of 8F (2.67 mm) or larger. The working length of OC35 catheter is 105 cm.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
The primary patency rate at 12 months post index procedure
Time Frame: 12 month
Defined as freedom from occlusion by thrombosis AND freedom from surgical or endovascular intervention on target vessel which are found to have restenosis or stent occlusion to maintain patency AND freedom from in-stent stenosis more than 50% by DUS.
12 month
Major Adverse Events within 30 days post index procedure
Time Frame: 30 days

The composite endpoint of below major adverse events (MAE) adjudicated by the Clinical Events Committee (CEC) within 30 days post index procedure:

  • Device or procedure-related death
  • Device or procedure-related bleeding at the target vessel and/or the target lesion or at the access site requiring surgical or endovascular intervention or blood transfusion ≥ 2 units
  • Device or procedure-related arterial or venous injury occurring in the target vessel segment and/or target lesion or at the access site requiring surgical or endovascular intervention
  • Device or procedure related acute DVT outside of the target vein segment
  • Clinically significant pulmonary embolism defined as being symptomatic with chest pain, hemoptysis, dyspnea, hypoxia etc… AND be documented on CT
  • Stent embolization
  • Stent thrombosis
30 days

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Anticipated)

December 31, 2021

Primary Completion (Anticipated)

December 31, 2022

Study Completion (Anticipated)

December 31, 2023

Study Registration Dates

First Submitted

April 12, 2021

First Submitted That Met QC Criteria

April 20, 2021

First Posted (Actual)

April 23, 2021

Study Record Updates

Last Update Posted (Actual)

November 1, 2021

Last Update Submitted That Met QC Criteria

October 25, 2021

Last Verified

October 1, 2021

More Information

Terms related to this study

Other Study ID Numbers

  • S2484

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

Undecided

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.