A Study of Agreement Between Oxygen Saturation Measured by Two Probes in Infants on the Neonatal Unit (Ag-OS)

September 7, 2023 updated by: University of Nottingham
On the neonatal unit in Derby, two types of probes are in routine use: Mindray 520 N and Mindray 521 N. Feedback from parents and staff show that there are concerns that the two probes given very different oxygen saturation readings and, often, do not agree with each other. This has raised concerns that infants' clinical care may be affected by the choice of probe. It is important to know if the two probes give similar results or not to ensure that infants get the appropriate monitoring and respiratory support as needed for good neonatal care. In this study, we will compare the reading made by the two probes and determine whether the readings of the two probes agree with each other and if any disagreement is such that clinical decision making is affected by the difference.

Study Overview

Status

Completed

Conditions

Detailed Description

Continuous monitoring of oxygen saturation level is a key component of neonatal intensive care and NICE recommends that all neonatal services "ensure that systems are in place for preterm babies to have a target saturation level of 91% to 95%" and that this is monitored using "continuous pulse oximetry" (NICE Quality statement 4 in Quality Standard QS193: https://www.nice.org.uk/guidance/qs193/chapter/Quality-statement-4-Oxygen-saturation). Oxygen saturation monitoring is used to provide optimal respiratory support to preterm infants to ensure adequate oxygenation of tissues while preventing risks of conditions such as retinopathy of prematurity and bronchopulmonary dysplasia [1]. It is equally important in term born infants who require intensive care such as for determining adequate oxygenation for respiratory support and screening for congenital heart disease [2]. For these reasons, most infants in neonatal units are continuously monitored using pulse oximetry, a simple and painless test that measures blood oxygen levels by passing a small beam of light through the infant's hand or foot. This measures the changes in light absorption to calculate the levels of oxygenated and deoxygenated blood. To do this the light (or probe) has to be attached to the infant's hand or foot using a tie or a sticker. The probe must be attached firmly for the saturation monitor to pick up the correct readings. As neonatal skin is delicate and it is often difficult to stick objects to it, neonatal oxygen saturation probes are carefully designed. Many different types of saturation probes are in routine use in neonatal units. It is important to ensure that they give accurate and comparable readings. In the neonatal unit in Derby, two types of probes are in routine use: Mindray 520 N and Mindray 521 N. Feedback from parents and staff show that there are concerns that the two probes given very different oxygen saturation readings and, often, do not agree with each other. This has raised concerns that infants' clinical care may be affected by the choice of probe. It is important to know if the two probes give similar results or not to ensure that infants get the appropriate monitoring and respiratory support as needed for good neonatal care. In this study, we will compare the reading made by the two probes and determine whether the readings of the two probes agree with each other and if any disagreement is such that clinical decision making is affected by the difference.

Study Type

Interventional

Enrollment (Actual)

20

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Derbyshire
      • Derby, Derbyshire, United Kingdom, DE3 0ED
        • University Hospitals of Derby and Burton

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

7 months to 1 year (Child)

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  • Infants who need oxygen saturation monitoring for clinical care
  • Infants born preterm (<37 weeks' gestation) who are < 7 days of age and need respiratory support including supplemental oxygen
  • Infants born preterm (<37 weeks' gestation) who are in supplemental oxygen at >34 weeks' gestation
  • Infants born at term gestation (< 37 weeks' gestation) who require respiratory support including supplemental oxygen
  • Infants with suspected cardiac disease

Exclusion Criteria:

  • Infants who have intravenous or arterial lines attached to either foot.
  • Infants who have any condition that may cause differential oxygen saturation or blood flow to the lower limbs such as venous thrombosis.
  • Infants who have local infections or skin lesions on either foot.
  • Infants older than 5 months of age

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: Quadruple

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: A
Right foot no.520 Left foot no.521

Two oxygen saturation probes will be used in this study:

Mindray 520 N is a disposable oxygen saturation sensor for neonates and adults. It is a disposable, single-use sensor that is wrapped around the infant's arm or foot. Its ends stick together giving a firm, close contact with the skin.

Mindray 521 N is another disposable oxygen saturation sensor for neonates and adults that is used on the neonatal unit in routine practice. As Mindray 520 N and 521 N are CE marked and used within their intended purposes we do not require a letter of no objection from the MHRA for that usage. The Mindray probes will be used without modification and there are no safety issues to the patient. Therefore, a letter of no objection from the MHRA is not required and this is not a medical device study under the Medical Devices Directive.

Experimental: B
Right foot no.521 Left foot no.520

Two oxygen saturation probes will be used in this study:

Mindray 520 N is a disposable oxygen saturation sensor for neonates and adults. It is a disposable, single-use sensor that is wrapped around the infant's arm or foot. Its ends stick together giving a firm, close contact with the skin.

Mindray 521 N is another disposable oxygen saturation sensor for neonates and adults that is used on the neonatal unit in routine practice. As Mindray 520 N and 521 N are CE marked and used within their intended purposes we do not require a letter of no objection from the MHRA for that usage. The Mindray probes will be used without modification and there are no safety issues to the patient. Therefore, a letter of no objection from the MHRA is not required and this is not a medical device study under the Medical Devices Directive.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
oxygen saturation of babies using Minday 520 N and Minday 521 N
Time Frame: 12 months
Oxygen saturation readings using Minday 520 N and Minday 521 N probes simultaneously
12 months

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
clinical decision making
Time Frame: 12 months
clinical decisions made, such as giving supplemental oxygen to infants on the neonatal unit following oxygen saturation readings from Minday 520 N and Mindray 521 N
12 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: shalini ojha, University of Nottingham

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

July 15, 2021

Primary Completion (Actual)

January 11, 2023

Study Completion (Actual)

January 14, 2023

Study Registration Dates

First Submitted

August 4, 2021

First Submitted That Met QC Criteria

September 28, 2021

First Posted (Actual)

October 4, 2021

Study Record Updates

Last Update Posted (Actual)

September 8, 2023

Last Update Submitted That Met QC Criteria

September 7, 2023

Last Verified

September 1, 2023

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • IRAS 297715

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Neonatal Disease

Subscribe